Download presentation
Presentation is loading. Please wait.
1
Arizona’s First University. Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis
2
Physicists (University of Arizona) RJ Hamilton CJ Watchman J Gordon NJ Stanley Radiation Oncologist (University of Arizona) B Stea Neurosurgeon (Center for Neurosciences, Tucson, AZ) A Sanan Feasibility of Image-Guided SRS for Trigeminal Neuralgia with Novalis
3
Accuracy of Frame-Based Novalis SRS for TGN Rahimian et al (J. Neurosurg S3 101:351-55, 2004) –Image Fusion 0.4 mm –MR 0.2 mm –Gantry 0.3 mm –Couch 0.6 mm –Overall 0.9 mm (0.6 mm w/o couch) Target position 0.2 mm Laser 0.2 mm Phantom Radiographic Test 0.7 ± 0.1 mm
4
Frame-Based SRS for TGN Human Issues –Pins fix frame to skull –Uncomfortable during treatment delivery too –Patient remains in department for several hours waiting for treatment –Difficult to abort procedure (machine problems) –Scheduling: patient, neurosurgeon, radiation oncologist, physicist, treatment time, etc…
5
Frame-Based SRS for TGN Technical Issues –Frame-based systems require that the position of a patient in the frame is unchanged from application of the pins through treatment completion –The stereotactic frame (not the patient) is aligned with the machine isocenter for treatment –Differential torque on the head ring deforms it –Head ring is bolted to the table assembly, patient may be displaced relative to it.
6
Image-Based SRS for TGN Human Issues –Plastic mask (BrainLab U-PLAST mask fixation system) helps patient remain still –Scheduling: specialists (neurosurgeon, radiation oncologist) may work at their convenience, treatment time set for patient
7
Image-Based SRS for TGN Technical Issues –Positioning accuracy Depends on the accuracy of the registration of images between the planning system (DRRs) and imaging system (ExacTrac) Depends on the calibration of the imaging system with the mechanical isocenter –Variations of the position of the patient in the mask (stereotactic coordinate space) are not important provided that the imaging system can correct for it (ExacTrac X-Ray 6D and Robotics) –Once positioned for treatment, the patient must remain fixed.
8
Feasibility of Image-Based SRS for TGN Demonstrate System Accuracy –Anthropomorphic Phantom Experiment Demonstrate that variations of the patient position in the mask are not important –Anthropomorphic Phantom Experiment Confirm for each patient that deviations of their head within the mask may be corrected by the ExacTrac X-ray 6D and robotics –Perform planning CT scan plus two addition CT scans. The final scan is performed 1 hour prior to the scheduled treatment time. Confirm that the patient does not move during treatment –Repeat the ExacTrac X-ray positioning 3-4 times (every 3 rd couch position) during treatment delivery
9
Demonstrate System Accuracy Anthropomorphic Phantom –0.25” BB embedded in head –BrainLab U-PLAST mask constructed –Axial CT scan (35 cm FOV 1.5mm slice) with BrainLab Head & Neck Localizer –BrainScan Plan –ExacTrac 6D X-Ray and Robotics
10
Anthropomorphic Phantom CT
13
Anthropomorphic Phantom Plan
14
Accuracy of Isocenter Determination Two planners contoured the BB in the planning system and then placed the isocenter at the center. Repeated the procedure 10 times each.
15
Pixel Size Pixel Size = 0.5 mm
16
Pixel Size Pixel Size = 0.5 mm
17
Pixel Size Pixel Size = 0.5 mm
18
Pixel Size Pixel Size = 0.5 mm 0.5 mm
19
ExacTrac Calibration
30
Anthropomorphic Phantom Treatment Head Holder
31
Anthropomorphic Phantom Treatment
32
Select Isocenter
33
Anthropomorphic Phantom Treatment Automatic Preposition
34
Anthropomorphic Phantom Treatment
35
Automatic Preposition Film Verification Difference = ( 0.3, -0.2 ) = 0.4 mm 10 mm cone BB shadow
36
Anthropomorphic Phantom Treatment
40
Film Verification Difference = ( 0.1, 0.1 ) = 0.1 mm 10 mm cone BB shadow
41
Phantom Displacement Experiment Phantom moved by exactly 2.0 cm in all three directions as determined by a ruler Perform X-ray Alignment with ExacTrac Shifts Found by ExacTrac –AP 19.85 mm –Lateral20.05 mm –Vertical20.40 mm
42
Phantom Moved: Not in Mask !!! Head Rotated
43
Phantom Moved: Not in Mask !!! No Head Holder Head Lower and Rotated Head Holder Removed
44
Phantom Moved: Not in Mask !!! Position using standard procedures with array
45
Phantom Moved: Not in Mask !!!
48
Film Verification Difference = ( 0.1, 0.1 ) = 0.1 mm 10 mm cone BB shadow
49
Phantom Moved: Not in Mask !!! Rotation too large for robotics
50
Phantom Moved: Not in Mask !!!
51
Uncorrected Rotations Isocenter position is still fine Uncorrected rotation changes geometry of beams relative to skull and therefore changes the calculated doses
52
Patient Repeat CT Scans Original CT scan includes mask fabrication Second CT scan scheduled between date of original scan and treatment day Third CT scan obtained one hour prior to scheduled treatment time
53
Patient Repeat CT Scans Results The mean observed isocenter displacements from repeat CT localizations in the mask –AP 0.9 mm –Lateral 0.6 mm –Vertical 1.7 mm
54
Patient Motion During Treatment ExacTrac localization was preformed during treatment every three arcs and at the end of the procedure to evaluate the extent of patient movement in the mask during treatment.
55
Patient Motion During Treatment: Results The mean displacements from isocenter –AP 0.3 mm –Lateral 0.5mm –Vertical 0.3 mm The mean rotations were all less than 0.4 degrees.
56
Dosimetric Effects of Isocenter Misplacement Five patients treated to 90 Gy at the isocenter with a 4 mm cone Isocenters of original plans were displaced towards the brainstem by 0.5, 1.0, 1.5, and 2.0 mm (worst case scenarios). Maximum brainstem and CN5 doses computed
57
Isocenter Misplacement Results: CN5
58
Isocenter Misplacement Results: Brainstem
59
Conclusions Accuracy is better than image pixel size Mask reproducibility allows accurate positioning with ExacTrac 6D and Robotics Patients move minimally during treatment. Monitoring the position adds little to the overall treatment time and insures accuracy Patients tolerate the procedure well and have not experienced complications
60
Thank You!
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.