1 AMBIENT Mass Spectrometry

2 Ion Sources in the Mass Spectrometer - Analyzed samples are ionized prior to the analysis in the mass spectrometer - A variety of IONIZATION techniques are used for Mass spectrometry - Electron Ionization (EI) & Chemical Ionization (CI) are suitable for gas-phase ionization - EI and CI use is therefore limited to compounds sufficiently volatile and thermally stable

3 Ion Sources in the Mass Spectrometer - A large number of compounds are thermally labile or do not have sufficient vapor pressure - Therefore, ions of these compounds should be extracted directly from the condensed (liquid, solid) phase to the gas phase

4 Liquid and Solid Phase Direct Ion Sources - Direct ion sources are divided to: 1. Liquid-phase ion sources Analyte is in solution 2. Solid-state ion sources Analyte is an involatile deposit

5 Liquid Phase Direct Ion Sources - Analyte is in solution - The solution is introduced by nebulization, as droplets into the mass spectrometer through vacuum pumping - Electrospray (ESI), thermo spray and Atmospheric Pressure Chemical Ionization (APCI)

6 Solid Phase Direct Ion Sources - Analyte is an involatile deposit - Analyte obtained by methods that introduce a matrix that can be either a solid or a viscous fluid - The deposit is irradiated by energetic particles or photons that desorb and ionize ions near the surface of the deposit - Such approaches are termed desorption/ionization methods (DI)

7 Desorption Removal of a substance from a surface on which it is adsorbed

8 Solid Phase Direct Ion Sources - The ions are extracted by electric field and focused towards the mass analyzer condensed phase-to-gaseous ion transfer - Examples: Matrix-assisted laser desorption ionization (MALDI), secondary ion mass spectrometry, field desorption, plasma desorption, fast atom bombardment ionization methods all use this strategy to produce ions

9 Generating Ions in Vacuum or in Air - Vacuum conditions are a simple choice for creating and maintaining ions - Vacuum environment is not absolutely necessary - Ions can also be generated in air

10 AIR/Ambient Ionization - A recent innovation of mass spectrometry - The ability to record mass spectra on ordinary samples, in their native environment - No sample preparation or pre-separation by creating ions outside the mass analyzer instrument

11 AIR/Ambient Ionization A few techniques developed recently that allow ions to be created under ambient conditions and then collected and analyzed by MS

12 AIR/Ambient Ionization Ambient ionization methods retain the characteristic advantages of MS: - Speed - Chemical specificity - Low detection limits - Applicability to complex mixtures through MS-MS

13 AIR/Ambient Ionization All the important characteristics of MS described before are used in ambient MS and do not require sample preparation

14 AIR/Ambient Ionization Applications in for example: - Explosives on luggage - Drug metabolites in urine - Lipids in intact tissue - Active ingredients in pharmaceutical tablets

15 AIR/Ambient Ionization - A few techniques developed recently that allow ions to be created under ambient conditions and then collected and analyzed by MS The main one is DESI

16 - A fine spray of charged droplets hits the surface of interest (ESI-like) - It picks-up small molecules and large biomolecules and ionizes them -De-solvated ions are delivered into the mass spectrometer Desorption ElectroSpray Ionization (DESI)

17 e.g. PTFE

18 Cotte-Rodriguez et al., Anal. Chem. 2005 Desorption ElectroSpray Ionization (DESI)

19 - DESI applies to both large and small molecules - The charged micro droplets (used as projectiles) pick up proteins and other large biomolecules from the surface ionize them and transport them to the MS - The process results in ESI like mass spectra of proteins Desorption ElectroSpray Ionization (DESI)

20 Gas phase solvent ions in the spray protonate or react with analyte small molecules on the surface Desorption ElectroSpray Ionization (DESI)

21 This results in the generation of ions from compounds with low desorption energies such as: 1. Volatile and semi-volatile compounds (aromatic hydrocarbons) 2. Low polarity molecules of small size (e.g. terpenes, lipids) 3. Low molecular weight polar compounds (amino acids) Desorption ElectroSpray Ionization (DESI)

22 In the MS: - Gas phase ions characterized by m/z ratio - By MS/MS spectra - Soft ionization (minimum fragmentation), intact molecular ions Desorption ElectroSpray Ionization (DESI)

23 - Quantitative accuracy is limited by matrix effect (vary with analyte) - Internal standard can make it better but it cannot be used for solid samples - Detection limits are very low (10 -15, femtomole) - Peak stability can be more than 20% RSD

24 Desorption ElectroSpray Ionization (DESI) Takats et al., Journal of Mass Spec. 2005

25 Desorption ElectroSpray Ionization (DESI) Takats et al., Journal of Mass Spec. 2005

26 Desorption ElectroSpray Ionization (DESI) Takats et al., Journal of Mass Spec. 2005 Dependence of signal intensity on surface temperature and surface potential (melittin is a small peptide)

27 Desorption ElectroSpray Ionization (DESI) Takats et al., Journal of Mass Spec. 2005 Dependence of signal intensity on SPRAY IMPACT ANGLE, SPRAY VOLTAGE, GAS INLET PRESSURE & FLOW RATE

28 Takats et al., Journal of Mass Spec. 2005 Signal intensity vs. the amount of peptides deposited Pure peptides 5 orders of magnitude Peptides mixture (tryptic digest), much less dynamic range and ion suppression effect Dynamic Range of DESI

29 Takats et al., Journal of Mass Spec. 2005 Detection of Different Metabolites with DESI

30 Reactive DESI - Adding reagents to the DESIO spray to cause interface reactions of the reagents with the condensed phase analytes - Thus, adding chemical selectivity - Examples: 1. Redox reactions of metal complexes 2. Covalent modifications of particular functional groups 3. Formation of non-covalent complexes

31 Applications/examples of DESI

32 Example of a DESI Application - DESI of a dry urine spot (10µl) on a paper - No sample preparation, just deposition on surface - Spray with Methanol/water/acetic acid - Rate of one analysis per second

33 Example of a DESI Application Human Urine sample

34 Example of a DESI Application

35 Detection of Explosives with DESI - RDX explosive - 1ng RDX on mm 2 - Desorbed from an insulating tanned leather surface - Negative ion DESI - Acetonitile/methanol/TFA spray solvent

36 Detection of a Chemical Warfare Agent Simulant with DESI - The analyte- DMMP - Positive ion DESI - Nitrile gloves that were exposed for less than 1 sec. to DMMP vapor and then washed and dried

37 Detection of Dried E.Coli Culture with DESI DESI mass spectrum of a dried sample of E.coli cells In the future: bacteria on skin, food stuffs and biological fluids will be measured by these type of methods

38 Spatial Analysis of Native Surfaces with DESI (Plant or animal tissues) - Seed section of Conium - Positive ion DESI - M/Z/ 126 corresponded to the protonated alkaloid Coniceine

39 Spatial Analysis of Native Surfaces with DESI (Plant or animal tissues) Distribution of the alkaloid Coniceine across a stem cross section

40 Spatial Analysis of Native Surfaces with DESI (Plant or animal tissues) Tomato fruit skin analysis

41 Spatial Analysis of Native Surfaces with DESI (Plant or animal tissues) Tomato fruit skin analysis

42 In Vivo Sampling of Living Tissues Surfaces with DESI - Person taking 10mg of over-the-counter the antihistamine Loratadine - 40 min after taking the tablet, molecule became detected in Saliva or skin - Surface concentration remained above the detection limit for another 50 min.

43 In Vivo Sampling of Living Tissues Surfaces with DESI Direct Tissue Profiling of Liver Tumor Tissue Cooks et al., Science 2006

44 In Vivo Sampling of Living Tissues Surfaces with DESI

45 DESI-MS/MS Data for Main Ingredients in Tablets Chen et al., Anal. Chem. 2005

46 HIGH THROUGHPUT DESI-MS/MS Data (Claritin Tablets) Chen et al., Anal. Chem. 2005 Speed of 0.76 sample/s

47 DESI FINGERPRINT

48 DESI of Flower Tissue Detected essential oils, carotenoids and antioxidants Hibiscus flower

49 Other Ambient MS Methods DART: Direct Analysis in Real Time (different projectiles) DAPCI: Desorption Atmospheric Pressure Chemical Ionization (different projectiles) ELDI: Electrospray-assisted Laser Desorption/Ionization (uses laser for desorption)

50 The End