29.1.2003Sensor meetingA. Furgeri 1 Status report IEKP – University of Karlsruhe Alexander Furgeri
2
29.1.2003Sensor meetingA. Furgeri 2 QTC Optical inspection of (16/40) –KA-Set 9 (2.) W5B (5/5) –KA-Set 10/1 W5B (3/8) –KA-Set 11/1 W6B (4/10) –KA-Set 11/2 W6B (4/4) Electrical measurements of KA-Set 9 (2.) IV & CV of KA-Sets 10 to 11 (18/35) Summary IQC
3
29.1.2003Sensor meetingA. Furgeri 3 Optical inspection KA-S9 Batch 32965 2. Delivery 26.11.2002 W5B KA-S10/1 Batch 31850 W5B (3/8) 9.1.2003 KA-S11/1 Batch 32961 W6B (4/10) 9.1.2003 KA-S11/2 Batch 39728 W6B 9.1.2003
4
29.1.2003Sensor meetingA. Furgeri 4 Impressions of opt. inspection IV&CV OK ! ?
5
29.1.2003Sensor meetingA. Furgeri 5 Summary of optical inspection Scratches ! On surface and on backplane ! – many at the same region of the sensors (handling facility ?) – some are really deep ! Antistatic sheets to small for W5B No -stamps on backplane No sensor failed
6
29.1.2003Sensor meetingA. Furgeri 6 IV-curves of KA-Set 9 (2.) KA-S9 Batch 32965 2. Delivery 26.11.2003 No failures !
7
29.1.2003Sensor meetingA. Furgeri 7 CV-curves of KA-S9 KA-S9 Batch 32965 2. Delivery 26.11.2003
8
29.1.2003Sensor meetingA. Furgeri 8 Strips of KA-S9 No failures !
9
29.1.2003Sensor meetingA. Furgeri 9 IV-Curves of KA-Set 10/2 2 failures ! W5B Delivered 9.1.2003
10
29.1.2003Sensor meetingA. Furgeri 10 CV-curves of KA-Set 10/2
11
29.1.2003Sensor meetingA. Furgeri 11 IV-curves of KA-Set 11/1 No failures ! W6B Delivered 9.1.2003
12
29.1.2003Sensor meetingA. Furgeri 12 CV-curves of KA-Set 11/1
13
29.1.2003Sensor meetingA. Furgeri 13 IV-curves of KA-Set 11/2 1 failure ! W5B Delivered 9.1.2003
14
29.1.2003Sensor meetingA. Furgeri 14 CV-curves of KA-Set 11/2
15
29.1.2003Sensor meetingA. Furgeri 15 Summary of QTC KA-S9/2KA-S10/1KA-S10/2KA-S11/1KA-S11/2 Optical inspected 5/53/80/134/104/4 measured5/50/8 10/13 Only IV/CV 4/10 Only IV/CV 4/4 Only IV/CV Failures Opt/IV/ strips 0/0/00/0/00/0/00/0/00/2/00/2/00/0/00/0/00/1/00/1/0 Accepted %100 8410075
16
29.1.2003Sensor meetingA. Furgeri 16 IQC Status of cyclotron: –Mechanical part ready –Electrical part problematically Old components (1962) Many improvements (automatic temperature control, flux measurement ) Comparison of Cint-ramps between Louvain and Karlsruhe Statement of main technician: „Ask in 2 weeks“
17
29.1.2003Sensor meetingA. Furgeri 17 Cint-bumps Effect only on irradiated ST-mini´s seen Caused by low Rint ( that´s the problem) 1 mini sent to Louvain 3 mini´s sent from Louvain to us, but storage over christmas (3 weeks) at unknown temperature
18
29.1.2003Sensor meetingA. Furgeri 18 Comparison KA-Louvain KA-mini irradiated with p Karlsruhe Louvain Same behavior! But different values !? Strip 92
19
29.1.2003Sensor meetingA. Furgeri 19 Cint ramps 1 8 1 8 Cint Dissipation
20
29.1.2003Sensor meetingA. Furgeri 20 Rint ramps 1 8 Maximum of Cint corresponds to Minimum of Rint
21
29.1.2003Sensor meetingA. Furgeri 21 Ileakage ramps Profil from 400 V Udep
22
29.1.2003Sensor meetingA. Furgeri 22 Cint ramps (2) 17 27 In middle of sensor „constant“
23
29.1.2003Sensor meetingA. Furgeri 23 Rint ramps (2) 1 27 In middle of sensor „constant“ Lowest value ~60MOhm
24
29.1.2003Sensor meetingA. Furgeri 24 Ileak ramps (2) Ileak ramps (2) -12°C -10°C -11°C Ileak-bumps are no effect of changing temperature during measurement
25
29.1.2003Sensor meetingA. Furgeri 25 Summary of bumps No temperature effects Still in specs @400V „Calibration“ between KA and Louvain –Same behaviour –Different absolute values (due to LCR-mode ?)
26
29.1.2003Sensor meetingA. Furgeri 26 Outlook Comparison between Cp-D mode and Cp- Rp mode in KA Measurement of Louvain irrad. Mini in KA –Comparison of Rint ! –Comparison of Rpoly increase (see Eric) Perhaps irradiations ?