1/13/2004USCMS Tracker Meeting Sergey Korjenevski U of R Status Report IV curve and “second kink” QA stand status - R poly measurement has been debugged.

Similar presentations


Presentation on theme: "1/13/2004USCMS Tracker Meeting Sergey Korjenevski U of R Status Report IV curve and “second kink” QA stand status - R poly measurement has been debugged."— Presentation transcript:

1 1/13/2004USCMS Tracker Meeting Sergey Korjenevski U of R Status Report IV curve and “second kink” QA stand status - R poly measurement has been debugged

2 1/13/2004USCMS Tracker Meeting Sergey Korjenevski IV curve IV curve shows integrated leakage current for all strips If strips working properly the IV curve reaches plateau, stays constant until breakdown Ideally total leakage current would be expected to be around 500 nA for ST OB2 In reality IV curve always goes up due to early breakdown some “leaky” strips

3 1/13/2004USCMS Tracker Meeting Sergey Korjenevski IV Curve

4 1/13/2004USCMS Tracker Meeting Sergey Korjenevski Strip Leakage Profile

5 1/13/2004USCMS Tracker Meeting Sergey Korjenevski “Second Kink” Abnormal change

6 1/13/2004USCMS Tracker Meeting Sergey Korjenevski Together with strips

7 1/13/2004USCMS Tracker Meeting Sergey Korjenevski “Leaky” strips break together Good Strip

8 1/13/2004USCMS Tracker Meeting Sergey Korjenevski Second Kink Voltages 30210431830707 NO 30210431830713 NO weak 200V30210431830715 weak 200V weak 190V30210431830718 weak 190V 30210431830801 obvious 185V 30210431830802 NO (vacuum) 30210431830803 obvious 190V 30210431830805 obvious 195V 30210431830808 obvious 190V 30210431830809 obvious 200V 30210431830810 obvious 160V 30210431830811 NO 30210431830812 obvious 175V weak 185V30210431830813 weak 185V weak 175V30210431830816 weak 175V 30210431830819 BAD strip 130 weak 185V30210431830822 weak 185V Out of 17 4 good; 5 sort of OK; 7 have 2 nd kink; one BAD

9 1/13/2004USCMS Tracker Meeting Sergey Korjenevski “Second Kink” Second kink indicates partial breakdown. Contributing strips are not necessary qualified as bad strips though probably are not reliable at all. If bad “leaky” strip definition would be strip current > 5 nA (currently > 100 nA), ALL ST sensors will fail.

10 1/13/2004USCMS Tracker Meeting Sergey Korjenevski Status report on QA stand Noise while testing Rpoly was because of the way value of current was taken. Bug was found and fixed. Next would be optimization of the program eliminating not necessary delays and crosschecks. This doesn’t interfere with testing in general.


Download ppt "1/13/2004USCMS Tracker Meeting Sergey Korjenevski U of R Status Report IV curve and “second kink” QA stand status - R poly measurement has been debugged."