Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection.

Similar presentations


Presentation on theme: "Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection."— Presentation transcript:

1 Calibration of the VA2 B Pal, R Mountain, M Artuso 6/29/2010 Goal: get an absolute calibration of a VA2 channel to measure diamond charge collection

2 Calibration fixture is replaced by single channel fixture

3 Single channel fixture schematic 1.0 pF

4 Channel Gain Channel 141 previously used in diamond measurements

5 Input Bias Nominal gain from specs is 30 mV/fC, which is ½ of the gain we measured in ch 141. Nominal value value Used Vfp-0.2 to 0.5 V-0.4 V Vfs700 mV Pre_Bias100 mV99.8 mV Sha_Bias22 mV22.3 mV i_buf14 mV13.8 mV

6 Gain vs. I_buf

7 To do Calibrate the ADC (have adc/fC now), Check the gain after ADC calibration, Fix the diamond holder, Start studies of “unusual” diamond behavior.