LNLS Experimental facilities 15 beam lines in operation : X-Ray diffraction (XRD) Crystallography (MX) Small Angle X-Ray Scattering (SAXS) X-Ray Absorption.

Similar presentations


Presentation on theme: "LNLS Experimental facilities 15 beam lines in operation : X-Ray diffraction (XRD) Crystallography (MX) Small Angle X-Ray Scattering (SAXS) X-Ray Absorption."— Presentation transcript:

1

2

3 LNLS Experimental facilities 15 beam lines in operation : X-Ray diffraction (XRD) Crystallography (MX) Small Angle X-Ray Scattering (SAXS) X-Ray Absorption and Fluorescence (XAFS) For our measurements we use 3 beam lines : SAXS1 (Pinhole experiment) Energy Resolution 10% (8 +-0.8 keV) 1E12 photons/s Divergence 900 x32 urad @ 8keV XAFS1 (Energy scan for single pixel calibration) Energy range 4-15 keV Resolution : 0.0001% Beam Size 1x1 mm2 1e9 photons /s @ 7 keV MX1 (Pinhole and Fluorescence calibration) Energy = 7.7keV + harmonics 1e12 photons /s Various targets for calibration

4 Fluorescence measurements (MX1) TargetCoCrCuFeMnNiTiVZn E (kα) in keV 4.514.955.4145.896.46.937.478.048.63 X-Ray beam (7.9,15.8 keV) 1e12 photons/ s Target Fluorescent photons field ~ 1000 single pixel clusters per pixel / 50k frames (1ms integration) -> ~30-45 min per fluo Only tested with one assembly (A06-W0110) but possibility to run multiple assemblies at the same time (Large Fluorescence radiation field ) Timepix+FITPix

5 Single Pixel TOT Spectrum First Harmonic scattered

6 Calibration Curve

7 Single Pixel Energy Spectrum First Harmonic scattered

8 Application to testbeam data Old CalibrationNew Calibration By Sophie Redford

9 Application to testbeam data Old CalibrationNew Calibration By Sophie Redford

10 Impact on resolution Resolution in X, 2 hit cluster: 5.124 um -> 4.893 um Resolution in Y, 2 hit cluster: 4.986 um -> 5.265 um Resolution in X 4 hit cluster: 6.269 um -> 6.166 um Resolution in Y 4 hit cluster: 6.211 um -> 6.481 um

11 Energy Measurements with monochromator (XAFS1) X-Ray incident beam goes to Crystal Monochromator and is then delivered to the experimental area. Energy can be selected with 0.1keV accuracy along the 5-15 keV range. 2 sensors (C04-W0110,D09-W0126) were tested in this beamline to obtain single pixel calibration for a subset of pixel (beam 2.5x3mm2) Analysis on one of the sample will be presented by Mateus Vincente for UFRJ

12 Single Pixel TOT Spectrum I cant see the 5-8 keV photon peak, Double count peaks are however visible

13 Global Surrogate Fit

14

15

16


Download ppt "LNLS Experimental facilities 15 beam lines in operation : X-Ray diffraction (XRD) Crystallography (MX) Small Angle X-Ray Scattering (SAXS) X-Ray Absorption."