Low Frequency and Medium Frequency Currents
Foundations in Electro 2 Low Frequency and Medium Frequency Currents
2
OBJECTIVES Review on the difference between high, medium, and low medium frequency currents and their therapeutic/clinical implications
3
OBJECTIVES Be familiar with terms used in electrotherapy current modulations particularly with WAVEFORM, (electric) PULSE, FREQUENCY, CURRENT INTENSITY, PULSE DURATION
4
OBJECTIVES Enumerate the characteristics of the three types of low/medium frequency currents Formulate guidelines in the selection and/or prescription of the most appropriate (low/medium frequency current) electrical modality
5
OBJECTIVES Describe the basic design features of electrical stimulators Be familiar with the clinical importance of the design features of electrical stimulators
6
OBJECTIVES Identify the common controls present on electrical stimulator units Be familiar with the parameters regulated by each control present on the electrical stimulator units
7
A review on the differences…
High Frequency Currents Medium Frequency Currents Low Frequency Currents
8
HIGH FREQUENCY CURRENTS
Frequency is >6000 Hz Short wavelengths (<10 mm) Effects occur only at superficial structures General effect = HEATING Sample modalities: US, MWD, SWD, IRR, UVR, LASER
9
MEDIUM and LOW FREQUENCY CURRENTS
Frequency ranges from 1 to 6000 Hz Longer wavelengths (>10 mm) Effects occur at deeper structures General effects: MFC: blocks pain LFC: nerve stimulation
10
MEDIUM and LOW FREQUENCY CURRENTS
Sample modalities: Electrical stimulators, Diadynamics, Biofeedback, Iontophoresis, TENS, IT
11
Definition of some relevant terms…
ELECTRIC PULSE PULSE DURATION CURRENT INTENSITY WAVEFORM FREQUENCY
12
ELECTRIC PULSE A unit of stimulating current
Otherwise known as a PHASE (current phase)
13
ELECTRIC PULSE Can be more fully described according to DURATION (pulse duration expressed in seconds), INTENSITY (current intensity expressed in amperes or volts), and SHAPE (waveform)
14
PULSE DURATION Amount of time needed for the rise and fall pattern to occur at a given pulse Expressed in SECONDS (millisecond=ms)
15
CURRENT INTENSITY Rate of flow of electrons
Usually expressed in AMPERES (milliamperes = mA)
16
WAVEFORM Describes the rise-and-fall pattern of a pulse
The shape of the waveform reflects the time required for the current to reach the maximum intensity
17
WAVEFORM Waveforms with sudden rise in intensity are suitable for innervated muscle Waveforms with slowly rising intensity are best suited for denervated muscle
18
FREQUENCY Rate of change of an electrical pulse
Expressed in HERTZ (Hz)
19
Therapeutic/Clinical Uses…
MEDIUM and LOW FREQUENCY CURRENTS
20
MEDIUM and LOW FREQUENCY CURRENTS
Assists in functional training Assists in muscle force generation and contraction Decreases unwanted muscle activity Increases rate of healing of open wounds and soft tissues
21
MEDIUM and LOW FREQUENCY CURRENTS
Helps maintain muscle integrity after surgery Modulates and/or decreases pain Decreases or eliminates soft tissue swelling, inflammation, or restriction
22
TYPES OF MEDIUM-LOW FREQUENCY CURRENTS
Direct Currents Alternating Currents Pulsed Currents
23
Noted Characteristics…
QUANTITATIVE: Frequency (Hz) Pulse duration
24
Noted Characteristics…
QUALITATIVE: Number of PHASES Shape and symmetry of WAVEFORMS Other qualitative characteristics
25
Direct Current Refers to a current passing continuously in the same direction (unidirectional current)
26
Direct Current (cont.) Synonyms: Constant Current Galvanic Current / Galvanism Galvanic stimulation is useful only for stimulating denervated muscles
27
Direct Current (cont.) Interrupted Direct Current (IDC) is used to stimulate innervated muscles Direct current is also used in IONTOPHORESIS
28
Direct Current (cont.) 2 Types of IDC: Long Duration IDC > 1 ms For sensory and motor nerve stimulation (denervated)
29
Direct Current (cont.) Short duration IDC (Faradic-Type) < 1 ms For pain control and nerve stimulation (innervated)
30
Direct Current (cont.) Physiological effects: Sensory stimulation Hyperemia Electrotonus Relief of pain Acceleration of healing Tissue destruction
31
Alternating Current Defined as continuous or uninterrupted bidirectional flow of charged particles
32
Alternating Current (cont.)
2 Types: Sinusoidal Current Evenly alternating sine wave currents of 50 Hz For pain relief, edema, and improvement of circulation
33
Alternating Current (cont.)
Diadynamic Current Rectified monophasic sinusoidal current For pain relief, tissue healing, muscle re-education and improvement of circulation
34
Pulsed Current Defined as the uni- or bi-directional flow of charged particles that periodically ceases for a finite period of time
35
Pulsed Current (cont.) Types: Symmetrical Biphasic Balanced Asymmetrical Biphasic Unbalanced Asymmetrical Biphasic Monophasic
37
GUIDELINES… Selecting, Prescribing, or Purchasing the MOST APPROPRIATE Electromodality
38
GUIDELINES… Determine your treatment goals
Note for the presence of contraindications Determine the usual conditions of or problems presented by patients of the facility/area
39
GUIDELINES… Consider the market availability of the modality and its cost Consider the requirements for maintenance of the modality
40
BASIC DESIGN FEATURES and CONTROLS…
Electrical Stimulators
41
BASIC DESIGN FEATURES Path from power source to the unit (plugs and cables) Control knobs and/or buttons Electrodes (with cables) Alternative power source Safety features
42
BASIC DESIGN FEATURES Controls or adjustment knobs/buttons for:
Frequency Intensity Mode (continuous or pulsed) Pulse Duration and Intervals Treatment Duration
43
Basic Electrode Systems
Malleable metal electrodes Electrodes that conform to the body surfaces Water Bath
44
Malleable Metal Electrodes
Made of tinplate or aluminum with pad of lint, cotton gauze or sponge at the end Pad/gauze/sponge is wet with water before being applied to skin Electrodes kept in place with bandages / straps
45
Malleable Metal Electrodes (cont.)
If unequal in size, the smaller electrode is active & most effects will occur here; the other electrode is the indifferent or dispersive electrode
46
Electrodes that Conforms to the Body Surface
Made of carbon-impregnated silicone rubber Used with sponge pads or thin layer of conducting gel Kept in place with strap or adhesive tape
47
Electrodes that Conforms to the Body Surface
Less efficient in passing current than metal electrodes Has lower impedance than polymer electrodes
48
Water Bath Used for hand, forearm, foot and leg which is placed between the electrodes Provides a large area for the indifferent electrode & for applying muscle stimulating currents Current density depends on location of electrodes
50
Methods of Electrode Placement
UNIPOLAR BIPOLAR QUARDRIPOLAR
51
Unipolar Motor Point Stimulation
One small active electrode & one large dispersive electrode Site of stimulation: motor point for stronger response
52
Unipolar Motor Point Stimulation
Same amount of current passes thru each electrode Smaller sized electrode has higher current density (stronger effect)
54
Unipolar Motor Point Stimulation (cont.)
Used for innervated and denervated muscles Indications: Peripheral nerve injuries Tendon transplants
55
Unipolar Motor Point Stimulation (cont.)
Contraindications: Cases wherein active motion is prohibited Patients with pacemakers Directly over superficial metal implants
56
Unipolar Motor Point Stimulation (cont.)
Contraindications: Active bleeding over treatment site Malignancies over treatment site
57
Unipolar Motor Point Stimulation (cont.)
Precautions: Sensory loss over treatment site Open wounds Extreme edema
58
Bipolar Motor Point Stimulation
Equally sized electrodes Effect of stimulation is dependent on electrode placement Current density is equal in both electrodes Effective for stimulating muscle groups or very large muscles
60
Bipolar Motor Point Stimulation (cont.)
Used for innervated and denervated muscles Indications: Peripheral nerve injuries Inhibition of muscle activity due to joint pain and effusion
61
Bipolar Motor Point Stimulation (cont.)
Indications (cont.): UMN lesions to decrease spasticity & facilitate active contraction Disuse atrophy Immobilization Orthopedic & neurological cases with LOM
62
Bipolar Motor Point Stimulation (cont.)
Contraindications & Precautions: Same as Unipolar application
63
Quadripolar Motor Point Stimulation
Electrodes from two or more circuits positioned so that currents geometrically intersect Used for Interferential Stimulation Technique (MFC)
65
Quadripolar Motor Point Stimulation (cont.)
Indications: Pain & muscle spasm Edema Hematoma Chronic ligamentous lesions Urinary stress incontinence
66
Quadripolar Motor Point Stimulation (cont.)
Contraindications & Precautions: Same as Unipolar application
67
Break? or Break?
68
Importance of Stimulation Parameters
The effect of electrical stimulation on the tissue will depend on the rate of change of the electrical pulse: No change / Slow change in electric pulse IONTOPHORESIS / DIRECT CURRENT
69
Importance of Stimulation Parameters (cont.)
Very fast change of rate HIGH FREQUENCY CURRENTS Rate of change between nos. 1 & 2 LOW & MEDIUM FREQUENCY CURRENTS
70
Importance of Stimulation Parameters (cont.)
The current intensity determines the extent of physiological changes When stimulating a muscle at a constant frequency the only way to increase the force produced is to recruit more motor units by increasing the intensity of stimulation
71
Importance of Stimulation Parameters (cont.)
A single pulse is described by their: Duration Seconds / Milliseconds / Microseconds Intensity Milliamps / Volts Shape Illustrates the change of intensity with time
72
Importance of Stimulation Parameters (cont.)
The relationship between time and current intensity is the rate of change in current
73
Current used in Galvanic current and Iontophoresis
74
Current used for Nerve Stimulation
75
Current used for producing Single Nerve Impulse
76
Current used for TENS & Faradic Stimulators
77
Surged current producing Muscle Contraction
78
Current Flow in the Tissues
The quantity of current that flows in the tissues and the path it follows will depend on the impedance of that pathway Generally, watery tissue such as blood, muscle and nerve has low ohmic resistance
79
Current Flow in the Tissues (cont.)
Bone and fat has higher ohmic resistance The epidermis has the highest ohmic resistance, which is determined by: Thickness & nature of skin Inter-electrode distance
80
Current Flow in the Tissues (cont.)
This electrical resistance can be reduced by: Washing & wetting the surface Warming the skin
81
Skin Irritation as an Adverse Response
Skin irritation may be caused by: Electrical reaction Electrochemical response Allergic response to electrodes, gel, or tape
82
Skin Irritation as an Adverse Response (cont.)
Skin irritation may be caused by: Mechanical irritation caused by shearing forces between adhesive substances and the skin
83
Hazards in Electrotherapy
Chemical damage due to inadequate skin protection when direct or interrupted current is used Disruption of stimulating devices due to proximity of diathermy output Skin irritation Electric shock
84
Contraindications to Electrical Stimulation
Strong muscle contraction might cause joint/muscle damage; detachment of thrombus; spread of infection; and hemorrhage Stimulation of autonomic nerves might cause altered cardiac rhythm or other autonomic effects
85
Contraindications (cont.)
Currents might be unduly localized due to open wounds or skin lesions Currents might provoke undesirable metabolic activity in neoplasms or in healed tuberculous infections
86
Contraindications (cont.)
Current is not evenly phasic, leading to possible skin damage or irritation, especially if there is loss of sensation
87
Principles of Application
Conduct general safety checks with respect to the equipment Check the patient for contraindications Explain the treatment fully to the patient
88
Principles of Application (cont.)
Collect the necessary equipment ES, electrodes, wiring Soap & water for cleaning the skin Contact gel / sponge, tape / straps / Velcro
89
Principles of Application (cont.)
Position the patient in the comfortable position The skin should be uncovered & examined for any contraindications to treatment Test the equipment as appropriate; demonstrate the technique to the patient
90
Principles of Application (cont.)
Wash the skin over the region of electrode contact. Soaking the skin for 3-4 min either in a bath or with a warm, damp pad may reduce skin resistance Select appropriate treatment parameters
91
Principles of Application (cont.)
Always turn all intensity dials to zero before beginning the treatment Place the electrodes as appropriate for the treatment
92
Principles of Application (cont.)
Increase intensity until desired result is produced Never lift the active electrode from the skin or replace it without turning the intensity to zero
93
Principles of Application (cont.)
Terminate the treatment; check the skin condition Keep a full record of the treatment
94
THANK YOU!