Download presentation
Presentation is loading. Please wait.
1
Probestation 1: 6“ (200mm * 300mm + 360° rotation + 15mm Z) QTC Instruments: HP4274A (LCR), K6517 (electrometer), HP6614C (voltage supply), K595 (quasistatic CVmeter) Status: Operational (used to measure Hamamatsu) Software (user friendly and some error protections) Probestation 2: 6“ (200mm * 100mm + 15mm Z) IQC Instruments: LCR, electrometer, voltage supply Status: Operational Misc: Homemade relay system (17 channels each) Optical Inspction: Zeiss CMM (500*500*500) Res.: 1 m (Z~5 m) Hardware setup
2
Chuck
3
Probestation
4
Error estimation 1000 measurement runs on one strip (ST teststructure) ParameterMean MinMax Rpoly (M ) 1,0640,001371.06051.0687 Leakage current (pA) 62,12,259,064,0 Coupling capacitance (pF) 108,30,78106,13108,57
5
Software – Front panel Testing time ~ 2h
6
Probing marks Hamamatsu Karlsruhe
7
Pick up tool
8
Packaging (Hamamatsu) Sensor Teststructures Sandwich packaging in envelope Bad to handle Optical inspection HANDLING against specification Takes time Air cusion effect while placing gliding
9
Packaging 2 (ST) Optical inspection very good Easy to take out All 4 TS available
10
Optical survey and dimensional measurements
11
Check Sheet:
12
Pictures 25 m
13
More pictures Scratchpad used RpolyAdd DC pad 3 AC pads
14
Electrical characterization of sensors Rpoly with 10VRpoly with 2V Changes in measurement with respect to specs
15
Electrical characterization of sensors Summary
16
Electrical characterization of sensors CV & IV
17
Electrical characterization of sensors Rpoly
18
Electrical characterization of sensors coupling capacitance
19
Electrical characterization of sensors leakage current
20
Comparison of our test results with industry results (based on 5 sensor of 8 ) IV even lower than company measurement CV perfect agreement Some Rpoly higher than company measurement (but company only measured 1 per batch or even TS only) No single strip fault (agreement) (neglect Rpoly)
21
Relevant problems Nothing Serious V depletion not calculated Improvements: Automatic web publishing (ready 50%) Implement Pattern recognition
22
Optical survey of ST prototype Pinhole or bad coupling capacity No strange electrical behavior
23
Optical survey of ST prototype 2 Z ( m) Bowl shape! Asymmetric cut! Global size OK!
24
Pads
25
Electrical Survey Very high total leakage current Depletion voltage fine One region (7 strip) with a very high leakage current Some coupling capacitors out of specs
Similar presentations