Presentation is loading. Please wait.

Presentation is loading. Please wait.

Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph.

Similar presentations


Presentation on theme: "Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph."— Presentation transcript:

1 Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph

2 Vitamin C (Ascorbic Acid) formula weight 176.13g Water Soluble melting point 190-192° C, unstable in solutions containing oxygen as it is readily oxidized

3 Why is Vitamin C important? It is involved in over 300 biological processes in the body for example  Antioxidant  Collagen manufacture & neurotransmitters  Immune function It is a Nutrient not synthesized by the body Complete deficiency results in scurvy Deficiency symptoms  Bleeding gums  Joint pain  Bruising & wounds that don’t heal  Fatigue & muscular weaknesses

4 Recommended Daily intake is 30-60mg The RDI is what you need to consume to prevent a deficiency 100ml (which is less than ½ a cup) of the Popper Juice orange Juice is all you would need to consume excess vitamin C is usually eliminated through the urine An excess of Vitamin C can cause diarrhea as water is pulled into the colon from the circulatory system

5 Fruit Juices Expected concentrations of Vitamin C in Fruit Juices

6 Samples Popper Juice Apple and Blackcurrant Popper Juice Orange Juice Popper Juice Apple Juice Popper Juice Apple And Mango Juice WESTCLIFF Apple Juice And V8 Vegetable Juice

7 Metrohm Polarograph Photographed by Shay

8 Procedures Differential Pulse Polarography Standard Addition Method Dilute sample accurately to approximate concentration of 40mg/L Place in sample cell with water and buffer solution Run scan -0.4 to +0.1 V 3 standard additions Normal Calibration Method Prepare 0, 5, 10, 25mg/L standard solutions in buffer solution Dilute sample accurately to an approximate concentration of 10mg/L in buffer Runs scans as above

9 Differential Pulse Polarography Standard Addition Results SampleWESTCLIFF Apple Juice200mg/L standard Solution of Ascorbic acid Average concentration 263 mg/L235 mg/L Recovery Check 174.1 % recovery Standard Deviation 68%51% Relative Error 17%

10 Differential Pulse Polarography Normal Calibration Results SampleAverage concentration (mg/L) Recovery Check (% recovery) Standard Deviation WESTCLIFF Apple Juice2731176% JUST JUICE Apple Juice17491- POPPER JUICE – Orange Juice811108- POPPER JUICE – Apple and Blackcurrant Juice 633-- POPPER JUICE – Apple and Mango Juice 731-- POPPER JUICE – Apple Juice662-- V8 – Vegetable Juice332155-

11

12 Concentration Range of Vitamin C

13

14

15 Choice of method normal calibration  simpler for sample delivery (no pipetting required)  more consistent results  better RC values

16 What went wrong Standard Addition Method Lack of precision in the results obtained Caused by Micropipettes Human error 3 aliquots to be transferred to sample cell per run

17 What went wrong Normal Calibration Method  An aliquot of the V8 vegetable juice could not be taken by pipette as the juice was so viscous  Had to weigh a mass and do a density calculation before diluting the juice JUST JUICE – Apple Juice concentration was outside the concentration range of the standards used in the calibration graph Easily fixed – just recalibrate

18 Polarography Disadvantages Not time efficient Must use a buffer solution

19 Conclusion The normal Calibration polarographic method for the analysis of vitamin C is better than the standard addition method All sample analyzed had considerably more vitamin C than specified on the packet except Just Juice apple juice which had about a 1/5


Download ppt "Rebekah Livesley, 2008 Analysis of Vitamin C in Fruit Juices Instrument Metrohm Polarograph."

Similar presentations


Ads by Google