6890 Flame Photometric Detector Introduction and Theory August8, 2003
2
6890 FPD June 19, 2003 Agilent Restricted Page 2 6890 FPD - Top Level Assembly AUX 2 Temp Heater Back DET Temp Heater Vent Tube Photo-Multiplier Tube (PMT)
3
6890 FPD June 19, 2003 Agilent Restricted Page 3 6890 FPD Installation AUX 2 Temp Heater Connection Back DET Temp Heater Connection Vent Tube Photo-Multiplier Tube (PMT) Ignitor Connection High Voltage Connection Signal Connection
4
6890 FPD June 19, 2003 Agilent Restricted Page 4 6890 FLAME PHOTOMETRIC DETECTOR Flame Photometric Product No. : G2333A/G2334A or HP 6890 option 240/241 Single and Dual Signal options. EPC Only. Same design as the 5890 FPD. (Most part numbers the same) Phosphorus and Sulfur filters available.
5
6890 FPD June 19, 2003 Agilent Restricted Page 5 6890 FLAME PHOTOMETRIC DETECTOR Installation: Can be retrofitted on older instruments but may require top cover modification. PMT mounted left to right Single FPD - Fitted in back position (ease of access to front) Dual FPD - Fitted in back : Uses both detector PCB slots : Cannot use ALS in back position. Uses both DETB and AUX2 heated zones to heat transfer line and heated block, but only DET temp entered. Requires AUX actuator solenoid bracket and wiring harness. Requires 2.1x 6890 firmware and 4.02 Chemstation (single) 3.0x 6890 firmware and 5.01 Chemstation (dual)
6
6890 FPD June 19, 2003 Agilent Restricted Page 6 6890 FLAME PHOTOMETRIC DETECTOR
7
6890 FPD June 19, 2003 Agilent Restricted Page 7 6890 FLAME PHOTOMETRIC DETECTOR Dual Wavelength option PMT 2 PMT 1 Phosphorus Filter Sulfur Filter Air/H 2 Lines Jet
8
6890 FPD June 19, 2003 Agilent Restricted Page 8 6890 FLAME PHOTOMETRIC DETECTOR Dual Wavelength option
9
6890 FPD June 19, 2003 Agilent Restricted Page 9 6890 FLAME PHOTOMETRIC DETECTOR
10
6890 FPD June 19, 2003 Agilent Restricted Page 10 6890 FLAME PHOTOMETRIC DETECTOR
11
6890 FPD June 19, 2003 Agilent Restricted Page 11 6890 FLAME PHOTOMETRIC DETECTOR
12
6890 FPD June 19, 2003 Agilent Restricted Page 12 6890 FLAME PHOTOMETRIC DETECTOR Photo Multiplier Tube Theory Electron Multiplication by the Dynodes Typical Gain 100,000 - 1,000,000 Photo-emissive Cathode Anode/Collector Dynodes Photon of Light e - 10 6 Secondary electrons 850 VDC e - Photo-electron DC Voltage Signal out to electrometer
13
6890 FPD June 19, 2003 Agilent Restricted Page 13 6890 FLAME PHOTOMETRIC DETECTOR
14
6890 FPD June 19, 2003 Agilent Restricted Page 14 6890 FLAME PHOTOMETRIC DETECTOR
15
6890 FPD June 19, 2003 Agilent Restricted Page 15 6890 FLAME PHOTOMETRIC DETECTOR Unique featuresDual Flame PPC H2 / air Dual PMTPulsed FPD Dual PMT EPC Also PFPD EPCHP GCone or two FPDs Dual PMT Price$11,595$7245$7400