X-RAY MEASUREMENT AND DOSIMETRY Salman Albeshan
2
EXPOSURE 2 It is the quantity that expresses the concentration of radiation delivered to a specific point. Old unit: ROENTGEN(R). SI unit: COULOMB/Kg of air. 1R=2.58x10(-4) 1C/Kg=3876R
3
ABSORBED DOSE. It is the quantity that expresses the concentration of radiation energy at a specific point within the body tissue..OLD UNIT : RAD.SI UNIT : GRAY 1Gy=100Rad 1Rad=1/100Gy 3
4
SEVER. This is the unit used only in radiation protection and it is equal to gray multiply By quality factor (for x-ray=1) 1SEVER=100REM 4 INTEGRAL DOSE.It is the total amount of energy absorbed in the body..UNIT:GRAM –RAD.
5
EXPOSUR FACTORS 1.Kilo-voltage. 2.Milliamperes. 3.Secoeds. 4.FFD. 5.Focal spot size. 6.Type of film. 7.Grid. 5
6
KILO-VOLTAG.The technicians have four matter in mind before selecting the KVp value: *Penetration,image contrast, image un-sharpness and radiation protection. 6 MILLIAMPERE-SECOND (MAS).It control the image density.
7
FOCUS FILM DISTANCE(FFD).The FFD used for most of examination is approximatly100cm..Short distance increase the intensity..In the lateral projection of cervical spine and chest the FFD is 180cm. 7
8
GRID. It is a device, which is used to enhance the image contrast by Removing the scatter. Is made up of lead strips and It is Placed between the patient and image receptor. *Types of grid: 1.Linear. 2.Parallel. 3.Focused. 4.Cross. 5.Moving. 8
9
Conventional X-ray Imaging. X-ray Production. Electrons from cathode filament are accelerated towards and impact the rotating anode. Rapid deceleration produces heat (~ 99%) and x-rays (~1%) High Electrical Potential Electrons - + Exposure Recording Device Radiation Penetrate the Sample