Download presentation
Published byCassandra Evans Modified over 9 years ago
1
Liquidborne Particle Counting using Light Obscuration and Light Scattering Methods
2
What has been . . . Focus has been on injectable liquids
Possibility to block capillaries and arteries Red Blood cells are about 5 µm Capillary (5 to 10 µm) Large veins (10 to 50 µm) Threat of microbial infection Allergic reaction to foreign substances
3
Contaminants Reported in IV Solutions
Extrinsic
4
Definition of Particulate Contaminants
Unwanted insoluble matter that exist as “randomly-sourced extraneous substances” Excludes homogeneous monotonic materials that exist as a precipitate or suspension i.e. colloids, drug degradation or otherwise derived from a defined source and can be analyzed by chemical means Regarded as “contamination” and “adulteration” under Food and Drug Act the chemical composition of the particulate is varied, and would not be declared on the label Examples: bits of paper fiber, fragments of filler material, etc
5
Liquid Particle Counting Applications
Final Product Testing – USP <788> SVP or SVI (Small Volume Parenteral/Injectable) Ampoules, Vials LVP or LVI (Large Volume Parenteral/Injectable) IV (Intravenous) solutions Process contamination studies Decomposition studies (stability) DI or WFI Water Precision Cleaning – Medical Devices Aqueous Other Chemicals
6
Other Applications for Particle Counting
Medical Devices Cleanliness of manufacturing environment Cleanliness of device before implantation pacemakers, stents, artificial arteries Cleanliness of reclaimed devices Design of particulate-based medicines Inhalation therapies Intentional occlusion of blood flow to cancers Time-based dosages Transdermal absorption
7
Global Regulations: Particles in Liquids
USP <788>, EP , JP XV, KP, CP Primary method Optical Particle Counter [OPC] Light Obscuration Counter Secondary method Optical microscope Subjective Labor intensive Requires more time to process samples
8
Proposed: USP 787, USP 1787 USP <787> Under discussion
Focused on reducing necessary test volumes due to concerns of biotechnology manufacturers of cost for test Expensive and often very small dose factory for example: 500 uL pre-filled syringe
9
Proposed: USP 787, USP 1787 USP <787> Primary method ?
Optical Particle Counter [OPC] Light Obscuration Counter Secondary method ? Optical microscope Subjective Labor intensive Requires more time to process samples
10
Proposed: USP 787, USP 1787 USP <787> Small sample volume
- 1 mL ?? Concerns with variability - within production lots - in analytical methods
11
Optical Particle Counter
Optical Instrument Must move fluid through sensor Can quantify particles from 100 nm to 5000 µm Counts particles individually (one at a time) Cannot tell you composition But results are immediate 11
12
Alumino-silicate with K and Ti
Many shapes and sizes Alumino-silicate with K and Ti Talc
13
Sizing Particles by Microscope
Largest Dimension d Martin’s Diameter Area A Area B d Ferret’s Diameter d Projected Area d 13
14
Challenges of Protein-based Products
Handling can change material !!! Agitation Heat and Light Contaminates Container: Vials versus syringes/cartridges Shear forces Key concern is Aggregation Reduction of native form (impacts efficacy) Introduction of homogeneous aggregates Introduction of heterogeneous aggregates
15
Challenges of Protein-based Products
Transparency of most proteineous entities Refractive index NIST working on calibration material Not “contamination” but instead a shift from native form Not a solution as with small-molecule therapeutics Formation of quaternary structures [dimers, etc.] Protein complexes Reconstitution of lyophilized product
16
Refractive Index Key is the ability to distinguish between the particle and the surrounding fluid - needs to be great enough Optical response is proportional to comparative index
17
Refractive Index Key is the ability to distinguish between the particle and the surrounding fluid - needs to be great enough Optical response is proportional to comparative index
18
Refractive Index NIST working on protein-like calibration material
Probably 2 years away Exploring 2 methods of manufacture Need thread-like material Indices near water Stable over reasonable period
19
II. Sample Handling
20
Settling/Agitation Entrained gas
- sonication probably not ideal with protein structures - light vacuum seems to work OK Settling Limits collection of particles - especially of greater mass - dependent on time and viscosity - improved collection with slanted containers
21
Consistency of sample characteristics
Temperature Settling Probe position
22
Issues with Sampling Particles in Liquids
Sampling Errors Account for most problems Accidental Contamination or Alteration by Technician 1. System Preparation Initial Cleanliness Calibration 2. Sample Preparation Contamination - Particles - Gases - Liquids 3. Sample Handling Aggregation Settling Cavitation
23
Particle Counter vs. Microscope
Sizing of Particles Microscope: Subjective size measurement Electronic: consistent for specific particle type Particle orientation and movement Electronic: large, long particles can be mis-read Calibration How to establish accurate sizing and counting Consistency (repeatability)
24
Sizing Particles by Microscope
Largest Dimension d Martin’s Diameter Area A Area B d Ferret’s Diameter d Projected Area d 24
25
HIAC Liquid Particle Counters
Example: HIAC 9703 The industry standard liquid particle counter since 1997 USP <788> was written specifically around HIAC technology Every major manufacturer of particle calibration standards uses the HIAC 9703
26
HIAC Liquid Particle Counters
Example: HIAC 9703+ Improved sample mounting method for small vials or containers Detection of usual conditions such as bubbles or contamination Proven syringe sampler SVI and LVI sampling Addresses non-compendial applications, e.g. R&D and other low frequency, small sample volume applications Reproducibility Repeatability
27
Light Obscuration Light Scattering
Detection Ranges 0.1µm 1µm 10µm 100µm Light Obscuration Light Scattering nm
28
Light Obscuration Light Obscuration Sensors and system
also known as Light Extinction also known as Light Blocking
29
Principles: Light Obscuration
Detector Output
30
Principles: Light Obscuration
Detector Output
31
Principles: Light Obscuration
32
Particle Detection Like an air particle counter, the larger the particle, the larger the pulse that is created
33
Principles: Light Scattering
Detector Detector Output Light Trap Laser Diode Mirror
34
Principles: Light Scattering
Detector Particle Light Trap Laser Diode Mirror
35
Advantages: Light Scattering
Good sensitivity from 0,1µm to 50µm Wide range of sample concentration Good rejection of false counts High sample flow rates
36
Disadvantages: Light Scattering
More complicated construction = higher cost Characteristics of particle surface (shiny, color) affect response
37
Effect of colors and surfaces on Light Scattering
Dark Light Shiny
38
Alumino-silicate with K and Ti
Talc
39
Sizing Particles by Microscope
Largest Dimension d Martin’s Diameter Area A Area B d Ferret’s Diameter d Projected Area d
40
General Comments on Liquid Counting
Particle Counters Report Size But measure an Optical Response Difference in reported size compared to microscope Calibration Relates the Optical Signal to Size Difference between calibration material characteristics and “real world” particles Projected Area d 40
41
General Comments on Liquid Counting
Particle Counters Report Size But measure an Optical Response Differences in reported size compared to microscope Calibration Relates the Optical Signal to Size Difference between calibration material characteristics and “real world” particles 41
42
LO results versus LS results
Light Obscuration [LO] Good immunity to variations of surface and morphology Very stable Limit of quantitation circa 1.2 – 1.3 microns Light Scattering [LS] Results affected by surface characteristics and coloring Good stability Limit of quantitation sub-micron Problem can occur in the attempt to correlate results of these two methods above 1 micron
43
Sampling Particles in Liquids
Good, consistent results depend on: Well-trained operators Careful technical methodology Constant effort needed: Review ways to avoid or reduce errors due to sampling methods Usually cause “false positives”
44
System Preparation 2-step Verification - optional:
Run 2 test solutions Blank (“particle-free” water) Aqueous solution containing known counts at 10 µm or 15 µm In the range of 1000 to 3000 counts per mL Frequency – based on risk analysis Each morning Shift change Change of operator Other interval
45
System Preparation Check for bubbles in sample lines and syringe
Affects flow rate and calibration Verify correct calibration curve installed Different flow rates for same sensor Change of syringe size Some companies have multiple sensors Verify calibration is current Sensor resolution and response curve IST tests conducted [USP, JP]
46
System Preparation Instrument Standardization Tests [IST]
Five tests of system Volume accuracy Flow rate accuracy Calibration of sensor Resolution Count accuracy Required by USP and JP but not EP USP <788> 31 “…at intervals of not more than six months.” JP <24> “…at least once a year.”
47
Sample Preparation Removing residue of previous sample
If previous sample contained particles, may be good plan to run a “particle-free” blank before the next sample Use liquid that is compatible with sample fluid An aqueous blank could trigger false counts in an oil-based sample by causing immiscible droplets Potential residue from previous sample can cause change of counts Data from first run of series is often discarded
48
Sample Preparation Contamination Particles Gases Liquids
49
Sample Preparation True Particle Sources False Particle Sources
Residue on glassware and equipment Ambient air Paper dust Glass Diluent Residue from previous sample Colloidal suspensions False Particle Sources RF signals or other electronic interference Bubbles from entrained gases
50
Sample Preparation Work in controlled environment
Use particle-free gloves Let water flow for 200 mL or more after opening a valve / tap Opening / closing valve generates particles Wipe the outside of containers before sampling Particles on surface of vials or ampoules Open vials and ampoules away from beaker or flask Particles from activity can fall into open container Wash outside of containers to reduce potential particle source
51
Degassing sample Three common methods Allowing to stand in ambient air
Risk of large particles settling Sonification [ultrasound] 80 to 120 watts [USP <788>] 30 seconds [USP <788>] Vacuum Bell jar or dessicator 0.6 – 0.8 atmospheres for 2 to 10 minutes
52
Sample Preparation Possibility of decreasing true particle counts
Settling Lack of agitation Position of probe in sample vessel Remaining material from previous sample run Sample with lower counts Blank
53
Sample Handling Aggregation Settling Entrained gases
54
Sample Handling Aggregation
Samples held over time or at extremes of temperature can develop aggregates of smaller particles Exposure to light can also trigger this reaction Sub-micron particles can thus add to particle counts above 10 µm Suggestion: Profile counts under 10 µm [e.g. at 2 or 5 µm] in addition to standard count sizes at 10 and 25 µm
55
Sample Handling Settling Undercounting caused by
Gravitational settling Failure to suspend particulate matter adhering to walls or stopper of container Standards have recommended procedure for agitation Multiple inversions of original container before decanting Constant motion of liquid during sampling “Gently stir the contents of the container by hand-swirling or by mechanical means…” USP <788>
56
Optical Particle Counter
Instrument & sensor Must move fluid through sensor Can quantify particles from 100 nm to 5000 µm Counts particles individually (one at a time) Cannot tell you composition But results are immediate 56
57
Questions?
58
Patent Pending
59
New Hardware: 9703+ Key features Auto stop for sensor elevator arm
Small vial holding clamp Sample probe with reduced dead volume Back-flush and forward flush from front panel Supports MC-05 sensor (0.5 micron sensitivity)
60
Software: PharmSpec 3 Key new features
Compendial test support continues USP, EP, JP, KP looks same as previous PharmSpec versions Uses same log on as for Windows Improved Report format Improved Error Detection and Display
61
Still the HIAC 9703 you know and trust – only better!
Syringes 1 ml, 10 ml, 25 ml Flow rate settings 10 to 100 ml Sensors MC-05 is added Sampling Probes added shorter small-bore probe Instrument size / shape 50%+ of instruments are placed in laminar flow cabinets. Smooth, curved surfaces create less turbulence for the air flow
62
Easier, Faster and Confident Sampling
Use less sample, save valuable time - protect your investment Small vial clamp ensures that sample does not spill during testing Probe needle safety switch prevents probe damage New small needle probe with industry’s smallest tare volume
63
Easier, Faster and Confident Sampling
Reduce uncertainty from data anomalies Bubble alarm notification Contamination alarm notification Advanced notification when service or calibration is due Invalid configuration notification
64
HIAC just got easier! Less time needed for clean-up
Automated flushing and cleaning routines Push a button, walk away and return to a clean sensor Export your data with ease Select one, several or all of your historical data records with our batch export utility Select PDF, Word, Excel, or text files Save time with electronic signature Stricter interpretation of 21CFR Part 11 electronic signature process…. WITHOUT more manual inputs Remembers user Login ID
65
HIAC Flexibility Interchangeable sampling probes, syringes, and sensors Ensure you have one instrument to manage all applications Now supports MC05 sub-micron sensor Change configuration with no impact to instrument validation Customized reporting Customize the number of reviewers and approvers for compendial test reports Add company logo, user-defined descriptors Customized test recipes Procedure Builder enables the development of unique test recipes for your application Enables testing to marketing license-specific applications
66
New Sampling Probe 3 probes available Tare New small / short probe
¼” ID =1.2 ml tare volume 1/16” ID = ml tare volume New small / short probe 1/16” ID = 0.09 ml tare volume Tests can be performed on 1 ml of product!
67
Small Vial Clamp Small Vial Clamp Platform Can be retrofitted
Can be removed Ease of Use Use one hand to compress lever arms Use other hand to place sample in central location Decompress hand Clamp auto-centers and holds sample container
68
Docking Module Docking Module
Enables removal/disabling of the stir bar mechanism Enables field installation of small vial clamp outside of the laminar flow cabinet Avoid potential of re-qualification that can occur if instrument is moved Future developments to expand 9703 applications
69
Sampling Safety Switch
Ensures the sampling probe does not crash (and bend or break) into the docking module Ensures the probe does crash into or tip the sample container
70
Liquidborne Particle Counting using Light Obscuration and Light Scattering Methods
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.