Refining Corneal Trephination and Donor Preparation for Artificial Cornea Surgery World Cornea Congress VII April 15 - 17 2015, San Diego 2015 Behrooz Azizi, MD; Ronan Conlon, MD; Salina Teja. MD; Joshua C Teichman, MD, MPH; Season Yeung, MBBS, FRANZCO; Setareh Ziai, MD, FRCSC; Kashif Baig, MD, MBA, FRCSC University of Ottawa Eye Institute, Ottawa, Ontario, Canada
2
Financial Disclosures Drs. Azizi, Conlon, Teja, Yeung, Teichman, & Ziai have no financial relationships to disclose Dr. Baig has the following relationships: Financial Interest/ AffiliationName of Company(s) Grant/Research Support Allergan, Bausch & Lomb, Merck, Moria Employment/Honoraria/Consulting Fees/Travel Expenses Alcon, Allergan, Bausch & Lomb, Labtician Major Stock Shareholder None Member: Advisory Panel, Standing Committee, Board of Directors Allergan, Bausch & Lomb Other Financial or Material Interest None
3
Introduction Corneal transplantation is a standard surgical treatment for many corneal diseases However, there are situations in which replacement of the diseased cornea is best managed with an “artificial cornea” The Boston keratoprosthesis (Kpro) is the most commonly utilized artificial cornea
4
Difficulty in achieving concentric punches Khalifa YM, Moshirfar M. Improved centration of the type 1 Boston Keratoprosthesis in donor carrier tissue. Clin Ophthalmol. 2010 Aug 19;4:931-3
5
Difficulty in achieving concentric punches Khalifa YM, Moshirfar M. Improved centration of the type 1 Boston Keratoprosthesis in donor carrier tissue. Clin Ophthalmol. 2010 Aug 19;4:931-3
6
Difficulty in achieving concentric punches John T. Use of trypan blue corneal staining and centration ring in Boston keratoprosthesis surgery. Techniques in Ophthalmology. 2009 Dec. 7(4):127-130
7
Purpose To devise a simple and cost-effective device for the preparation of the donor corneal tissue for Kpro surgery. This device will perform the outer 8mm and inner 3mm cuts with increased accuracy and precision to improve centration.
8
Methods 30 donor corneas that were not suitable for transplantation 3 arms (1:1:1 design) 3mm followed by 8mm punch 8mm followed by 3mm punch Paired 3mm & 8mm punches utilizing the prototype Computer software (Adobe Photoshop, San Jose, CA) was used to measure the centration of the punches by a masked examiner Descriptive & inferential statistics were performed
9
Methods
10
Results – Decentration Error bars = ± SEM * P<0.0001
11
Results – Decentration
12
Conclusions Our study demonstrates that pairing the two trephinations on the donor cornea with the use of a single device leads to increased predictability & ease of trephination