Presentation is loading. Please wait.

Presentation is loading. Please wait.

The X-ray Imaging System

Similar presentations


Presentation on theme: "The X-ray Imaging System"— Presentation transcript:

1 The X-ray Imaging System
Week 4-5

2 Bucky slot cover During fluoroscopy the Bucky tray is moved to the end of the table This leaves an opening in the side of the table about 5cm Approximately at what level is the bucky?

3 Bucky slot cover The opening should automatically be
covered with at least 0.25 mm Pb equiv..

4 Fluoroscopy

5 The Control Console The control console is device that allows the technologist to set technical factors (mAs & kVp) and to make an exposure. Only a legally licensed individual is authorized to energize the console.

6 Control Panel All the electric circuits connecting the meters and controls are at low voltage to minimize the possibility of shock.

7 Operating Console Controls:
Line Compensation, kVp, mA and time Quantity = # of x-rays Milliroentges (mR) or (mR/mAs) Quality = the pentrability Kilovolts peak (kVp)

8 Operating Console has meters to measure
kVp, mA, & exposure time Modern units only display mAs Units with ACE’s will have a separate meter for mAs

9 Control Panel

10 AEC Automatic Exposure Control Uses an ionization chamber
Technologist sets kVp, mA, back-up time & sensors Exposure terminates the IR has proper OD Patient positioning must be absolutely accurate

11 AEC Sensors

12 APR Anatomically Programmed Radiography (Ch 15)
Radiologic Technologist selects on the console a picture or a written description of the anatomic part to be imaged and the patient body habitus A computer selects the appropriate kVp and mAs.

13 APR The whole process uses an AEC
Precise patient positioning over the phototiming sensor is critical

14 APR

15 X-RAY CIRCUITY Contributions by Mosby, Thompson Publisher, Carlton, Bushberg, and the WWW.

16 3 Divisions of Circuit Board
PRIMARY (CONTROL PANEL) yellow SECONDARY (HIGH VOLTAGE) blue FILAMENT (LOW CURRENT) purple

17

18

19 Functional Position Control Console Transformers Tube

20 Line Compensation Most imaging systems are designed to operate on 220 V. (some 110 V or 440 V) However power from the wall is not always accurate continuously

21 Line Compensation Wired to the autotransformer is the line compensator
Designed to maintain the accurate voltage required for consistent production of high-quality images Today’s line compensators are automatic and are not displayed on the control panel

22 Line Compensator

23 Autotransformer The power for the x-ray imaging system is delivered first to the autotransformer The autotransformer works on the principle of electromagnetic induction It has one winding and one core There are a number of connections along its length

24 Autotransformer A’s = primary connections & power into the transformer
Other connections allow for variations of voltages

25 Autotransformer Is designed to step up voltage to about twice the input voltage value The increase in voltage is directly related to the number of turns

26 kVp selection

27 X-ray tube current or Filament circuit
A separate circuit crossing from cathode to anode Measured in milliampers (mA) What determines how many x-rays are created?

28 X-ray tube current or Filament circuit
# of e- is determined by the temperature of the filament. The hotter the filament the more e- Are their any limiting factors to thermionic emission?

29

30 mA selection

31 Filaments Operate at currents of 3 to 6 amperes (A)

32 What is directly proportional to the number of x-rays reaching the IR?
Question? What is directly proportional to the number of x-rays reaching the IR?

33 Exposure Timers The timer circuit is separate from the other main circuits of the imaging system It is a mechanical or electronic device whose action is to “make” and “break” the high voltage across the x-ray tube This is done on the primary side of the high voltage transformer.

34 mAs Timers Monitors the product of mA and exposure time
Terminates the exposure when the desired mAs value is reached Located on the secondary side of the high-voltage transformer since actual tube current must be monitored

35 mAs Timers Designed to proved the highest mA for the shortest exposure
Modern X-ray machines have falling-load generator Automatically adjusts to the highest mA at the shortest exposure time possible

36 AEC Control AEC measure the quantity of radiation reaching the IR
Automatically terminates when the IR has received enough radiation for desired OD Two types are common

37 Flat, parallel plate ionization chamber
Located between the patient and the IR Made radiolucent Ionization w/in the chamber creates a charge; calibrated to produce a given OD on the IR

38 Photomultiplier (Photodiode) detector assembly
Located behind the IR Contains a fluorescent screen and a photomultiplier The photomultiplier detects the light from the fluorescent screen until the desired OD on the IR is reached terminating the exposure

39

40 AEC’s Upon instillation must be calibrated by the service engineer
Technologists selects the desired OD which then sets the mA & kVp

41 AEC’s A back up timer usually automatically set to prevent over exposure if the AEC fails Should be set to 1.5 times the expected exposure time Why? When the ionization chamber or photodiode reaches the preset level, a signal is returned to the operating console to terminate the exposure

42 High-Voltage Generator
Responsible for increasing the output voltage from the autotransformer to the kVp necessary for x-ray production 3 parts: High-voltage transformer (step-up), filament transformer (step-down) and rectifiers

43 High voltage transformer
Or step up transformer Connected to the Major and Minor kVp selector Increases the volts from the autotransformer to kilovolts

44 Step Up Transformer

45 Voltage Rectification
Converts AC to DC current During the negative cycle current can only flow from anode to cathode E- must travel cathode to anode – DC current keeps e- traveling in the correct direction, cathode to anode Attracted to the positive anode

46 Voltage Rectification

47 X-Ray Tube Circuit

48 Filament transformer Or step down transformer
Reduces the current to the filament

49 High-Voltage Generation – converts 220 volts of AC to kilovolts of DC
The generator is a FIXED component of the imaging system, not under the control of the technologist Three basic types: single phase, three phase, and high frequency The generator affects the quality and quantity of photons produced

50

51 How does this effect technique selection?

52 Questions on imaging systems?


Download ppt "The X-ray Imaging System"

Similar presentations


Ads by Google