RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 1 14 – RSH Test Plans Lars Seimetz,

Similar presentations


Presentation on theme: "RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 1 14 – RSH Test Plans Lars Seimetz,"— Presentation transcript:

1 RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 1 14 – RSH Test Plans Lars Seimetz, CAU, Kiel, Germany RSH Packaging Peer Review

2 RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 2 RSH Model Philosophy RSH Prototype „Pathfinder“ RAD EM RAD FM/FS Lab-Level Mockup‘s REB Prototype/EM REB EM

3 RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 3 Test Plans for RSH-Models Test / Procedure Lab- Mockup‘s „Pathfinder“RAD EMRAD FMRAD FS CalibrationVarious Sources at Lab-Level Sufficiently flight-like for Calibration Yes (see G. Reitz Talk for Details) Limited parallel Calibration with FS Full Calibration VibrationVibe TestProtoflight Vibe- Testing Full Qualification Acceptance Level Tbd Thermal/ Thermal-Vac Subsystem Testing No (e.g. softening Point BC430) Full Qualification Thermal Vac Acceptance Level Tbd Screened Parts InvestigationUnscreened but idendical Parts Yes ProcessesDevelop Processes Qualify Processes Qualified Processes MaterialsInvestigationDeterminationSystem-Level Qualification./.

4 RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 4 Testing Lab-Level Tests: –Electro-Optical Tests: Efficiency of Diodes with various sources –Mechanical Tests: Pressure, Tension, Torque –Thermal: Cycling (Gradients +/- 5 K/min @ -85(LN2)..+65 (max.+150°) Tests with LN2 Vibration Testing: Single Axe Sine sweep, Manual Setting possible

5 RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 5 Testing Restriction in CAU Testing Capabilities due to shortage in manpower SwRI offers great support for CAU with Testing Infrastructure and Resources in –Hardware Testing –Vibration Testing –Programmatics (e.g. Scheduling, Documentation) –Screening –Equipping E-Boards (Foto Stephan)

6 RAD – The Radiation Assessment Detector for MSL Seimetz DLR November 30 – December 1, 2006 RSH Packaging Peer Review 6 Calibration Refer to G.Reitz PDR-Talk for Details