Relaxation Exponential time constants T1 T2 T2*
Longitudinal (spin-lattice) Inversion recovery experiment T2 Transverse (spin-spin) relaxation Spin-echo experiment T2* T2 with field inhomogeneities Gradient echo experiment TE = 15 ms TE = 20 ms TE = 30 ms TE = 60 ms Relaxation
2
Exponential decay = 30 ms Relaxation
3
T2 TE = 15 ms TE = 20 ms TE = 30 ms TE = 60 ms = T2= 30 ms
Relaxation
4
Exponential recovery Inversion recovery Saturation recovery Relaxation
5
T1 Inversion recovery T1 = 2000 ms TI = 50 ms 100 ms 400 ms 800 ms
Relaxation
6
Pulse sequence diagrams
Magnetization preparation Excitation Read-out Nrep RF GSS GPE GRO DAC Relaxation
7
Spin Echo EPI pulse sequence (not to scale)
Excitation pulse Refocusing pulse RF GSS … GPE … GRO … DAC Relaxation
8
Spin Echo measures T2, not T2*
Excitation pulse Refocusing pulse Echo Read-out MT T2 T2* S Relaxation
9
Spin Echo EPI pulse sequence
TE/2 TE/2 RF GSS GPE GRO DAC Relaxation
10
Inversion Recovery Spin Echo EPI
TE/2 TE/2 RF GSS GPE GRO DAC Relaxation
11
Inversion Recovery FLASH
NPE RF GSS GPE GRO DAC NRO Relaxation
12
Spin Echo: erasing magnetic field imperfections
Imaging signal comes from protons on water molecules. Frequency map, zoomed in on lateral temporal cortex Hz On resonance Sensitive to macro- and microscopic variations in B0. 100 Hz off resonance 250 Hz off resonance Relaxation
13
Spin Echo: erasing magnetic field imperfections
Summing all spins (e.g. axial slice) creates rapid signal decay Relaxation
14
Spin Echo: erasing magnetic field imperfections
Applying a 180 pulse at TE/2 refocuses the inhomogeneity-induced dephasing at TE t = 0 ms t = TE/2 t = TE Relaxation
15
SE EPI: reduction of through-slice dephasing
Gradient echo Spin echo Relaxation