Download presentation
Presentation is loading. Please wait.
Published byHilda Garrison Modified over 9 years ago
1
RECTIFICATION
2
X-RAY CIRCUIT
3
Whenever the art of medicine is loved, there is also love for humanity. Hippocrates
4
RECTIFICATION PROCESS BY WHICH ALTERNATING CURRENT IS CHANGED TO PULSATING DIRECT CURRENT
5
ALTERNATING CURRENT 1 PHASE A/C ( UNRECTIFIED)- FREQUENCY 60 Hz
6
60 POSITIVE PULSES/SEC 60 NEGATIVE PULSES/SEC
7
60 POSITIVE PULSES 60 NEGATIVE PULSES 120 PULSES 60 FULL CYCLES
8
BATTERY – CURRENT OUTPUT STEADY D/C
9
D/C HALF WAVE RECTIFIED D/C HALF-WAVE – 60 POSITIVE PULSES/SEC NEGATIVE PULSES ARE SUPPRESSED 1 PULSE/CYCLE
10
D/C FULL WAVE RECTIFIED D/C HALF-WAVE – 120 POSITIVE PULSES/SEC 2 PULSES/CYCLE
11
D/C -3 PHASE 6 PULSE CURRENT D/C 3 PHASE 6 PULSE – 6 PULSES/CYCLE
12
D/C - 3 PHASE 12 PULSE D/C 3 PHASE 12 PULSE – 12 PULSES/CYCLE
13
DIODES OR VALVE TUBES HELP RECTIFY A/C CURRENT DIODE ELECTRONS VALVE TUBE
15
X-RAY TUBE FORWARD BIAS
16
REVERSE BIAS + - NO X-RAY
17
FORWARD vs REVERSE
18
RECTIFICATION TYPES ► HALF-WAVE ► FULL WAVE ► THREE PHASE SIX PULSE ► THREE PHASE TWELVE PULSE ► HIGH FREQUENCY SINGLE PHASE
19
RECTIFICATION NO RECTIFICATION HALF-WAVE FULL-WAVE
20
3 PHASE RECTIFICATION
22
HIGH FREQUENCY RECTIFICATION
23
HALF-WAVE RECTIFICATION
25
FULL-WAVE RECTIFICATION
28
3 PHASE 6 PULSE
29
3 PHASE 12 PULSE
30
HIGH FREQUENCY
31
VOLTAGE RIPPLE ► HALF WAVE------------------- 100% ► FULL WAVE ------------------ 100% ► 3 PHASE 6 PULSE ----------- 13% ► 3 PHASE 12 PULSE ---------- 4% ► HIGH FREQUENCY ---------- <1%
32
RIPPLE 100%
33
RIPPLE-3 PHASE 13%
34
RIPPLE-HIGH FREQUENCY <1%
35
Home-made x-ray apparatus
36
CHECKING FAILURE OF RECTIFICATION SYSTEM ► SPINNING TOP TEST – 1 PHASE EQUIPMENT ► SYNCHRONOUS SPINNING TOP TEST- 3 PHASE WITH LONGER EXPOSURE TIMES ► OSCILLOSCOPE- 3 PHASE EQUIPMENT WITH SHORT EXPOSURE TIMES
37
SPINNING TOP TEST
39
RESULT OF SPINNING TOP TEST- 1 PHASE - DASHES
40
HALF- WAVE RECTIFICATION # OF DASHES = TIME (SEC) x 60 PULSES
41
FULL- WAVE RECTIFICATION # OF DASHES = TIME (SEC) x 120 PULSES
42
RESULT OF SYNCHRONOUS SPINNING TOP TEST-3PHASE ARC
43
3 PHASE RECTIFICATION- LONGER EXPOSURE TIME DEG. OF ARC = 360°/ sec x TIME (sec)
44
UTILIZING OSCILLOSCOPE
45
UTILIZING OSCILLOSCOPE RECTIFIED SYSTEM