1
Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton
2
Test Confirm option is activated Press RECORD for 1.0 - 2.0 seconds DIAG should appear on screen.
3
If DIAG does not appear... YOU MUST LOAD THE OPTION
4
To activate diagnostic quality option... Locate LP 10 programming key
5
Place battery into slot one. Turn power off Insert LP 10 programming key into AUX port
6
Turn Power Switch to BATT 1 & Remove key
7
Verify o01 appears on status display Press ECG SIZE to change option status to o01
8
Verify 1 appears in upper right screen Press QRS VOL until # 1 appears in upper right screen
9
Set Option Press CAL, BEEP will sound. Option is now loaded
10
Confirm option is activated Press RECORD for 1.0 - 2.0 seconds DIAG should appear on screen.
11
Obtaining the Diagnostic ECG Go directly to these steps if DIAG option is already activated on your LP10
12
Place Electrodes
13
Place Lead Wires
14
Select Lead 1
15
Select DIAG option This activates the recorder Press RECORD for 1.0 - 2.0 seconds DIAG should appear on screen.
16
Scroll through leads 1,2,3 (recorder will be on)
17
Watch for baseline to stabilize Obtain one full cardiac cycle for each lead
18
Leave switch in lead 3 position Manually obtain V-leads
19
Move from V-1 through V-6
20
Watch for baseline to stabilize Obtain one full cardiac cycle for each lead
21
Stop recorder when complete
22
Mark leads with pen
23
Cut and Paste ECG
24
Diagnostic 9 Lead Advantages Can be done with existing equipment Provides adequate diagnostic picture for AMI Requires no special equipment
25
Diagnostic 9 Lead Disadvantages Does not produce a “standard format” ECG Take more time than conventional 12-lead Harder to mount and label Often rejected by ED staff Lacks leads aVR, aVL, and aVF No computer interpretation