Lic. Pedro Antonio Pérez.  To develop a simple and versatile dosimetric method capable of determining changes in matter (Xylenol Orange added Fricke.

Similar presentations


Presentation on theme: "Lic. Pedro Antonio Pérez.  To develop a simple and versatile dosimetric method capable of determining changes in matter (Xylenol Orange added Fricke."— Presentation transcript:

1 Lic. Pedro Antonio Pérez

2  To develop a simple and versatile dosimetric method capable of determining changes in matter (Xylenol Orange added Fricke gel) due to ionizing radiation.  To design and construct a preliminar device (hardware and software) in order to quantify absorbed dose by means of visible light absorption.  To establish applicability limits of extremely simple devices. P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry2

3  Nowadays, the development of novel dosimetric techniques has become great interests.  Optical analysis: Standard Fricke solution presents maximum absorption at 294nm (ultraviolet).  X. O. Fricke gel has maximum absorption at 585nm (visible).  Spatial resolution can be obtained fixing the X.O. Fricke solution to a gel matrix.  X. O. Fricke gel dosimeters, which constitute tissue- equivalent media for radiotherapy purposes, are capable of providing real 3D dose mapping. 3 P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

4  Boltzmann Radiation Transport Equation  Parallel beam   a  s 4 P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

5  Contiuous light source  Then,  Wich solution is the Lambert-Beer eq. 5 P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

6  From L-B  Light absorption law  z≈Cd, then  And the Optic al Density differences result: 6 P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

7  Dosimeter preparation:  Chemical composition  Protocol 7 P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

8  Experimental Set Up:  Dispositive description 8 P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

9  Developed software: 9 Get imageGet red channel Calculate X CM Calculate 2° moment Calculate  (Rm) Get ROI Get definitive image Calculate VM on new ROI and new  P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

10 10 Radiation facilities: Conventional X-Ray tube  W anode. Mo cathode. V=40kV. 40mA CLINAC  Varian 6/100 Materials and methods P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

11 11 Results Experimental Set Up characterization. Preliminary tests. Repetitivty and stability of the sistem source-dosimeter-camera Results: Stability: I=25,6+0,6Repetitity: I=26,7+0,4 Diference on both modes: 4% P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

12 12 Software validation by Gauss function  Some results  Some distributions (i c,j c ) y  escrito(i c,j c ) calculado (50,25) y 10(50,25) (25,50) y 10(25,50) (50,50) y 5(50,50) (75,75) y 5(75,75) Measurement and data processing Experimental Set Up characterization. Preliminary tests. P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

13 13 Software validation  2 VM known functions  Constant I píxel to píxel (VM)  Variable I pixel to pixel (ROI) Results:  Fixed I: VM=1,  =0  Variable I: Measurement and data processing Experimental Set Up characterization. Preliminary tests. P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

14 14 Characterization of the developed device for optical analysis Results:  Serie 1:  OD=0,16+0,01. Error: 6,7%.  Serie 2: characteristic curve “exposition time-response” Measurement and data processing On X-Ray tube. P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

15 15  Two elaboration series to be irradiated by means of CLINAC: characteristic curve of “dose-response” and “depth dose” (to compare with standard dosimetric system -ionization chamber).  One elaboration serie to be irradiated by means of X-Ray tube and repeat the experience to obtain characteristic curve “exposition-response” Measurement and data processing On CLINAC P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

16 16 Irradiated samples Dose on depth and comparison with ionization chamber Preliminary measurements On CLINAC – in-depth dose P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

17 17 Discussion, conclusions and outlook Software validation  It has been verified the correct determination of relevant qquantities (VM, Rm, CM and ROI).  It has been established the consistency for the realization of the diferent steps relative to communication with hardware, signal acquisition and data interpretation and subsequent processing.  Response of the integral system (source-dosimeter-software)  It was possible to stablish the REPRODUCIBILITY (non-distinguishable), stability (less than 1.6%) and repetitivity (less than 2.4%).  The reults show clearly linearity region, as it was expected, even with supra and sub linear zones.  Response curves, which have been found to be indistinguishable, demonstrated the reproducibility of the whole dosimeter elaboration procedure. In addittion  Device dimensions (20cm side box) allow its portability, therefore improving its comfortability for CLINAC measurements.  Preliminary PDD determination, as compared with ionization chamber, show hopeful results. P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry

18 18 Main conclussions  It was demonstrated that, even considering extremely simplcity of the the device component, it is possible to develop suitable systems capable of determining absorbed dose by means of visible light transmittance measurements, particularly with radiochromic dosimeters. Perspectives  A lot of future improvements are been considered for the developed device regarding its applicability limits and accuracy. Specially, a better detector quality may improve significantly the overall performance.  Several techniques devoted to 1D and 2D dose mapping are under investigation. Discussion, conclusions and future lines P. Pérez, V. Galván, M. Valente - Preliminary study about the feasibility of developing novel analysis methods based on visible light absorption for Fricke dosimetry