A Computer Aided Detection System For Mammograms Based on Asymmetry and Feature Extraction Techniques By Mohammed Jirari Benidorm, Spain Sept 9th, 2005
2
Why This Project? Breast Cancer is the most common cancer and is the second leading cause of cancer deaths Mammographic screening reduces the mortality of breast cancer But, mammography has low positive predictive value PPV (only 35% have malignancies) Goal of Computer Aided Detection (CAD) is to provide a second reading, hence reducing the false positive rate
3
Basic Components of the System Mammogram Normalization Mammogram Registration Mammogram Subtraction Feature Extraction –Morphological Closing –Morphological Opening –Size Test –Border Test ROC Analysis
4
What is a Mammogram? A Mammogram is an x-ray image of the breast. Mammography is the procedure used to generate a mammogram The equipment used to obtain a mammogram, however, is very different from that used to perform an x-ray of chest or bones
5
Mammograms (cont.) In order to get a good image, the breast must be flattened or compressed In a standard examination, two images of each breast are taken: one from the top (CC) and one from the side (MLO)
6
Mammogram Examples Mammogram of a left breast, cranio-caudal (from the top) view Mammogram of a left breast, medio- lateral oblique (from the side) view
7
Purpose of CAD Mammography is the most reliable method in early detection of breast cancer But, due to the high number of mammograms to be read, the accuracy rate tends to decrease Double reading of mammograms has been proven to increase the accuracy, but at high cost CAD can assist the medical staff to achieve high efficiency and effectiveness The physician/radiologist makes the call not CAD
8
Proposed Method The proposed method will assist the physician by providing a second opinion on reading the mammogram, by pointing out area(s) that are different between the right and left breasts If the two readings are similar, no more work is to be done If they are different, the radiologist will take a second look to make the final diagnosis
9
Data Used The dataset used is the Mammographic Image Analysis Society (MIAS) MINIMIAS database containing Medio- Lateral Oblique (MLO) views for each breast for 161 patients for a total of 322 images Each image is: 1024 pixels X 1024 pixels
10
Normalization The images were corrected/normalized to avoid differences in brightness between the right and left mammograms
11
Mammogram Registration Thermodynamic concepts are used Match a model M with a scene S (M must be deformed to resemble S as much as possible) Use diffusion process technique as follows:
12
Mammogram Registration (cont.) 1. Select pixels to be demons 2. For each demon, store displacement then apply Gaussian filter 3. Use trilinear interpolation to estimate intermediate intensities 4. The demon force is given by optical flow
13
Registration Example Mammogram of left breast Mammogram of right breast
14
Registration Example (cont.) Registered images Grid of displacement
15
Mammogram Subtraction Simple linear subtraction is used Flipped right – left Most common gray level is 0 Masses in right breast are in lower gray level region of subtraction image histogram, while left breast masses are in the higher gray level region
16
Mammogram Subtraction Example Flipped right breastLeft breast showing mass
17
Mammogram Subtraction Example (cont.) Subtraction imageSuperimposed subtraction image
18
Feature Extraction Many features are not masses Morphological filtering using a 3X3 kernel Size test (100 pixels) Border test for border misalignment
19
Avg. # of areas after each stage of the detection process Stage in detection processAvg. # of detected areas After subtraction13.65 After morphological filtering7.80 After size test5.42 After border test2.17
20
Results 102 registered pairs of mammograms used Verified by expert radiologists Recognition %93% False positive1.26 TPF0.9605 FPF0.0962 Az0.95
21
ROC curve showing Az=0.95
22
Future work Use more features like brightness and directionality Try and reduce False Negatives on the basis of region characteristics size, difference in homogeneity and entropy Use larger database that contains both MLO and CC to train/learn, since most commercial CADs use hundreds of thousands of mammograms to try and recognize foreign samples
23
Thank you
24
Questions