Work Package 2 – Overview of instrumentation, data gathering and calibration issues Lidar Calibration Ewan O’Connor, Anthony Illingworth and Robin Hogan Progress meeting 3 –Reading – 24-25 Oct 2002
2
Calibration technique ’=1/2 k Use Mie theory for spherical droplets. k is constant for cloud droplets (D=8-30 m).
3
Multiple scattering Calculate as a function of range using Eloranta’s model (1998). Need to know laser divergence and telescope FOV.
4
Aerosol k in aerosol variable therefore extinction unknown effect on calibration unknown Current solution: Reject profile for calibration if a > than some predetermined value
5
Validation – UV Raman lidar Calibrate UV Raman lidar with molecular backscatter. To validate – need to calculate atmospheric transmission at both wavelengths and the value of for each instrument.
6
Validation – UV Raman lidar Calibrate UV Raman lidarCompare calibrated CT75K ceilometer
7
Validation – UV Raman lidar Step 1. Calibrate both lidars Step 2. Adjust for differences in atmospheric transmission Step 3. Adjust for differences in multiple scattering
8
Validation in space – LITE
9
CT75K – solar background