MUPUS Team Meeting, Graz, 24-25 Oct-2013 A. Schaffer, G. Kargl, N. Kömle Space Research Institute Austrian Academy of Sciences Calibration strategy for the PEN Flight Spare
2
Infrared images VarioCam HR research 600 Resolution: 640x480 IR-pixel Accuracy: ± 0.8°C Distance to probe: 1.5m Env. temp.: 24 to 25°C Rel. humidity: 55 – 60% Trigger interval: 30sec Captured: approx. 1200 images Attenuation coeff.: 0.85 (VarioCam)
3
Infrared images Sensor 12 Powerlevel 30All sensors Powerlevel 30 Time
4
Infrared images Max. T and IR-pic Sensor 10 Example: Sensor 10 max. temperature graph and corresponding IR-picture „Thermal middle“ or „Hot spot“ – region determined for each sensor
5
Usefull for: model purposes layer detection graphical representation Conclusion Infrared images
6
Heat sink
7
Aluminium Dim.: 425 x 60mm Cooperation MFL Liezen Cut with water jet Copper cooling tubes Heat sink
10
Assembled condition
11
Measurement setup NEW calibration file Thermal control Refrigeration bath circulator Huber CC-902 Heat sink Agilent data acquisition 6x PT100 class A 4x PT100 class B Vaccum chamber High performance rotary vane pump DUO 20 M Small yet turbopump HiPace@ 80 Operation range: 10E-6 mbar EXTASE system SON10#MP02 EXTASE data logger
12
Measurement setup
15
Planned temperature profile, can be programmed in CC-902 Constant temperature level kept for 1h, except 0°C for 2h Av. change of temperature in ramps : ±0.35 K/min ±0.02 K/min
16
Measured data First measurement series Date: 10.10.2013 to 11.10.2013 Range: +30°C to -50°C in 3 stages (CC-902 limitations) Chamber pressure: air pressure (≈1013 mbar) no heating of the probe REFPT100: mean of HSE1 to 5 (ΔT ≅ 0.6°C @0°C)
17
Calibration approach Mathematical background
18
Measured data
19
Sensor 5: R 0 = 73.74 Ω
20
Measured data
24
Comparison of OLD and NEW EXTASE calibration data
25
Measured data
28
Second measurement series - VACCUM conditions Date: 15.10.2013 to 16.10.2013 Range: +30°C to -50°C in 3 stages (CC-902 limitations) Chamber pressure: High vaccum: 2.6e-6mbar to 7.0e-6mbar No heating of the probe REFPT100: mean of HSE1 to 5 (ΔT ≅ 0.6°C @0°C)
29
Measured data
32
VACUUM - Comparison of OLD and NEW EXTASE data
33
Measured data
36
Unique calibration file for vacuum conditions Correction factor for sensors 1 to 3 Conductivity test measurements (air pressure and HV) Temperature gradient measurement to get information about the conductivity of the probe Heating in high vacuum with no sample Measurements at temperatures below -50°C Softwareupgrade for CC-902 Outlook and conclusion Measured data