Testing Calorimeter with cosmic rays By: Matteo Pallocca (Liceo Scientifico A.Landi, Velletri) Agnese Capobianco (Liceo Scientifico E.Amaldi, Roma) Federica Forcesi (Liceo Scientifico Farnesina, Roma) Tutor: Marco Mirazita Technician: Walter Pesci
2
Stage objectives Testing a lead/scintillator sampling calorimeter using cosmic muons. Calculation of Voltage Curves and Calibration Line.
3
Our Detector PAD 1 PMT CALORIMETER PAD 2
4
Electronic Acquisition System
5
Power Supply Discriminator Logic AND Delays QDC FAN IN/ FAN OUT
6
Example of a QDC spectrum Background Signal Peak, released by a muon
7
Voltage Curves
8
Signal after the equalization
9
Calibration line
10
Conclusions We set up an acquisition system to test a calorimeter using cosmic muons. We calculated voltage curves for four different voltages. We equalized the 4 central PMTs. We calculated the calibration line of our calorimeter and verified its linear behavior.