Quality Control (Level 0) for Tail-Catcher Brian Wood University of Illinois Eric Téllez Universidad Nacional Autonoma De Mexico
2
Extrusion Run 07/08/04 Total amount extruded: 676 strips Accepted Scintillator: 218 strips Yield: 32.2% (twenty plus darkened strips)
3
Measured the Width and Thickness
4
Visual Test on Dopant Presence Ultraviolet Lamp (365 and 254 nm)
5
Evaluated all the permissible strips
6
….continued
7
Width:
8
Thickness:
9
Attenuation Length Measurement Setup
10
Attenuation Length from Random Sampling
11
Number:Batch:L1 (Long) [Excel]: L2 (Short) [Excel]: 1 1 41.3 cm 38.0 cm 101 6 59.9 cm 48.8 cm 121 (dark) 7 69.0 cm 49.3 cm 161 9 64.1 cm 51.4 cm 182 10 31.3 cm 24.1 cm Ref (previous run) - 43.7 cm 24.1 cm
12
Conclusion We made measurements for 218 strips of extruded scintillator during which : We accumulated experience and we currently have a procedure for quality control level 0.