Download presentation
1
Hemodynamic Monitoring Part I (ABP, CVP, Ao)
MICU Competencies
2
What is Hemodynamic Monitoring?
Non-invasive = clinical assessment & NBP Direct measurement of arterial pressure Invasive hemodynamic monitoring
3
Noninvasive Hemodynamic Monitoring
Noninvasive BP Heart Rate, pulses Mental Status Mottling (absent) Skin Temperature Capillary Refill Urine Output
4
Proper Fit of a Blood Pressure Cuff
Width of bladder = 2/3 of upper arm Length of bladder encircles 80% arm Lower edge of cuff approximately 2.5 cm above the antecubital space
5
Why A Properly Fitting Cuff?
Too small causes false-high reading Too LARGE causes false-low reading
6
Indications for Arterial Blood Pressure
Frequent titration of vasoactive drips Unstable blood pressures Frequent ABGs or labs Unable to obtain Non-invasive BP
7
Supplies to Gather Arterial Catheter Pressure Tubing Pressure Bag
Pressure Cable Pressure Bag Flush – 500cc NS
8
Supplies to Gather Sterile Gown (2) Sterile Towels (3) Sterile Gloves
Suture (silk 2.0) Chlorhexidine Swabs Mask
9
Leveling and Zeroing Leveling Before/after insertion
If patient, bed or transducer move Zeroing Performed before insertion & readings Level and zero at the insertion site
10
Potential Complications Associated With Arterial Lines
Hemorrhage Air Emboli Infection Altered Skin Integrity Impaired Circulation
11
Documentation Insertion procedure note ABP readings as ordered
Neurovascular checks every two hours (in musculoskeletal assessment of HED) Pressure line flush amounts (3ml/hr) Tubing and dressing changes
12
Central Venous Pressure Assesses . . .
Intravascular volume status Right ventricular function Patient response to drugs &/or fluids
13
Central Venous Pressure (CVP)
Central line or pulmonary artery catheter Normal values = 2 – 8 mm Hg Low CVP = hypovolemia or ↓ venous return High CVP = over hydration, ↑ venous return, or right-sided heart failure
14
Leveling and Zeroing Leveling Before/after insertion
After patient, bed or transducer move Aligns transducer with catheter tip Zeroing Performed before insertion & readings Level and zero transducer at the phlebostatic axis
15
Phlebostatic Axis 4th intercostal space, mid-axillary line
Level of the atria (Edwards Lifesciences, n.d.)
16
More on Leveling and Zeroing
HOB 0 – 60 degrees No lateral positioning Phlebostatic axis with any position (dotted line) (Edwards Lifesciences, n.d.)
17
Dynamic Flush Dynamic flush ensures the integrity of the pressure tubing system. Notice how it ascends - forms a square pattern - and bounces below the baseline before returning to the original waveform. Check dynamic flush after zeroing any pressure tubing system
18
System Maintenance Change tubing and fluid bag q 96hrs
No pressors through CVP port Antibiotics, NS boluses, blood, & IV pushes are allowed through the CVP line
19
Troubleshooting Improper set-up and equipment malfunction are the primary causes for hemodynamic monitoring problems Retracing the set-up process or tubing (patient to monitor) may identify the problem and solution quickly Use your staff resources: Help All, Charge Nurse, Educator, Preceptors, MICU experts
20
Troubleshooting Damped Waveforms Pressure bag inflated to 300 mmHg
Reposition extremity or patient Verify appropriate scale Flush or aspirate line Check or replace module or cable
21
Troubleshooting Inability to obtain/zero waveform
Connections between cable & monitor Position of stopcocks Retry zeroing after above adjustments
22
Continuous Airway Pressure (Ao)
Also known as Paw, Ao Purpose: Improves accuracy of hemodynamic waveform measurements Identification of end-expiration Positive waveform deflections = positive pressure ventilation Negative deflections = spontaneous inspiratory effort
23
Supplies to Gather Pressure Cable Pressure Tubing Connector
(Edwards Lifesciences, n.d.)
24
Setting up the Ao Discard infusion spike end & cap port
Connect pressure tubing to vent tubing (using connector opposite heating cable) Connect cables Zero the tubing (leveling not necessary)
25
Troubleshooting Ao Do not prime tubing with fluids!
Damping will occur with fluid or secretions To evacuate any fluids, disconnect pressure tubing from vent tubing and push air through the pressure tubing with a 10 ml syringe connected at one end until fluid-free
26
Pressure Measurement CVP=13 1) Record Ao and CVP on the same strip
2) Find end-expiration by drawing a vertical line with a straight edge 200 ms prior to the rise or dip in Ao (1 large box) associated with a breath. 3) Draw a horizontal line through the visually assessed average vascular pressure starting at end-expiration going backward 200 ms (1 large box). 4) Read the pressure at the horizontal line. 15 10 5 -5 CVP=13
27
Assist-Control 200 ms { Ao { CVP 200 ms
28
CPAP with Pressure Support
200 ms { Ao { 200 ms CVP
29
CPAP without Pressure Support
200 ms { Ao { 200 ms CVP
30
Incorrect method! This point was identified as end-expiration for a pt. who did not have an Ao set up. 40 30 20 10 -10 Correct method! 30 sec after the above tracing, Ao was added & true end-expiration clearly identified.
31
Same patient 20 minutes later
40 30 20 10 -10 40 30 20 10 -10
32
15 10 5 -5 CVP=13
33
Summary Record Ao with CVP
Read mean CVP at end-expiration as described. No need read vascular pressure at any particular time in the cardiac cycle
34
Documentation of CVP Include on waveform strip
Position of the HOB Vasopressors and rates Amount of PEEP Scale CVP measurement Signature of the nurse (post in green chart behind graphics tab)
35
References & Resources
Burns, S. M. (2004). Continuous airway pressure monitoring. Critical Care Nurse, 24(6), Chulay, M., & Burns, S. M. (2006). AACN Essentials of critical care. McGraw-Hill: New York. Edwards. (2006). Pulmonary Artery Catheter Educational Project. Edwards Lifesciences. (n.d.) Educational videos. MICU Routine Practice Guidelines. MICU Bedside Resource Books MICU Education Kits (Red cart in conference room) MICU Preceptors, Help All Nurses, & Charge Nurses VUMC policies.
Similar presentations
© 2024 SlidePlayer.com. Inc.
All rights reserved.