ACD calibrations Pedestals Measured from online script Measure PHA w/ HV off, no charge injection Use cyclic triggers ~100-300 ADC counts, very narrow (rms 2-3 counts) Gains Measured from self-triggered data Look for MIP peak w/ HV on W/ full LAT use TKR extrapolation W/ ACD only use coincidence to reduce angular distribution
2
PHA distribution in a single channel MIP peak Low energy noise Seen in both PMT (probably x-rays in same shower as muon) Zero suppression is on for these data (15 cts above pedestal) Veto threshold (0.3 MIP)
3
Fitting the MIP peak To determine fit range 1)Find local min between low energy noise & MIP 2)Find MIP peak 3)Find ½ max value after MIP peak Fit between 1) and 3) Following procedure from ACD group, fit to polynomial Around MIP peak and Extract peak value
4
Robustness of fitting procedure Procedure is pretty robust for tiles. On first calibration run tried it only failed in 1 out of 178 PMTs. Need to pick a fallback option anyway. Ribbons don’t have a clear MIP peak Need more work on these
5
Scatter in MIP peak values