FYS 4250 Kap.14 Electrical safety
2
Electric current, perception- and hazard thresholds
device related patient injuries in USA each year Three phenomenon with current through the body: Electric stimulation of excitable tissue (nerve & muscle) Resistive heating of tissue Electrochemical burns and tissue damage for direct current and very high voltages Perception = tingling sensation
3
Electrovibration (1uA 50Hz)
4
Statistical scattering of sensitivity for electric current
5
Frequency dependency and spread of the let-go threshold
Let go current rais below 10 Hz, muscles can partially relax during part of each cycle
6
Ventricular fibrillation thresholds in animals
Duration of applied current is an important factor.
7
Macro- og microshock Macroshock 100 mA Microshock 10 uA
8
Electromedical devices must be CE-marked (Europe)
9
Patient near PC (B-apparatus)
10
American electric power distribution
11
4-leads (norske), og 5-leads (europeiske) electrical circuits fra ”Medisinsk-teknisk sikkerhet på sykehus” Grimnes&Jensen 2003 Norwegian elctrical network is floating with respect to ground = IT network (No current dropout for the first ground short circuit European TN.
12
Isolated system
13
Macroshock due to ground fault
14
Microshock, two error situations
15
Ventricular fibrillation and pump failure vs catheter area
16
Micro-shock A device with a ground fault that does not open the circuit breaker. 5 A flow to the distribution panel on the ground wire. Ground wire have 0.1 ohm resistance -> 500 mV could appear across the patient. Resistance of a patient’s body and liquid filled catheter is less than 50 kOhm -> current = 10 uA
17
System for grounding
18
Ground fault circuit-interupter
Disconnect the source of electric power when a ground fault greater than 6 mA occurs. Senses the difference between the hot conductor and neutral conductor Trip time varies inversely with the magnitude of the ground-fault current
19
Floating Patient
20
Electrical isolation Perfect isolation barrier modeled by the isolation capacitance and resistance. IMRR = isolation mode rejection ratio (potential between the input common and the output common)
21
Electric isolation
22
Isolated pressure transducer
23
Receptacles tests
24
Ground pin to chassis test
Should not exceed 0.2 Ohm
25
Chassis leakage cuttent test
26
Patient leakage current test
27
Leakage current between patient leads
28
Isolation test for AC isolation
Equipment manufacturer only, may be omitted for testing in health care facilities