1 Electrical Surgical Unit
Dr Fadhl Al-Akwaa Please contact Dr Fadhl to use this material

2 Please contact Dr Fadhl to use this material
ESU Importance Radio frequency current has been used for many years to reduce time under anesthesia and improve the clinical result. The ability to simultaneously cut tissue and coagulate blood vessels greatly reduces morbidity and mortality, and enables a large number of procedures that are otherwise not possible, especially in neurosurgery. Please contact Dr Fadhl to use this material

3 Please contact Dr Fadhl to use this material
AGENDA 81.1 Theory of Operation 81.2 Monopolar Mode 81.3 Bipolar Mode 81.4 ESU Design 81.5 Active Electrodes 81.6 Dispersive Electrodes 81.7 ESU Hazards 81.8 Recent Developments Please contact Dr Fadhl to use this material

4 Please contact Dr Fadhl to use this material
Terminology Cutting Coagulation= fulgurate + desiccation Desiccation OR contact coagulation. تجفيف: process of drying food Fulgurate: (fulgur is Latin for lightning). destroy using electricity (such as a tumor, etc.) Hemostasis: stoppage of bleeding Blended mixed, merged, combined choice of blend waveforms to allow the surgeon to select the degree of hemostasis desired. Bleeding RF frequency 300 kHz and 5 MHz Cautery: process of burning or scarring with a hot iron or other implement الكي Ablating: remove by cutting Please contact Dr Fadhl to use this material

5 Electrosurgery vs electrocautery
Electrosurgery is sometimes erroneously referred to as electrocautery. In cauterization the essential physical mechanism behind the treatment is conduction heat transfer from a hot object placed on the surface to raise the temperature high enough to denature the tissue proteins. Cutting and coagulation by means of electrosurgery is also accomplished by heating tissue to high temperatures, but the essential difference is that the primary mechanism is electrical power dissipation directly in the affected tissues themselves, rather than heat transfer from an external hot object on the tissue surface. Please contact Dr Fadhl to use this material

6 Electrosurgery vs Diathermy
Electrosurgery is also referred to as surgical diathermy, particularly in Europe. diathermy literally means through-heating, such as might be applied in physical medicine for the relief of pain or in hyperthermia therapy for tumor treatment. Electrosurgical devices in operating rooms are designed and built for surgical use only. Please contact Dr Fadhl to use this material

7 Please contact Dr Fadhl to use this material
Clinical application Electrosurgery is commonly used in dermatological, gynecological, cardiac, plastic, ocular, spine, ENT, Ear Nose & Throat orthopedic, urological, neuro- and general surgical procedures. Please contact Dr Fadhl to use this material

8 Duty cycle=on time/off time
Continuous Interrupted Please contact Dr Fadhl to use this material

9 Crest Factor The Crest Factor is equal to the peak amplitude of a waveform divided by the RMS value. Electrical engineering — for describing the quality of an AC power waveform

10 Current density (high or low)
Please contact Dr Fadhl to use this material

11 Electrosurgical modes
Please contact Dr Fadhl to use this material

12 Electrosurgical modes
Monopolar mode AND Bipolar mode The most noticeable difference between these two modes is the method in which the electric current enters and leaves the tissue. In the monopolar mode, the current flows from a small active electrode into the surgical site, spreads through the body, and returns to a large dispersive electrode on the skin. In the bipolar mode, the current flows only through the tissue held between two forceps electrodes. The monopolar mode is used for both cutting and coagulation. The bipolar mode is used primarily for coagulation. Please contact Dr Fadhl to use this material

13 Please contact Dr Fadhl to use this material
History Development of the first commercial electrosurgical device is credited to Dr. William T. Bovie, who worked on it from 1914 to 1927 while employed at Harvard University[8] [9] The first use of an electrosurgical generator in operating room occurred on October 1, The surgery was performed by Dr. Harvey Cushing. Please contact Dr Fadhl to use this material

14 Theory of Operation: Temperature
What is the general effects of heat on biologic tissue? What happen when tissue temperature increase from normal body temperature to 45°C? neither microscopic nor macroscopic changes, some cytochemical changes do in fact occur. the cells return to their normal function when the temperature returns to normal values. Please contact Dr Fadhl to use this material

15 Please contact Dr Fadhl to use this material
Theory of Operation What is the general effects of heat on biologic tissue? Above 45°C, Normal cell functions was inhibited and lead cell was death. Further increasing the temperature up to 100°C leads to tissue drying; that is, the aqueous cell contents evaporate. This process is called desiccation. If the temperature is increased beyond 100°C, the solid contents of the tissue reduce to carbon, a process referred to as carbonization. Please contact Dr Fadhl to use this material

16 Please contact Dr Fadhl to use this material
Bioheat equation Where T and To are the final and initial temperatures (K), σ is the electrical conductivity (S/m), ρ is the tissue density (kg/m3) C is the specific heat of the tissue (Jkg-1k-1) J is the current density (A/m2), and t is the duration of heat applications Please contact Dr Fadhl to use this material

17 Please contact Dr Fadhl to use this material
Bioheat equation The bioheat equation is valid for short application times where secondary effects such as heat transfer to surrounding tissues, blood perfusion, and metabolic heat can be neglected. the surgeon has primarily three means of controlling the cutting or coagulation effect during electrosurgery: the contact area between active electrode and tissue, The electrical current density, and the activation time. output power vs. tissue impedance Please contact Dr Fadhl to use this material

18 Please contact Dr Fadhl to use this material

19 Please contact Dr Fadhl to use this material
Output values for ESUs, Please contact Dr Fadhl to use this material

20 Please contact Dr Fadhl to use this material
Waveform Interrupted waveforms, such as exponentially damped sinusoids (obtained from spark gap or other relaxation-type oscillators) are effective for coagulation techniques requiring fulguration, or intense sparking (fulgur is Latin for lightning). Continuous sine waveforms (e.g., those obtained from vacuum tube or transistor oscillators) have proven most effective for cutting. Please contact Dr Fadhl to use this material

21 Please contact Dr Fadhl to use this material
Cut/Coag Cut/Coag Most wet field electrosurgical systems operate in two modes; Cut where the tissue is vaporized and Coag where the tissue is dried. Cut If the voltage level is high enough then the heat generated can generate a vapour pocket. The vapour pocket typically reaches temperatures of approximately 400 Degrees Centigrade which vaporizes soft tissues. Coag When the system is operating in coag mode the voltage output is usually lower than in cut mode and less power is delivered. This therefore generates less heat and a vapour pocket is not generated. Please contact Dr Fadhl to use this material

22 Please contact Dr Fadhl to use this material
Why high frequency? sensitivity to electrical stimulation decreases with increasing frequency in the kHz range and above. Neural and muscle cells are electrically-excitable, i.e. they can be stimulated by electric current. In human patients such stimulation may cause acute pain, muscle spasms, and even cardiac arrest. To minimize the effects of muscle and neural stimulation, electrosurgical equipment typically operates in the radio frequency (RF) range of 100 kHz to 5 MHz. Please contact Dr Fadhl to use this material

23 high-frequency power amplifiers
vacuum tube/ spark gap 1907, Lee De Forrest, power transistors, solid-state electrosurgical generators(Valleylab and EMS) 1970 Please contact Dr Fadhl to use this material

24 Please contact Dr Fadhl to use this material
ESU Design RF Generators for Electrosurgery “high-frequency power amplifiers” control input switches, safety monitor. Microcontroller Power supply Microcontroller Electrodes Active electrode return electrode, Please contact Dr Fadhl to use this material

25 Block diagram for typical solid state ESU
Please contact Dr Fadhl to use this material

26 Please contact Dr Fadhl to use this material

27 Please contact Dr Fadhl to use this material
Safety Appliances. American National Standard for Electrosurgical Devices the return electrode losing its attachment to the skin>> skin burns at the dispersive, or return, electrode site. Patient return monitor circuit Please contact Dr Fadhl to use this material

28 Electrosurgery Analyzers
QA-ES Electrosurgery Analyzer Please contact Dr Fadhl to use this material

29 Please contact Dr Fadhl to use this material
Performance testing Tests include: • Automatic power distribution measurement • Crest factor measurement • RF leak measurement • Return electrode monitor (REM) test Please contact Dr Fadhl to use this material

30 Please contact Dr Fadhl to use this material
Power Output Tests output current (A), power (W), peak-to-peak voltage (V), crest factor values. Please contact Dr Fadhl to use this material

31 HF Current Leakage Test
This test checks whether the active and dispersive leakage currents are within acceptable limits. Please contact Dr Fadhl to use this material

32 Measuring Leakage Current with HF Isolated Equipment
Please contact Dr Fadhl to use this material

33 Measuring Leakage Current with Grounded HF Equipment
Please contact Dr Fadhl to use this material

34 Please contact Dr Fadhl to use this material
REM Alarm Test The REM (Return Electrode Monitoring) alarm test ensures that the ESU sounds an alarm if the resistance between the two neutral electrodes exceeds its specified limit Please contact Dr Fadhl to use this material

35 Electrosurgery Analyzers
Typical measurement parameters: RF current (ma) RF power (watts) Crest Factor (CF) – the measure of Vpeak / VRMS 1.414 to 1000 range Duty cycle (pulsed waveforms) 1% - 100% range Through “non-inductive” test loads from 0Ω to > 5000Ω 5th generation parameters include Frequency Pulsed parameters such as time on, time off, total cycle time, duty cycle