MINERvA Calorimetry Calibration with infinite detector Jaewon Park University of Rochester Jupiter/MINERvA Group Meeting, Nov 15, 2006.

Similar presentations


Presentation on theme: "MINERvA Calorimetry Calibration with infinite detector Jaewon Park University of Rochester Jupiter/MINERvA Group Meeting, Nov 15, 2006."— Presentation transcript:

1 MINERvA Calorimetry Calibration with infinite detector Jaewon Park University of Rochester Jupiter/MINERvA Group Meeting, Nov 15, 2006

2 2 Infinite ID gap dependence Angle: 0deg (perpendicular to detector) Gap difference gives ~1% contribution

3 3 Infinite DS-HCAL No gaps Proton beam energy: 1, 1.5, … 5GeV Angle: 0, 10, … 60deg  0deg means perpendicular to DS-HCAL

4 4 Infinite DS-HCAL

5 5 Infinite OD With gaps Iron width: 3, 5, 7, 10cm Proton beam energy: 1, 1.5, … 5GeV Angle: 30, 40, … 80deg  90deg means perpendicular to OD

6 6 Infinite OD (3cm iron)

7 7 Infinite OD (5cm iron)

8 8 Infinite OD (7cm iron)

9 9 Infinite OD (10cm iron)

10 10 Infinite OD (summary)

11 11 Summary Calibration constants are calculated for ID, DS- HCAL, OD ID  Gap dependence  Average over beam energy DS-HCAL  Average over beam energy and angle(0 to 50deg) OD  Iron width dependence  Average over beam energy and angle(30 to 80deg) Next step is to test those calibration constants with composite detectors (ID+DSHCAL and ID+OD)

12 12 ID + infinite DS-HCAL No gaps in ID and DS-HCAL Angle: 0deg (perpendicular to detector) Proton beam energy: 1, 1.5, … 5GeV ID track length: 476, 376, 276, 176, 76, 26cm ~2% deviation from beam energy

13 13 ID + infinite OD With gaps in ID and OD 4 types of ID+OD: iron width=3, 5, 7, 10cm Proton beam energy: 1, 1.5, … 5GeV Angle = 50deg  This is middle point of previous calibration(30 to 80deg) ID track length: -100, -80, -40, 0, 40, 80, 100cm from center

14 14 ID+Infinite OD Iron width: 3cm Iron width: 5cm

15 15 ID+Infinite OD Iron width: 7cm Iron width: 10cm

16 16 ID+Infinite OD (summary) Calibration doesn’t work very well ~5-10% lesser than beam energy

17 17 Calibration without back-scattered particles When ID track length=0, it should be 1  Back-scattered particle can’t contribute DS-HCAL for composite detector When ID track length is large, it’s ID only geometry Two end point should be 1 With new calibration for DS-HCAL

18 18 Calibration without back-scattered particles It doesn’t help very much for OD With new calibration for OD

19 19 Next try I don’t want change ID calibration Relative OD calibration doesn’t help It would be best to make flat in the central region and to scale (ID+OD) 5cm OD

20 20 Another try First of all, I used different OD calibrations for 5, 7, 10cm OD to make close to 3cm OD curve Then, scale (ID+OD) to make it near 1  Scale factor was 1.03  I hope to use same ID calibration for DS-HCAL and OD

21 21 Summary Individual calibration didn’t work very well for ID+OD. It seems best to calibrating with combined detector which requires constant total energy over different ID track length. Next step is leakage study


Download ppt "MINERvA Calorimetry Calibration with infinite detector Jaewon Park University of Rochester Jupiter/MINERvA Group Meeting, Nov 15, 2006."