McDaniels – Oct 24, 2008
2
Outline MRI Image uncertainty Point uncertainty ADC uncertainty
3
MRI Image Uncertainty Firbank et al – SNR = 0.655 S/σ air – S is Image Signal (Average Intensity) – σ air is the standard deviation of intensity in air area – 0.655 is due to noise centered about zero in raw data, then skewed positive after made into a magnitude image
4
MRI Image Uncertainty Image J returns raw data histogram centered about zero with positive and negative values. Omit 0.655 factor Analyzed “air” noise for about 13K voxels per image
5
MRI Image Uncertainty
6
Typical values of σ (relative) – b=0, σ ≈ 5-8% – b=520, σ ≈ 6-9% – b=850, σ ≈ 8-12%
7
MRI Image Uncertainty
8
Point Uncertainty Points used for Linear Fit Each Point: – Determined from I n = -ln(I b /I b=o ) – Uncertainty in each point σ n 2 = σ b 2 + σ b=0 2 I n +/- σ n
9
Uncertainty in ADC Values Uncertainty in linear fit from Bevington σ 2 = 1/Δ*Σ1/σ i 2 Δ = Σ 1/σ i 2 * Σx i 2 / σ i 2 -(Σ x i / σ i 2 ) 2 e.g. for σ 2 =σ 3 =.2, x 2 =520, x 3 =850 Δ = (1/.04 +1/.04)*(520 2 /.04+850 2 /.04)- (520/.04+850/.04) 2 = 6.8e7 σ = √(1/6.8e7*(1/.04+1/.04)) = 0.00086 ADC ≈ 0.0015 +/- 0.00086 (57%)
10
ADC uncertainty Patient 5 Minimum +/- 18.4% Maximum +/-56.4% Average +/-33.7%
11
ADC Uncertainty