TOF calibration Y. Karadzhov. Y. Karadzhov, MICE CM26 24 March Slide 2 TOF Calibration - Trigger TOF1 TOF 0 TOF 1 TOF 2 The blue line encloses the calibrated.

Similar presentations


Presentation on theme: "TOF calibration Y. Karadzhov. Y. Karadzhov, MICE CM26 24 March Slide 2 TOF Calibration - Trigger TOF1 TOF 0 TOF 1 TOF 2 The blue line encloses the calibrated."— Presentation transcript:

1 TOF calibration Y. Karadzhov

2 Y. Karadzhov, MICE CM26 24 March Slide 2 TOF Calibration - Trigger TOF1 TOF 0 TOF 1 TOF 2 The blue line encloses the calibrated area in the TOF stations.

3 Y. Karadzhov, MICE CM26 24 March Slide 3 TOF Calibration - Trigger TOF1 Time resolution after the calibration : TOF0 – 51ps; TOF1 – 62ps; TOF2 – 52ps. Why the resolutions of TOF1 and TOF2 are different since the two detectors are identical?

4 Y. Karadzhov, MICE CM26 24 March Slide 4 TOF Calibration - Trigger TOF0 Time resolution after the calibration: TOF0 – 50ps; TOF1 – 63ps; TOF2 – 51ps.

5 Y. Karadzhov, MICE CM26 24 March Slide 5 Data quality February runstatustrigger 01467good1 01468bad1 01469good1 01470good1 01472good0 01474good0 01476good1 01477good1 01478good1 01479good1 01480bad1 01484good0 01485good0 01486bad1 01487good1 01488good1 01489bad1 March runstatustrigger 01507bad1 01513bad1 01514bad1 01515bad1 01516bad1 01417good1 01518bad1 We still suffer from the loss of synchronization between the crates of the DAQ. This problem is visible only in TOF2 and KL.

6 Y. Karadzhov, MICE CM26 24 March Slide 6 Conclusions 1.Calibration for trigger in TOF0 and TOF1 is available. 2.Why TOF1 resolution is worse? Puzzle that has to be solved. 3.Data quality is not good.


Download ppt "TOF calibration Y. Karadzhov. Y. Karadzhov, MICE CM26 24 March Slide 2 TOF Calibration - Trigger TOF1 TOF 0 TOF 1 TOF 2 The blue line encloses the calibrated."