Knee Arthroplasty
2
Degeneration of Knee
3
Degeneration of Knee (cont’d) Osteoarthritis is the most common cause Abnormalities of knee joint function resulting from –Fractures –Torn cartilages –Torn ligaments can lead to degeneration many years after the injury
4
Total Knee Arthroplasty Indications for surgery – Pain and disability at the point When ADL (standing, walking, and climbing stairs) cannot be done It is an artificial joint –Resurfacing of cartilage and underlying bone –A metal and plastic implant –Corrects deformity
5
Appearance of TKA
6
Objectives of TKA Function –Stability –Motion Long-term fixation of implants Correction of deformity –reduce wear
7
History of TKA 1863-1921 –Interposition of joint capsule (Verneuil) –Muscle (Ollier) –Fat and fascia (Murphy) –Pig bladder (Campbell) 1951-1958 –Acrylic hinge (Walldius) –Vitallium femoral hemiarthroplasty (Campbell) –Acrylic two-part prosthesis, first TKA (Judet) –Metallic tibial hemiarthroplasty (Townley) –Metallic hinge (Shiers) –Tibial unicompartmental designs (McKeever and MacIntosh) 1969 Polycentric TKA (Gunston) 1970 Bicruciate sacrificing arthroplasty (Freeman and Swanson) 1971 Improved, refined hinge (Guepar)
8
History of TKA 1972-1974 –Polyethylene metal bicondylar anatomic TKA (Townley) –Congruent geometric design (Coventry) –Unicondylar, unicompartmental arthroplasty (Boston and Brigham) –Total Condylar cruciate sacrificing tricompartmental TKA (Insall and Ranawat) 1975-1978 –Bicruciate retaining metal- backed tibial TKA (Cloutier) –Varus-valgus/ anteroposterior constraint TKA (Walker) –Posterior stabilized TKA (Insall and Burstein) 1980’s –Low Contact Stress, ACL sparing
9
Developments in TKA Design Early designs failed –Loosening, wear, osteolysis, stiffness, dislocation, instability, and extensor mechanism dysfunction In 1970’s –~ 300 TKA designs To provide rotation –Mobile-bearing implants in the 80’s To reduce wear –Poly concave design
10
Poly design Congruent femorotibial articulation –A larger area of contact –Reduces contact stresses High contact Stresses for “Curved-on-Flat” design LCS low contact stress distributed over a large area of polyethylene
11
Poly design (cont’d) Sphericity leads to congruency in coronal and sagittal plane –Reducing this mode of wear –Mobility of tibial bearing reduces Rotational torque Subsequent loosening of tibial component
12
Design criteria Material compatibility and wear Adequate mechanical strength Minimization of joint reaction forces Minimization of fixation interface shear Avoidance of fixation interface tension\
13
Design criteria (cont’d) Uniformity of interface compression Duplication of anatomical function Adequate fit for patient population Manufacturability Reasonable inventory costs
14
Implant Wear Level and type of stresses –On articulating surfaces Material properties –Imperfections of UHMWPE Coefficient of friction UHMWPE - ultra-high molecular weight polyethylene
15
Stresses on Implants Load –Peak tibiofemoral force during sport activities Around 7 times the body weight Contact stresses on UHMWPE –~ 3 times yield point
16
Stresses on Implants (cont’d) Plastic strain of multiple cycles –Material fatigue –Pits, cracks, and delamination –Flake-like wear particles in surrounding tissue Wear particle Polyethylene failure
17
Material properties UHMWPE fails first Wear resistance –Ultimate tensile strength and ductility Inversely proportional –Increase in ultimate tensile strength Reduction in toughness Increase wear rates Balance the two to increase wear resistance
18
Material properties (cont’d) Wear, crack nucleation, occurs due to –Fusion defects –Voids Quality of resins Manufacturing processes –Cyclic plastic deformation
19
Coefficient of Friction Coefficient of friction depends on –Material –Surface finish of articulating surfaces –Lubricating regimen Surface roughness can increase in vivo –Entrapment of third body particles Bone or bone cement Amount of wear particles can be reduced –If full-fluid lubrication is used
20
Knee joint components Knee joint implants consist of –Femoral –Tibial –Patellar component
21
Femoral Component Made of a strong polished metal –Cobalt chrome Radius –Single –Reduced
22
Femoral Component (cont’d) Single radius design has same femoral radius from extension to full flexion Reduced radius design has larger radius near extension and smaller radius at flexion
23
Tibial Component Proximal tibia is covered with a metal tray Tibial component is topped with –Disk-shaped polyethylene insert May be fixed Rotates about a stem in rotating platform
24
Patellar Component Replaces knee cap
25
Types of TKA Condylar TKA –Constrained PCL sacrificed –Non-Constrained Mobile TKA –May spare the ACL Uni Hinged
26
Differences between Condylar and Mobile TKA Condylar TKAMobile TKA Tibial component is topped with a flat metal piece that securely holds the polyethylene insert The polyethylene insert can rotate slightly, which gives knee implant a more natural interface between the surface of femoral component and polyethylene When the knee is in motion, the femoral component glides over the polyethylene Both gliding and rotating motion can be done Natural shape of condyle is not a perfect circle: Different amounts of stress is exerted on the insert at different points. Since PE can rotate, contact with FC is more even, less stresses and wear
27
Advantages of Mobile TKA Many components of mobile-bearing knee are same as traditional fixed knee implants –Same proven surgical procedures can be used Currently used preoperative and postoperative routines for patient are also same
28
Disadvantages of TKA Particles polyethylene wear –Lead to aseptic loosening and osteolysis –Destroys a tibial inlay in <10 years Unexplained pain –Infection Reduced flexion
29
Surgical Procedure An incision is made over the front of the knee and tibia Femoral condyles are exposed Bone cuts are made to fit the femoral component
30
Femoral IM Canal A reamer is passed through a hole near the center of joint surface of lower end of femur and into femur shaft
31
Cutting the Distal Femur A resection guide is attached to lower end of the femur 8-10 mm Osteo- cartilage surface is removed
32
Cutting the Distal Femur (cont’d) Another resection guide is anchored to end of femur Pieces of femur are cut off the front and back –As directed by the miter slots in guide Then cuts are made to bevel the end of femur to fit implant
33
Cutting the Distal Femur (cont’d)
34
Placing the Femoral Component Metal component is held in place by friction In the cemented variety –An epoxy cement is used
35
Cutting the Tibial Bone A resection guide is attached to front of tibia Direction of the saw cuts in 3D –AP tilt –LM tilt Upper end of tibia is resected
36
Cutting the Tibial Bone (cont’d)
37
Placing the Tibial Component Metal tray that will hold plastic spacer is attached to the top of the tibia
38
Placing the Plastic spacer Attached to the metal tray of tibial component
39
Preparing the Patella The undersurface of the patella is removed
40
Placing the Patella Component The patella button is usually cemented into place behind the patella
41
Completed Knee Replacement
42
X-Ray of Completed Knee
43
Animation for TKA http://www.hipandkneesurgery.net/knee.ht ml http://www.hipandkneesurgery.net/knee.ht ml
44
Unicompartmental KA Unlike total knee surgery this is –Less invasive procedure –Replaces only damaged or arthritic parts i.e. in either compartments
45
Advantages of Uni Preservation of the ACL Smaller incision Less blood loss Lower morbidity Shorter recovery time Lesser bone removed
46
Disadvantages of Uni Inferior survivorship –Error in proper placement of components –Loosening –Prosthetic wear –Secondary degeneration of opposite compartment
47
Animation for Unicompartmental KA http://www.hipandkneesurgery.net/repicci.h tml http://www.hipandkneesurgery.net/repicci.h tml
48
Modifications in TKA Design The New Jersey LCS Knee allows –Bicruciate or posterior cruciate ligament (PCL) retention Using gliding meniscal bearings or cruciate substitution with rotating platform design Also provides –Uni mobile-bearing –Mobile- bearing stemmed design
49
References http://www.orthobluejournal.com/supp/020 2/sorrells/ http://www.orthobluejournal.com/supp/020 2/sorrells/ http://www.orthobluejournal.com/supp/020 2/crossett/Default.asp http://www.orthobluejournal.com/supp/020 2/crossett/Default.asp http://www.orthobluejournal.com/supp/020 2/kuster/ http://www.orthobluejournal.com/supp/020 2/kuster/