Download presentation
Presentation is loading. Please wait.
1
Rafael Ortega, MD Boston University Medical Center External Pacemakers
2
From the Trenches 70-year-old woman Returns to OR for expanding neck hematoma Cardiac arrest after intubation Management
3
OBJECTIVES Review External Pacemakers (EP) Summarize EP historical evolution Clarify EP indications Explain the Philips/Agilent HeartStart Unit
4
External Pacemaker Manual Defibrillator Automatic Defibrillator 3 in 1 Biphasic Unit
5
Question What is “Biphasic”?
6
A Few Words on “Biphasic” Biphasic energy is delivered in two directions Require less energy for defibrillation Less myocardial injury and myocardial dysfunction 150 J should be used for defibrillating ventricular fibrillation (in adults)
7
Waveform Comparison Monophasic Biphasic
8
Indications for EP Temporizing measure in patients with symptomatic bradycardia Little benefit in pulseless situations May be useful for overdrive-pacing in certain tachycardias.
9
Early History 1791: Galvani electrically stimulates dead frog’s heart 1872: Duchenne resuscitates child tapping precordium with an electrode 1932: Hyman designs external pulse generator: "artificial cardiac pacemaker“ 1952: Zoll uses simultaneous precordial and transesophageal electrodes for pacing
10
46 Years later Hesselvik JF. Ortega RA. Simultaneous transesophageal atrial pacing and transesophageal echocardiography in cardiac surgical patients. Journal of Cardiothoracic & Vascular Anesthesia. 12(3):281-3, 1998 Jun.
11
Who was Dr. Zoll? A Boston native Harvard Medical School Internship BI Hospital In 1939 joins a research group at BI WW-ll on Dwight Harken's surgical team Paul Zoll Professor Emeritus HMS
12
Transvenous Pacing Effect Temporary and permanent implantable transvenous pacemakers (late 1950’s) superseded the use of external models. But, these took time to insert There was still a need for external pacing
13
1981: A New Era Zoll patents and introduces a transcutaneous external pacemaker Longer pulse duration and larger electrode surface Reduced current required for capture and increased comfort for the patient New model could be applied much more rapidly
14
Pulse Duration and Current Early models used short (1-2 msec) impulses resembling the action potential of skeletal muscles Zoll increased duration to 4 msec with a 3-fold reduction in threshold Stimulation 20% over threshold stimulates only the ventricles. External pacing requires 30-100 times greater current than transvenous pacing
15
Question What is the effect of EP onhemodynamics?
16
Hemodynamics EP can simultaneously stimulate all 4 heart chambers (in dogs) Madsen echo demonstrated that atrial stimulation was retrograde without opening the mitral valve. Atrial-pacing threshold in humans is much higher than for ventricles Net result: loss of the “atrial kick” Cardiac output is reduced
17
Reducing Discomfort Not a problem under GA If awake, place electrodes in the midline chest and just below the left scapula Considered sedation
18
Question Can you perform CPR and use EP at the same time?
19
EP During CPR CPR can be performed with EP pads in place. No electrical hazard to the person performing CPR However, turning the unit off during CPR is advisable.
20
No Heart Damage No enzymatic, EKG, or microscopic evidence of myocardial damage has been found after pacing (dogs and humans) for as long as 60 minutes Low risk of triggering ventricular fibrillation
21
Tutorial
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.