An Introduction to SIMS and the MiniSIMS ToF © Millbrook Instruments Limited Blackburn, UK www.millbrook-instruments.com www.minisims.com
2
Secondary Ion Mass Spectrometry (SIMS) & The Millbrook MiniSIMS
3
The SIMS Process simulation courtesy of Dr Postawa Zbigniew at the Jagiellonian University in Poland
4
A Conventional SIMS System
5
The MiniSIMS Instrument
6
Design Objectives Increase routine use of Surface Analysis –more affordable –more accessible Not a replacement for conventional SIMS –not state-of-the-art performance –restricted analysis conditions
7
Transform one of these …..
8
….. into one of these.
9
Operational Strengths Low Capital & Running Costs Fast, On-Site Analysis Compact Design, Single Electrical Supply Full Automation & Control for Ease of Use High Reliability Rapid Sample Throughput Simplified Data Interpretation Remote Control via Internet
10
Typical Application Areas Surface Coatings Surface Treatments Electronic Components Semiconductors Electrodes & Sensors Catalysts Adhesives Lubricants Packaging Materials Corrosion Studies
11
Alternatives to a MiniSIMS ? No Other Benchtop SIMS Instrument Conventional SIMS Systems –much more complex and expensive Contract Analysis Laboratories –not convenient, contamination during transport Other Surface Analysis Techniques –generally less sensitive
12
Comparison of SEM / EDS and SIMS
13
Advantages of SEM / EDS High magnification physical image Quantitative elemental information
14
Advantages of SIMS Surface specific analysis Organic structure identification Profiling for depth distribution Light element detection
15
EDX sampling depth is typically 1 micron For many applications surface sensitivity is needed… Not 10 xNot 100 x But 1000 x – SIMS can offer true surface analysis 509 m
16
Conclusions SIMS & EDS give complementary information SIMS has advantages for –Organic surface contamination –3 dimensional analysis of multi-layer structures SEM / EDS has advantages for –High magnification physical imaging –Quantitative analysis
17
MiniSIMS ToF
18
Typical Applications for the MiniSIMS ToF Analysis of unknown samples (failure analysis) Analysis of unique samples Improved analysis of organic materials Smaller area static SIMS analysis Retrospective experiments
19
MiniSIMS ToF Advantages over Quadrupole Instrument –Smaller area static SIMS analysis –Extended mass range –Higher mass resolution (organic v inorganic) –Retrospective experiments
20
MiniSIMS ToF Use of Continuous Primary Beam –Fast analysis (= low cost per sample) –No loss of image resolution in pulsing –Simplified depth profiling (single beam) Fast & simple static / imaging / dynamic SIMS in one instrument
21
Static SIMS (Surface Analysis)
22
Polycarbonate (Bisphenol–A)
23
Characterisation of Layer Structurally significant peaks
24
Characterisation of Layer Structurally significant peaks
25
Effect of Decreasing Area Analysis Area Dimension Mass Scale Quadrupole Data
26
Effect of Decreasing Area Analysis Area Dimension Mass Scale Time of Flight Data
27
Higher Mass Resolution
29
Extended Mass Range K9I8K9I8
30
ToF –ve ion mode: Irganox molecular ion at m/z = 1175 Da
31
Imaging SIMS (Spatial Analysis)
32
Surface Organic Contaminant Analysis
33
Retrospective Analysis
36
Dynamic SIMS (Depth Analysis)
37
Multi-Layer Coating (Three Dimensional Analysis)
38
Retrospective Profiling
39
Conventional Profile Sodium high at surface
40
Retrospective Cross-Section Sodium inclusion in layer Sodium high at surface
41
Retrospective Profiling
42
Retrospective Analysis Horizontal Image of inclusion
43
Summary
44
MiniSIMS Summary SIMS is a powerful technique for the 3-D analysis of the surfaces of materials and thin films Information easily available by SIMS may be difficult or impossible by any other technique SIMS is especially valuable for the detection of:- –organic species (e.g. silicones, fluorocarbons) –light elements (lithium, beryllium, boron …) –group IA & IIA metals, group VII halides
45
MiniSIMS Summary SIMS is a fast analysis technique, especially for imaging applications MiniSIMS ToF is most effective for a comparative analysis of samples Desktop MiniSIMS means SIMS is now affordable and accessible to all
46
ToF MiniSIMS ( v conventional ToFSIMS ) Use of Continuous Primary Beam –Fast analysis (= low cost per sample) –No loss of image resolution in pulsing –Simplified depth profiling (single beam) Fast & simple static / imaging / dynamic SIMS in one instrument Upgrade path from Quadrupole to TOF
47
ToF MiniSIMS ( v quadrupole MiniSIMS ) Improved Static SIMS from smaller areas Retrospective Experiment – 2D Imaging – 3D Imaging / Depth Profiling Extended Mass Range Higher Mass Resolution