Presentation on theme: "Transmission Electron Microscopy"— Presentation transcript:

1 Transmission Electron Microscopy
David Stokes 2DX Workshop University of Washington 8/15-8/19/2011

2 Kinds of Electron Microscopes
TEM transmissive resolution determined by optics analogous to bright field light microscope detector is film or CCD SEM reflective, surface imaging resolution determined by spot size x-ray microanalysis or EELS possible STEM transmissive, small spot with scan coils analogous to confocal detector is PMT

3 SEM vs. TEM

4 TEM sectioned cell

5 TEM metal shadowed molecules
250 nm

6 TEM of frozen, isolated macromolecules
Number of particles per micrograph and resolution needed determines magnification.

7 TEM of 2D membrane protein crystal

8 TEM images are 2D projections
3D Particle projected on 2D detector in TEM. Film, Phosphor screen or CCD.

9

10 Source of electrons Tungsten:
- outer electrons of W are made free by heating the metal. - W has excellent yield when heated just below its melting temp of 3653K. - Heat to 2600K for good yield without melting and evaporating tip (saturation point) - 100 hrs with W wire. - Make filament (cathode) at neg potential (e.g kV). - As electrons boil off the W, they are repelled by the cathode. - Potential difference relative to the anode provides the accelerating voltage - Make Weihnolt several hundred volts more negative electrons pass through the aperture in the Weihnolt cap. bias controls output of electrons and shapes field - Anode is grounded so electrons are accelerated towards it. - Electrostatic field causes crossover at the anode - this is effective source size LaB6 - has lower work function: amount of energy necessary to free electrons. - 1000K with higher electron yield and longer life. - get small beam crossover size with high flux (10x emission of W) Field emission - single crystal of W oriented relative its xtal lattice - no heating - series of high voltage anodes draw electrons out of lattice - electrostatic lens of anodes -> 10 nm source size - need very high vacuum -> contamination - Zirconium to reduce work function, heating to reduce contamination

11 from Reimer

12 Properties of Electrons
Energy of electrons is work W required to move electron from anode to cathode against force F=-eE: U=accelerating voltage charge on electron is x10-19 Coulombs, when accelerated through 1V potential difference has kinetic energy of 1.602x10-19 Nm (joule) define this kinetic energy as 1 eV. With accelerating voltage of 200kV – electrons have energy of 200keV (= 3.2x10-14 joules) compare to C-C bond of 5.8x10-19 joules (3.6 eV) or lattice binding energy of 1-2 eV Speed of electrons at 100kV: v=1.64x108m/sec  more than 1/2 speed of light Have to consider relativity 1MeV = 2Eo  m=3m0 Wavelengths of electrons () Å at 120kV, 0.025 Å at 200 kV, Å at 300 kV compare with wavelength of light: Å or X-rays: 1.5 Å for Cu K 1 Å synchrotron radiation diffraction limit for resolution of any optical system: D = /2

13 Force on an electron electrostatic: electromagnetic:

14 Typical electron lenses
single lens (e.g. condensor) split lens (e.g. objective)

15 Magnetic lenses like Glass lenses
are governed by Newton’s lens equation and

16 Focal length of Electromagnetic lens
f = focal length K = lens const, U = voltage, i = current electromagnetic lens: can change focal length by changing current glass lens: change focus by moving specimen up and down change magnification by switching lenses

17 Lens Strength affects focal length
f = focal length K = lens const, U = voltage, i = current A stronger lens demagnifies the image!!!

18 Trajectory of electrons through a homogeneous B field
where  is angle between v and B Energy of electron conserved – no change in |v| If force is  to v: electron moves in a circle If force is || to v, path unchanged. Generally, path is spiral, with v|| and v components So electrons passing through given point P will intersect again at point P' PP' = v||Tc = vTccos, where Tc is time to complete circle, which is independent of  P' P If many angles focused in same plane, get image formed For small angles of , cos ~1 and get image formed at P'. Larger angles of  reduces the distance to P' (cos = (1 - 2/ ) and thus gives rise to spherical aberration

19 Real lens is non-homogeneous field - so-called bell-shaped field
Aberrations: astigmatism, coma, spherical, field curvature, distortion (pincushion/barrel) chromatic and Contrast Transfer Function

20 TEM lens configuration
Condenser Lenses: Demagnifies source Define “spot size” Objective lens: Defines image focus Has fixed magnification (20-50x) Intermediate lens: Controls whether Image or diffraction pattern Is recorded Projector lenses: Controls image magnification or Diffraction “camera length”

21 Condenser lenses C1 lens produces image of source
C2 lens demagnifies onto sample C2 aperture reduces beam current and makes beam more parallel

22 C1 lens controls spot size
Strong C1 lens produces small probe size and weak beam Weak C1 lens produces larger probe size and bright beam This is known as “spot size” C1 crossover produces image of source and Mag = u/v for C2 lens

23 Beam convergence Increasing strength of C2 lens ~parallel convergent
Upper Objective lens ~parallel convergent divergent parallel

24 magnification of objective: 20-50x
angle for 2.5 Å resolution at 200kv:  = /d = 0.025/2.5 = 10 mrad = 0.5° most of mag range achieved with projection lenses Much smaller angles and aberrations therefore less important

25 Diffraction vs. Imaging controlled by Intermediate lens
Back focal plane of objective lens corresponds to diff pattern Selected area aperture is at first image plane Intermediate lens selects either BFP or 1st image plane as its object Diff pattern or 2nd image serves as object for projector lens

26 Coma For high resolution images, you want the apparent source aligned with the real source on the optical axis and the beam to run parallel to this axis. This is coma-free alignment.

27 Deflectors electrostatic electromagnetic
Deflectors produced by transverse fields: E field (parallel plate capacitor) B field (electromagnetic lens) electrostatic electromagnetic

28 deflector coils generally come in pairs
gun tilt coils pivot points aligned pivot points misaligned

29 Apertures source condensor lens condensor aperture specimen

30 A typical scattering event
Proportionality constant for scattering: is scattering cross-section for one atom (cm2/atom) elastic, inelastic, absorption, fission

31 Electron Scattering Elastic Scattering: Rutherford scattering
+ R z b F F electrostatic force: Z = atomic number, but, actually need to consider screening of electrons on nuclear charge Solving Scrodinger’s equation for the potential and making approximations for screening of nuclear charge by electron cloud: N.B. Elastically scattered electrons have same wavelength as incident beam and are most effective for generating phase contrast image because they interfere with the unscattered beam

32 Inelastic Scattering Energy transferred from incident electrons to the specimen Oscillations of molecular bonds and phonon excitations: 20 meV – 1 eV too small to generally observe given ~1 eV spread of beam energy Excitation of outer electrons and valence/conduction band electrons (metals): 1-50 eV 3. Ionization of inner electrons (K,L,M shells) to unoccupied shells: 10 – 100s eV

33 Why do we hate inelastic collisions?
Inelastic electrons are focused at “undesirable” places And therefore produce a blurred image due to Chromatic aberration Inelastic scattering produces chemical changes radiation damage

34 differential cross-section total scattering cross-section
scattered outside 2.4 A aperture unscattered inelastics within aperture differential scattering vs.  25 keV electrons with Ar gas elastics within aperture

35 The inelastic scattering “blurs” the low spatial frequencies
electrons X-rays

36 But they can be filtered out (after they damaged the specimen)