Download presentation
1
TIMERS
2
BASIC X-RAY SCHEMATIC
3
TIMER CIRCUIT: IS SEPARATE FROM THE OTHER MAIN CIRCUITS OF THE IMAGING SYSTEM. TIMER
4
TIMER A MECHANICAL OR ELECTRONIC DEVICE WHOSE ACTION IS TO MAKE OR BRAKE THE HIGH VOLTAGE ACROSS THE X-RAY TUBE
5
TYPES OF TIMING CIRCUITS TIMERS:
MECHANICAL SYNCHRONOUS ELECTRONIC mAs BACKUP AUTOMATIC EXPOSURE CONTROL
6
MACHANICAL TIMER OPERATE BY CLOCKWORK. SPRING MECHANISMS IS WOUND UP.
DURING EXPOSURE THE SPRING UNWINDS AND TERMINATES EXPOSURE.
7
SYNCHRONOUS TIMER IN SYNCH WITH FREQUENCY OF THE A/C CURRENT (60 Hz in the US). MINIMUM EXPOSURE TIME IS 1/60 s. TIMING INTERVALS INCREASE BY 1/60 s. MUST BE RESET AFTER EACH EXPOSURE
8
ELECTRONIC TIMER THE MOST SOPHISTICATED. THE MOST ACCURATE
IT’S CIRCUITRY CONSISTS OF THYRATRON TUBE OR THYRISTOR. ALLOW WIDE RANGE OF TIME INTERVALS. ALLOW SERIAL EXPOSURES. NOWADAYS CONTROLLED BY MICROPROCESSOR
9
THYRATRON CIRCUIT
10
mAs TIMER MONITORS THE PRODUCT OF mA & EXPOSURE TIME AND TERMINATES EXPOSURE WHEN THE DESIRED mAs VALUE IS ATTAINED.
11
BACKUP SET TO TERMINATE EXPOSURE IN THE EVENT OF FAILURE IN THE AEC CIRCUIT. MAX. mAs ALLOWABLE – 600 mAs UNLESS TUBE POTENTIAL IS LESS THAN 50 kVp – 2,000 mAs IS THE LIMIT.
12
AUTOMATIC EXPOSURE CONTROL
Automatic exposure control devices can assist the radiographer in producing consistent radiographic images from patient to patient, regardless of size or presence of pathology. The advantages of this consistency are numerous and include: decreased repeat rate; decreased patient exposure; and increased department efficiency. The most important benefit being decreased repeat rate
13
When an AEC device is used to terminate an exposure, the technologist sets the kVp and mA, but the time of the exposure is automatically determined by the machine. The AEC device differs from a manual timer because the AEC does not stop the exposure until the film has reached an appropriate density. Unlike manual timers, which simply stop the exposure after the preset time has elapsed.
14
A major benefit of the AEC device is its ability to consistently obtain accurately exposed radiographs, even in the presence of pathology. While manual timers terminate the exposure at the preset time, regardless of pathology or achievement of proper film density. The following example demonstrates the difference
15
AUTOMATIC EXPOSURE CONTROL (AEC)
FLAT IONIZATION CHAMBER PHOTOTIME
16
IONIZATION CHAMBER AEC
X-RAY TUBE X-RAY TABLE BUCKY
17
IONIZATION CHAMBERS ARE SITUATED
BETWEEN THE GRID MECHANISM & THE CASSETTE
18
CHAMBERS
19
CHAMBER SELECTION ON THE CONSOLE
DETECTOR SELECTION Along with proper
20
AEC
21
IONIZATION CHAMBER AEC
X-RAY TUBE X-RAY TABLE CASSETTE SCREEN
22
PM TUBE LIGHT ELECTRONS
23
PHOTOTIME MECHANISM IS SITUATED
BEHIND THE CASSETTE CASSETTE NEEDS SPECIAL DESIGN IT HAS OPENING IN THE BACK
24
PHOTOTIME CASSETTE
25
PHOTOTIME SCREEN X
26
Even though the ionization chamber and the phototimer operate differently, they both have the same function: convert radiation into an electrical signal which will be used to automatically stop the exposure when the film has reached the proper density
27
FACTORS INFLUENCING AEC
PATIENT POSITIONING DETECTOR SELECTION
28
POSITIONING!!!! AEC USES DETECTORS TO RECEIVE THE RADIATION THAT PASSED THROUGH THE PATIENT WE MUST POSITION THE ANAOMY OF INTEREST DIRECTLY ABOVE THE DETECTORS.
30
Without a technologist who is very knowledgeable about anatomy and positioning, automatic timers are worthless. In fact, they may actually decrease department efficiency because of the increased amount of repeat radiographs that will result if used improperly.
31
DETECTOR SELECTION
32
LAT. L-SPINE
33
C-SPINE
34
C-SPINE -AP
35
SHOULDER
36
KNEE
37
HIP
38
PELVIS
39
ABDOMEN
40
CHEST PA
41
LAT. CHEST
42
DENSITY SETTING D. SETTING % DENSITY CHANGE +4 +3 +2 +1 -1 -2 -3 -4
-1 -2 -3 -4 +100 +75 +50 +25 -25 -50 -75 -100
43
MINIMUM RESPONSE TIME SHORTEST EXPOSURE TIME POSSIBLE
44
EFFECTS OF mA, kVp, SID ON AEC
mA AEC TIME kVp AEC TIME SID AEC TIME
45
PROBLEMS WITH AEC PATIENT PATHOLOGY PROSTHETIC DEVICE COLLIMATION
STROBOSCOPIC EFFECT WRONG IMAGE RECEPTOR
46
CERTAIN PATHOLOGIES ALTER AEC READING
47
PROSTHESIS GIVES FALSE AEC READING
48
NOT ENOUGH COLLIMATION ALLOWS TOO MUCH SCATTER REACHING THE DETECTOR AND PREMATURE TERMINATION OF THE EXPOSURE TAKES PLACE
49
STROBOSCOPIC EFFECT TOO SHORT OF THE EXPOSURE TIME IS SELECTED BY THE AE, SHORTER THAN THE SPEED OF THE OSCILLATION OF THE GRID AND IT FREEZES THE GRID MOVEMENT.
50
WRONG SPEED OF IMAGE RECEPTOR!!!
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.