Download presentation
1
SELECTING AND MANAGING CONTRACT STERILIZERS AND TESTING LABORATORIES
Pacific BioLabs Inc. (510)
2
CONTRACT STERILIZERS Written agreement between the contract sterilizer and the device manufacturer is required by 21 CFR (e), Medical Devices, processing, labeling or repacking
3
WRITTEN AGREEMENT MUST INCLUDE:
Instructions for maintaining proper records Acknowledgement that the device is non-sterile and awaiting processing Detailed specifications for the sterilization process Division of responsibilities
4
DIVISION OF RESPONSIBILITY
The device manufacturer is ultimately responsible for the sterility of its products The contract sterilizer is responsible for a quality control system for steps under its control Therefore, ……..
5
RESPONSIBILITIES OF MANUFACTURER
Ensure devices are sterile by confirming quality control system of contractor Validation of the process Ensure process controls are appropriate Maintain records of specification & evidence they were followed Product release
6
RESPONSIBILITIES OF CONTRACT STERILIZER
All requirements of the quality system regulation that apply to its operations Equipment installation, maintenance, and equipment calibration Installation and operation qualifications Building adequacy Personnel training Environmental controls
7
RESPONSIBILITIES OF CONTRACT STERILIZER (cont.)
Quality system requirements: Process controls Documentation and record keeping Acceptance of manufacturing materials Procedures for handling process deviations Any other requirement of the written agreement
8
COMMON PROBLEMS FOUND BY FDA
Manufacturer does not assume responsibility for sterilization performed by the contract sterilizer Communication between the two parties of changes Product is released following sterilization without approval of device manufacturer
9
TESTING LABORATORY Manufacturing Facility Contract Laboratory
Contract Sterilizer
10
LAB ATTRIBUTES Quality System FDA Track Record Experienced Personnel
SOP Calibrated equipment Training FDA Track Record Experienced Personnel Industry Reputation Committed Client Services
11
STERILIZATION QUALITY TESTS
Bioburden Bioburden Recovery Validation Sterility Bacteriostasis and Fungistasis Biological Indicators Pyrogen or Bacterial Endotoxin Inhibition and Enhancement
12
STERILIZATION QUALITY TESTS (cont.)
Package Integrity Chemical Residuals Environmental Monitoring Microbial Identification
13
BIOBURDEN Webster’s Medical Device bio (Greek) – life, living
burden – that which is carried, load Medical Device Population (number and types) of viable microorganisms on a product and/or a package
14
BIOBURDEN TEST Three steps Bioburden Recovery Validation
Removal of microbes from a product Culturing on a medium appropriate for growth Enumeration Bioburden Recovery Validation Demonstrates the adequacy of the steps above
15
BIOBURDEN METHODS Pour plating Spread plating Membrane filtration
Automated microbiology systems
16
STERILITY TESTS Product Sterility Biological Indicators (BI) Sterility
Membrane filtration Direct transfer Biological Indicators (BI) Sterility Testing the BI by placing it in microbiological growth medium
17
STERILITY TEST VALIDATION
Bacteriostasis and Fungistasis (B & F) Test using known organisms to demonstrate whether the combination of the product and sterilization method do or do not inhibit microbial growth
18
STERILITY TEST USP vs. AAMI/ISO
Two microbiological media SCDM and FTM B&F (6 organisms) AAMI/ISO One microbiological medium SCDM B&F (3 organisms)
19
BIOLOGICAL INDICATOR TESTS
Three quality attributes may be measured: Species identification Population Resistance D-Value Z-Value
20
BACTERIAL ENDOTOXIN TEST
Endotoxin = Pyrogen Bacterial Endotoxin Test = LAL Test Gel clot method Photometric techniques Photochromogenic Turbidimetric
21
INHIBITION AND ENHANCEMENT TEST
Validation of the LAL test 3 lots of product must be tested
22
DEVICE PACKAGING 21 CFR “Each manufacturer shall ensure that device packaging and shipping containers are designed and constructed to protect the device from alteration or damage during the customary conditions of processing, handling, and distribution”
23
DEVICE PACKAGING 21 CFR Package integrity is required to maintain the microbial barrier and sterility of the product Expiration dating Required for IVD’s and certain other devices that may degrade over time If label includes expiration, you must have data to back it up
24
PACKAGE INTEGRITY TESTS
Dye Penetration Microbial Challenge Burst Creep Seal peal
25
CHEMICAL RESIDUALS (EO ONLY)
EO (ethylene oxide) ECH (ethylene chlorohydrin) EG (ethylene glycol) No longer required
26
ENVIRONMENTAL MONITORING
Conducted in-house Laboratory field technicians conduct sampling at manufacturing facility Samples collected by sponsor’s personnel and incubated/counted at testing laboratory
27
ENVIRONMENTAL MONITORING TESTS
Non-viable particles Viable particles Bacteria Soybean Casein Digest Medium (SCDM) 30 – 35 °C Yeast and Mold Sbouraud Dextrose Agar (SDA) 20 – 25 °C Growth promotion required
28
MICROBIAL IDENTIFICATION
Morphological characterization Colony description, Gram stain Identification to genus and species level Phenotypic Genotypic
29
MICROBIAL IDENTIFICATION
Biochemical Tests Vitek Fatty Acid Analysis MIDI Carbon Utilization Biolog Genetic Based Technology (preferred) RiboPrinter
30
MICROBIAL WATER TESTS Total Heterotrophic Counts Yeast and Mold Counts
Coliform Determinations Thermophiles Counts Anaerobes counts Pseudomonas screening
31
WATER TESTS Choice of Media Choice of Incubation Temperature
Plate Count Agar Soybean Casein Digest Agar R2A Tryptone Glucose Extract Choice of Incubation Temperature 30-35°C, 20-25°C Choice of Incubation Times 2, 3, 5 or 7 days
32
THANK YOU Q & A
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.