SIA of trace gases by CF-IRMS
2
Sample Types CO 2, N 2 O, N 2, CH 4 ; other gases by arrangement Atmospheric, soil, breath, and dissolved gases Natural abundance to 99 atom% (>50 atom% by prior authorization only) Dual-tracer method (N 2 & N 2 O)
3
Sample Preparation Labco Exetainers, 12mL Avoid “wide” exetainers (###W/W) Other vials suitable by prior arrangement Fill to ~ atmospheric pressure Butyl or butyl/silicone septa are best Overtightened septa can leak and clog autosampler needles Overtightening deforms septa
4
Sample Preparation If samples require poisoning, use ~1mL 50% w/v ZnCl 2 ; UCD SIF cannot accept samples containing HgCl 2 Please provide estimates of gas concentration and enrichment or provide test samples For DIC, prepared samples do not require poisoning and should not be frozen
5
Evacuate and flush-fill sample vials Do not buy “pre- evacuated” vials from Labco (hint: they are not empty). If vials will not be filled with sample gas, flush fill your vials with helium to near atmospheric pressure.
6
Sample Organization and Shipping If applicable, organize samples by increasing isotopic enrichment Label vials clearly; avoid adding labels on top of labels Protect glass from impact and label boxes as fragile
7
Paperwork Complete an Analysis Order Form for each analysis type Complete a sample list for each Analysis Order From Email forms to sif@ucdavis.edu Include printed copies with samples
8
SIA of trace gases by CF-IRMS Three dedicated trace gas instruments ThermoScientific and Sercon Atmospheric concentrations to pure gases
9
Chromatography
10
Data Review
11
Data Reduction and Corrections Blank Drift Size Scale expansion Corrections follow those of Kaiser and Röckmann 2008 Kaiser, J. & T. Röckmann 2008. Rapid Commun. Mass Spectrom.