BF DICDFPh Introduction to optical microscopy
2
Contrast techniques in optical microscopy Introduction to optical microscopy Some basic optics Properties of lens Types of illumination Resolution Contrast Phase contrast Differential Interference Contrast Thanks to Guy Cox (Electron Microscope Unit, University of Sydney) for contribution of many of these slides
3
Refraction –Direction change of a ray of light passing from one transparent medium to another with different optical density. A ray from less to more dense medium is bent perpendicular to the surface, with greater deviation for shorter wavelengths Refraction But it is really here!! He sees the fish here….
4
Refraction Light is “bent” and the resultant colors separate (dispersion). Red is least refracted, violet most refracted. dispersion Short wavelengths are “bent” more than long wavelengths
5
Reflection vs Refraction refractive index Snell’s Law: The angle of reflection (Ø r ) is equal to the angle of incidence (Ø i ) regardless of the surface material The angle of the transmitted beam (Ø t ) is dependent upon the composition of the material tt ii rr Incident Beam Reflected Beam Transmitted (refracted)Beam n 1 sin Ø i = n 2 sin Ø t The velocity of light in a material of refractive index n is c/n
6
Polarization and Phase: Interference Electric and magnetic fields are vectors - i.e. they have both magnitude and direction The inverse of the period (wavelength) is the frequency in Hz Wavelength (period T) Axis of Magnetic Field Axis of Propagation Axis of Electric Field Modified from Shapiro “Practical Flow Cytometry” 3 rd Ed. Wiley-Liss, p78
7
Interference Constructive Interference Destructive Interference A B C D A+B C+D Amplitude 0o0o 90 o 180 o 270 o 360 o Wavelength Figure modified from Shapiro “Practical Flow Cytometry” 3 rd ed Wiley-Liss, p79 Here we have a phase difference of 180 o (2 radians) so the waves cancel each other out The frequency does not change, but the amplitude is doubled
8
Microscopy:generating, collecting and recording photons from a small source. o
9
o
10
Properties of thin Lenses f 1 p + 1 q = 1 f f p q Magnification = q p
11
In a microscope the sample is further from the lens than the focus, so a real, magnified and inverted image is formed. Parallel rays are deflected through the focus; rays through the centre of the lens go straight through, so we can plot where this image will be.
12
If we move the lens closer to the object the image will get larger and further away - so with a greater 'tube length' - a longer microscope tube - we could get higher magnification with the same lens.
15
However having the image at different planes is inconvenient and screws up our optical corrections, so we use an objective of shorter focal length to form a more highly magnified image at the same plane. Shorter focal length = higher magnification.
16
However having the image at different planes is inconvenient and screws up our optical corrections, so we use an objective of shorter focal length to form a more highly magnified image at the same plane. Shorter focal length = higher magnification.
17
The eyepiece is placed close to the real image formed by the objective; since this is closer to it than its focus, it cannot form a real image. The rays are deflected towards the optic axis but are still diverging from each other.
18
These rays will therefore appear to an eye, placed beyond the eyepiece, as if they come from a virtual image which is further magnified but the same way up as the real image (and still upside-down relative to the object).
19
Types of microscopy and sources of light. Reflectance Surface reflection and absorption, colour Transmittance Absorption, colour, refractive index Incandescence Luminescence Fluorescence Excitation & emission
20
Transmission Optics:Conjugate planes in generalized microscope Köhler illumination sample
21
So what makes a “good” image? The most fundamental property of a microscope is its resolution. We are all familiar with the term but sometimes we don't really think about what it means. We may have the vague idea that it means ‘how small a thing we can see’, but this isn't really accurate. Actually we can see things that very tiny indeed but the important thing is the ability to distinguish objects as separate, and that is what resolution means. What determines resolution? One or two?
22
The first point is that it depends to some extent on your specimen! In fluorescence, the sample is made up, for imaging purposes, of a collection of luminous points (fluorochrome molecules) each emitting light independently (non coherent). In transmitted light the illumination makes a big difference to the resolution we get, because the individual points being imaged are not independent of each other. There is partial coherence.
23
The image results from the ‘convolution’ of the imaging devise with the object Transmitted image of a tree on the ground (usually called a shadow). But the imaging device has an unusual point spread function due to a lunar eclipse
24
Convolve to find edges The imaging device is modelled by a matrix of values called a mask or kernal. It can be used to emphasise different aspects of the image such as edges, for example.
25
Convolution is a mathematical operation of matrices combining flip, multiply, sum and scan http://www.eas.asu.edu/~karam/2dconvolution/
26
A lens forming a magnified image of a fluorochrome molecule - effectively an infinitely small point which emits light. ? The image at the plane of focus (lower left) is not a point but a disk - called an AIRY DISK - surrounded by a halo, in fact a long series of progressively dimmer haloes. Most of the light is in the disk so we often ignore the haloes. Above and below the focal plane (right) the light fans out in a hollow cone pattern. The complete 3D shape is called the Point Spread Function (PSF). Resolution of independent emitting objects
27
Stage sin = 0.25 sin = 0.75 What spreads out the light into a disc instead of a point is diffraction of light at the edge of the lens. The diameter of the disk depends on the path differences between rays from opposite edges of the lens. So the size of the disk depends on the size of the lens. To be precise, it depends on the proportion of the total light which the lens can collect - so a small, close lens is equivalent to a larger one further away. It therefore defined by angle - the radius of the disk, r, is given by: r = 0.61 / sin where is the half-angle of acceptance of the lens.
28
Rayleigh’s Criterion Rayleigh proposed that we could still tell two Airy disks apart until the first minimum of one was on the centre of the other. So the radius of the Airy disk becomes the definition of resolution. Note that the Airy discs are additive (no interference). r = 0.61 / sin
29
Transmitted light is partially coherent so diffracted waves can interfere Local domains of filament/arc emit light simultaneously and in phase. Waves that are split by diffraction maintain this coherence, thus may interfere if brought back together
30
Transmitted image resolution arises from interference of diffracted waves
31
When a light falls on a small object, or on the edge of a larger one, it is diffracted - scattered in all directions. Each particle will scatter light in all directions. However, if we look at the rays diffracted by each particle at the angle , we notice something rather interesting. Each will be out of step with its neighbour by exactly one wave- length - in other words, in step with its neighbour. Since these reinforce, the specimen as a whole scatters light strongly in this direction - the others cancel out. Clearly, in the diagram, / d = sin . The smaller the spacing, the larger the angle . n = d.sin
32
Two diffraction orders must be collected to resolve objects
33
This leads us to Abbe's formula for the resolution of a microscope in transmission mode. The minimum resolved distance (d) is given by: d = / 2n.sin What this makes plain is that when we are illuminating a specimen from below, how the light hits the specimen is crucial for good resolution! However wonderful and expensive our objective is, if we don't adjust the condenser properly it is wasted.
34
The visible spectrum is about 400 to 750 nm, but only about 500-625nm is bright enough to be useful for microscopy. So our scope for improving resolution by varying the wavelength is limited. The most important factor is to keep sin as large as possible. If = 90° then sin = 1 - this is the theoretical limit and practically inconvenient, but we can get close. Sin = 0.95 corresponds to = 72°, which is achievable, and the 5% difference is insignificant. For this case d = / 2n.sin works out to 290nm for = 550nm But to get maximum this resolution, we need contrast. In both cases, and are what determine resolution
35
Increased contrast in transmission microscopy Dark field-condenser angle greater than objective Phase Contrast Microscopy-cumulative shift of phase over light path Differential Interference Contrast (DIC)-gradient of refractive index at point of focus phase DIC
36
Darkfield The oldest technique for looking at unstained samples using diffracted light. We use a hollow cone of illumination, replacing the condenser diaphragm with a ring-shaped aperture which is large enough for the light to fall outside the acceptance angle of the objective lens. Light scattered up to an angle of 2 can enter the objective, so we are using its full resolution - but only if the condenser has a higher NA.
37
Phase Contrast The Dutch physicist Zernike hit upon the idea of making the diffracted rays half a wavelength out of step with the direct rays, so that anything in the specimen which scatters light will appear dark in the final image.
38
Assuming the refractive index of the cell contents puts the scattered light ~ ¼ behind, we need to slow it by a further ¼ to bring the path difference to ~ ½. Taking = 550nm (the centre of the spectrum), we want the path difference to be ¼, 550/4 which is 137.5nm. The depth t will therefore be twice this = 275nm. How deep must the groove be? For glass n=1.5, for air n=1. Path difference = depth (t) x difference in refractive indices (1.5 - 1), ie 0.5t.
39
water n = 1.3 cytosol n = 1.35 liposome n = 1.45 ray entirely in water – no phase shift ray goes through edge of cytosol – small phase shift ray goes through more cytosol – larger phase shift ray goes through edge of liposome – even larger phase shift ray goes through centre of liposome – largest phase shift
40
The final result - the more material is present the darker objects appear
41
Differential Interference Contrast Polariser Analyser Introduced by the Polish-French scientist Nomarski. Its basic principle is simple - make each ray of light interfere with another passing through the specimen a very small distance away from it. If the refractive index of the specimen is changing, there will be a path difference between the two rays, if it is uniform there won’t be. The contrast we see in the final image will depend on the local rate of change of refractive index in the specimen - hence the name differential interference contrast.
42
water n = 1.3 cytosol n = 1.35 liposome n = 1.45 both rays in water – no phase difference one in water, one cytosol – phase difference one sees more cytosol – small phase difference one in cytosol other in liposome – large phase difference Both in liposome + cytosol – no phase difference
43
one in water, one cytosol – phase difference one sees more cytosol – small phase difference one in cytosol other in liposome – large phase difference
44
The final result - a ‘relief’ effect – which is striking but entirely artificial
45
Bright Field Differential Interference Contrast Dark FieldPhase Contrast
46