Download presentation
Presentation is loading. Please wait.
1
Lecture 3 INFRARED SPECTROMETRY
Copyright ©The McGraw-Hill Companies, Inc. Permission required for reproduction or display.
2
INTRODUCTION TO INFRARED SPECTROMETRY
INFRARED SPECTRAL REGIONS INTRODUCTION TO INFRARED SPECTROMETRY
3
INFRARED ABSORPTION SPECTRUM
4
INTRODUCTION TO INFRARED SPECTROMETRY
Energy of IR photon insufficient to cause electronic excitation but can cause vibrational or rotational excitation Molecule electric field (dipole moment) interacts with IR photon electric field (both dynamic) Magnitude of dipole moment determined by (i) charge (ii) separation of charge INTRODUCTION TO INFRARED SPECTROMETRY
5
INTRODUCTION TO INFRARED SPECTROMETRY
Vibration or rotation causes varying separation INTRODUCTION TO INFRARED SPECTROMETRY
6
INTRODUCTION TO INFRARED SPECTROMETRY
Molecule must have change in dipole moment due to vibration or rotation to absorb IR radiation! Absorption causes increase in vibration amplitude/rotation frequency INTRODUCTION TO INFRARED SPECTROMETRY
7
Molecules with permanent dipole moments (μ) are IR active!
INTRODUCTION TO INFRARED SPECTROMETRY
8
INTRODUCTION TO INFRARED SPECTROMETRY
9
INTRODUCTION TO INFRARED SPECTROMETRY
Types of Molecular Vibrations: Stretch - change in bond length symmetric assymetric Bend – change in bond angle scissoring wagging rocking twisting/tortion INTRODUCTION TO INFRARED SPECTROMETRY
10
INTRODUCTION TO INFRARED SPECTROMETRY
STRETCH INTRODUCTION TO INFRARED SPECTROMETRY
11
GENERAL DESIGN OF OPTICAL INSTRUMENTS
BEND GENERAL DESIGN OF OPTICAL INSTRUMENTS
13
GENERAL DESIGN OF OPTICAL INSTRUMENTS
Only some modes may be IR active: Example CO2 O=C=O linear GENERAL DESIGN OF OPTICAL INSTRUMENTS
14
GENERAL DESIGN OF OPTICAL INSTRUMENTS
15
GENERAL DESIGN OF OPTICAL INSTRUMENTS
16
GENERAL DESIGN OF OPTICAL INSTRUMENTS
Classical vibrational motion: GENERAL DESIGN OF OPTICAL INSTRUMENTS
18
GENERAL DESIGN OF OPTICAL INSTRUMENTS
19
CLASSICAL VIBRATIONAL FREQUENCY
20
INFRARED SPECTROSCOPY
What about quantum mechanics? INFRARED SPECTROSCOPY
21
SAMPLE PROBLEM The force constant for a typical triple bond is 1.91 x 103 N/m. Calculate the approximate frequency of the main absorption peak due to vibration of CO.
22
INFRARED SPECTROSCOPY
23
= ±1 Vibrational Selection Rule Since levels equally spaced - should see one absorption frequency. SELECTION RULE
24
GENERAL DESIGN OF OPTICAL INSTRUMENTS
Anharmonic oscillator: Must modify harmonic oscillator potential for (i) electron repulsion (steeper at small distances) (ii) dissociation (bond breaks at large distances) GENERAL DESIGN OF OPTICAL INSTRUMENTS
25
New E-y curve: IR SPECTROSCOPY
26
IR SPECTROSCOPY
27
VIBRATIONAL MODES
28
How many vibrational mode? HCN, H2CO3
29
Coupling of different vibrations shifts frequencies Coupling likely when: (1) common atom in stretching modes (2) common bond in bending modes (3) common bond in bending + stretching modes (4) similar vibrational frequencies Coupling not likely when (1) atoms separated by two or more bonds (2) symmetry inappropriate IR SPECTROSCOPY
30
IR SPECTROSCOPY
31
IR SPECTROSCOPY - INSTRUMENTATION
SOURCES IR SPECTROSCOPY - INSTRUMENTATION
32
IR SPECTROSCOPY - INSTRUMENTATION
TRANSDUCERS IR SPECTROSCOPY - INSTRUMENTATION
33
IR SPECTROSCOPY - INSTRUMENTATION
34
IR SPECTROSCOPY - INSTRUMENTATION
35
FOURIER TRANSFORM INFRARED SPECTROMETER A photodiode array can measure an entire spectrum at once. The spectrum is spread into its component wavelength and each wavelength is directed onto one detector element. Fourier Analysis is a procedure in which a spectrum is decomposed into a sum of sine and cosine terms called a Fourier series.
36
IR SPECTROSCOPY - INSTRUMENTATION
Fourier Transform Instruments have two advantages: IR SPECTROSCOPY - INSTRUMENTATION
37
FOURIER TRANSFORM
38
FOURIER TRANSFORM
39
GENERAL DESIGN OF OPTICAL INSTRUMENTS
Unfortunately, no detector can respond on s time scale Use Michelson interferometer to measure signal proportional to time varying signal GENERAL DESIGN OF OPTICAL INSTRUMENTS
40
IR SPECTROSCOPY - INSTRUMENTATION
41
GENERAL DESIGN OF OPTICAL INSTRUMENTS
If moving mirror moves 1/4 (1/2 round-trip) waves are out of phase at beam-splitting mirror - no signal If moving mirror moves 1/2 (1 round-trip) waves are in phase at beam-splitting mirror - signal GENERAL DESIGN OF OPTICAL INSTRUMENTS
42
GENERAL DESIGN OF OPTICAL INSTRUMENTS
Difference in pathlength called retardation Plot vs. signal - cosine wave with frequency proportional to light frequency but signal varies at much lower frequency One full cycle when mirror moves distance /2 (round-trip = ) velocity of moving mirror GENERAL DESIGN OF OPTICAL INSTRUMENTS
43
GENERAL DESIGN OF OPTICAL INSTRUMENTS
44
GENERAL DESIGN OF OPTICAL INSTRUMENTS
45
GENERAL DESIGN OF OPTICAL INSTRUMENTS
46
GENERAL DESIGN OF OPTICAL INSTRUMENTS
47
GENERAL DESIGN OF OPTICAL INSTRUMENTS
48
GENERAL DESIGN OF OPTICAL INSTRUMENTS
APPLICATIONS: FTIR single-beam, dispersive IR double-beam, but FTIR advantages include • High S/N ratios - high throughput • Rapid (<10 s) • Reproducible • High resolution (<0.1 cm-1) • Inexpensive GENERAL DESIGN OF OPTICAL INSTRUMENTS
49
GENERAL DESIGN OF OPTICAL INSTRUMENTS
IR (especially FTIR) very widely used for • qualitative • quantitative analysis of • gases • liquids • solids GENERAL DESIGN OF OPTICAL INSTRUMENTS
50
GENERAL DESIGN OF OPTICAL INSTRUMENTS
Most time-consuming part is sample preparation Gases fill gas cell (a) transparent windows (NaCl/KBr) (b) long pathlength (10 cm) - few molecules Liquids fill liquid cell (a) solute in transparent solvent - not water (attacks windows) (b) short pathlength ( mm) - solvents absorb GENERAL DESIGN OF OPTICAL INSTRUMENTS
51
GENERAL DESIGN OF OPTICAL INSTRUMENTS
52
GENERAL DESIGN OF OPTICAL INSTRUMENTS
53
TRANSFORMATION OF ALKENES AND ALKYNES
55
SAMPLE
56
SAMPLE
57
Group Frequencies: • Approximately calculated from masses and spring constants • Variations due to coupling • Compared to correlation charts/databases
59
Quantitative Analysis: IR more difficult than UV-Vis because • narrow bands (variation in e) • complex spectra • weak incident beam • low transducer sensitivity • solvent absorption IR mostly used for rapid qualitative but not quantitative analysis
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.