Copyright © 2013 Lippincott Williams & Wilkins. Unauthorized commercial reproduction of this slide is prohibited Supplemental PowerPoint Slides In vivo analysis of insertional torque during pedicle screwing using cortical bone trajectory technique Keitaro Matsukawa, M.D.,1* Yoshiyuki Yato, M.D.,1 Takashi Kato, M.D.,1, Hideaki Imabayashi, M.D.,1 Takashi Asazuma, M.D.,2 and Koichi Nemoto, M.D.1 Affiliations: 1Department of Orthopaedic Surgery, National Defense Medical College, Tokorozawa, Saitama, Japan 2Department of Orthopaedic Surgery, National Hospital Organization, Murayama Medical Center, Tokyo, Japan
2
Key Points 1)The mean maximum insertional torque using the CBT technique is about 1.7 times higher than the traditional technique (p < 0.01). 2)Positive linear correlations between maximum insertional torque and BMD were found in both techniques. 3)The correlation coefficient between insertional torque and BMD of the traditional screws (r = 0.63, p < 0.01) was higher than that of the CBT screws (r = 0.59, p < 0.01).