Calibration of clinometers for the alignment of muon chambers in the CMS experiment. MOSCHOUTIS Thomas FAS meeting – Orlando, 12 March 2004.

Similar presentations


Presentation on theme: "Calibration of clinometers for the alignment of muon chambers in the CMS experiment. MOSCHOUTIS Thomas FAS meeting – Orlando, 12 March 2004."— Presentation transcript:

1 Calibration of clinometers for the alignment of muon chambers in the CMS experiment. MOSCHOUTIS Thomas FAS meeting – Orlando, 12 March 2004

2 Global view of the Compact Muon Solenoid

3 View of one Disk Dual-axis clinometers for monitoring orientation of transfer pates:

4 Calibration bench CIinometer sensing Liquid calibrated Linear mover Θ

5 First setup Clinometer’s place Linear mover

6 Error calculation Error in φ: Error in φ: φ =Tan -1 (L/H)φ =Tan -1 (L/H) Δφ=(∂φ/∂L)*ΔL+(∂φ/∂H)*ΔH =(1/(L+H))*ΔL-2L/(H 2 +H*L)*ΔHΔφ=(∂φ/∂L)*ΔL+(∂φ/∂H)*ΔH =(1/(L+H))*ΔL-2L/(H 2 +H*L)*ΔH Using the same formula: Using the same formula: ΔΩ=(1/(L+M))*ΔL- 2L/(M 2 +M*L)*ΔMΔΩ=(1/(L+M))*ΔL- 2L/(M 2 +M*L)*ΔM Taking ρ=cos-1(M/K) and M=0 Taking ρ=cos-1(M/K) and M=0 Δρ=ΔM/KΔρ=ΔM/K So the total error is: So the total error is: Δθ =(Δφ 2 +ΔΩ 2 +Δρ 2 ) 1/2Δθ =(Δφ 2 +ΔΩ 2 +Δρ 2 ) 1/2 M K ρ φ L H

7 First test curve

8 Zero degree measurement Pendulum clinometer Long rod to get more precision

9 Problem induce by the plate thickness

10 Second test curve

11 The 6 different mounting plates Clinometer’s place

12 Next setup Linear mover Adapter Tilting base Mounting plate Click here to see movie


Download ppt "Calibration of clinometers for the alignment of muon chambers in the CMS experiment. MOSCHOUTIS Thomas FAS meeting – Orlando, 12 March 2004."