EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Sudden jumps in the absolute energy scale and their origin.

Similar presentations


Presentation on theme: "EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Sudden jumps in the absolute energy scale and their origin."— Presentation transcript:

1 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Sudden jumps in the absolute energy scale and their origin

2 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale FF Mn-K Pos Quadrant 0 ± 5 eV

3 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Effect of the particle background

4 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale FF Mn-K Pos Quadrant 0 ± 5 eV 2005 Dec 20 – 2006 Jan 01 + 41.5 eV

5 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale FF AL-K Pos Quadrant 0 ± 5 eV 2005 Dec 20 – 2006 Jan 01 + 10.5 eV

6 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Al-K: 1.486 keV, +10.5 eV  +0.707% Mn-K: 5.894 keV, +41.5 eV  +0.704%  Change in the gain, not in the offset

7 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Energy resolution somewhat improved during high background  reduced charge loss noise  reduced charge loss  increased trap saturation Energy shift occurs also for first singles  precursors ‘invisible’  precursors below lower threshold or in preceding frame(s) low background high background

8 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale rev 711, high background Al-K Mn-K Al-K Mn-K The CTI during low and high background rev 709, low background  decreased CTI during high background (Mn-K)

9 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale low background

10 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale high background

11 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale  During episodes of high particle background the CTI is reduced.  This is most likely caused by saturation of traps due to precursors.  The charge of the precursors is below the 20 adu threshold and thus not available.  The usual CTI correction cannot be used.  But there may be a correlation with the background flux which can be used for the correction..

12 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Effect of the quadrant box temperature

13 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Gain/temperature correlation: earlier investigations

14 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α & Al-K position and quadrant box temperature

15 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Quadrant box temperatures (yellow circles: no FF mode)

16 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. temperature: CCD 1 & 2

17 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. temperature: CCD 3 & 4

18 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. temperature: CCD 5 & 6

19 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. temperature: CCD 7 & 8

20 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. temperature: CCD 9 & 10

21 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. temperature: CCD 11 & 12

22 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. Al-K position

23 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Deviations of the Al-K and Mn-K α positions

24 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α position vs. Al-K position

25 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α positions (after gain and CTI correction)

26 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mn-K α positions after temperature (and dCTI/dt) correction

27 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale “CCD offsets”

28 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Doy 2000 Quadrant box temperatures

29 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale

30 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale

31 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Standard correction

32 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale with 1 adu / 2000 d drop

33 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale with 1 adu / 2000 d drop

34 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Mean quadrant box temperature [C] (F1576..F1876) Mn-K α position [adu] (in quadrant 0, after 1 adu / 2000 d drop) slope: +0.43 adu / C

35 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale standard correction

36 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale after temperature.. correction

37 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale standard correction

38 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale after temperature.. correction

39 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Multiply all energies with a factor F: F = ( 1 - A(t) - B(T q ) ) C A = 4.24 * 10 -7 (t-t 0 )/[d] B = 3.65 * 10 -4 T q /[ o C] C = 1.0057 t 0 : 2000-Jan-01 T q : mean of T(F1576)..T(F1876)

40 EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale


Download ppt "EPIC Calibration Meeting, Mallorca K. Dennerl, 2006 October 26 EPIC pn: Jumps in the energy scale Sudden jumps in the absolute energy scale and their origin."