1
Community Preparation for Caring for Mechanical Circulatory Device Patients University of Wisconsin Hospital And Clinics Ventricular Assist Device Program
2
Mechanical Circulatory Support Device Overview MCDSs are devices to support the failing heart (one side or both) Most MCDS patients are anticoagulated with Coumadin and Aspirin All MCDSs have: – Pump (implanted inside the patient’s chest) – Computer to control the settings/ display VAD performance readings – Power source: batteries or console connected to AC power
3
HeartMate II Components Driveline (tube connecting pump to outside components ) with sterile dressing covering it Republished with permission of Thoratec Corporation Pump implanted inside the chest One of two batteries needed to power the pump Controller- computer runs device; lights alarms Console used to obtain power from AC outlet
4
HeartMate II VAD Readings 3 Readings: Flow= cardiac output from the pump – Target varies by patient Speed= how fast pump is running rpm - range 8000-10000 rpm Power= how much watts it takes to run the pump - normal is < 8 watts
5
HeartMate II Care Overview Cannot palpate a pulse Pulse oximeter may/ may not work Need to assess if pump is running: listen over the point of maximal impulse (PMI) area for a consistent mechanical hum Blood pressure: need a Doppler and sphygmomanometer – result is considered a MAP – acceptable MAP 60-85 mmHg
6
HeartMate II Care Overview EKG conduction- – does not impact VAD function unless RV fails – VAD will continue to receive blood supply until RV fails or pulmonary hypertension – Most of these patients have an ICD/ pacer. Some patients’ ICDs are off to avoid inappropriate shocks. – Most VAD patients tolerate the arrhythmias well and may have little symptoms for period of time. Assess for patient’s tolerance of situation (if able). Ask for symptoms: lightheadedness, shortness of breath, palpitations, bloating, pain (anywhere), fever, chills, diaphoresis
7
Test System button Battery Fuel Gauge Battery Alarm light (yellow and red) Power Symbol (green light lit when pump running) Red Heart Symbol lit: ? Pump running or not. Check PMI for mechanical hum. Controller Cell Symbol (yellow) Silence Alarm Button HeartMate II Controller Face
8
HeartMate II Newer Controller Face Plate
9
EMERGENCY CARE Call implant hospital: 608-263-6400 and ask for HEART FAILURE ATTENDING ON CALL Still need to follow Airway Breathing of ABC’s √ VAD stopped (listen over PMI area for consistent mechanical hum). If stopped significant risk for clot inside MCDS and thromboembolize. NO CHEST COMPRESSIONS Defibrillation- STANDARD METHOD Heparin bolus if VAD stopped (contact implant hospital first)
10
Contacts 608-263-6400 (Paging Operator) Heart Failure Attending Ventricular Assist Device Coordinator (non- emergency)