Tutorial on Computational Optical Imaging University of Minnesota 19-23 September David J. Brady Duke University www.disp.duke.edu
2
Lectures 1.Computational Imaging 2.Geometric Optics and Tomography 3.Fresnel Diffraction 4.Holography 5.Lenses, Imaging and MTF 6.Wavefront Coding 7.Interferometry and the van Cittert Zernike Theorem 8.Optical coherence tomography and modal analysis 9.Spectra, coherence and polarization 10.Computational spectroscopy and imaging
3
Lecture 10. Computational Spectroscopy and Imaging Compressive Optical Montage Photography Coded Aperture Spectroscopy
4
2. Multiaperture systems for imaging system miniaturization Prototype COMP-I visible Imager
5
Image reconstruction from subpixel shifts 4 Images taken by 2 by 2 array
6
Measurement Model
7
Inference Strategy
8
Reconstruction Best raw image Synthesized image
9
Reconstruction Zooms
10
Shot of COSI meeting One of 4 low res images Reconstructed image Zoom on speaker
11
3. Compressive Sampling Conventional imaging COMP-I imaging
12
Compressive Sampling Conventional imaging COMP-I imaging
13
Pixel Block Coding
14
Quantized Cosine transform Block-based transform maintains compressive property of DCT Arithmetic complexity O(n log n) Values from {-1,0,1} amenable to mask implementations
15
8x8 DCT
16
8x8 meanQCT
17
8x8 signQCT
18
MTF and Registration Image shift coding produces fundamentally bandlimited images Focal plane coding can recover high resolution images Registration may be addressed by PSF coding
19
Characterization of sub-pixel optical PSF Approximately 4x undersampled Green: MTF from sampling grid Blue: FT of opt psf + grid Red: fit to MTF based on 3 m opt psf 1.8mm focal length lens, 3 m psf width, 5.6 m pixels Scan showing pixel-pixel progression of impulse Pixel number time
20
Demonstration of sub-pixel PSF in IR System Approximately 5x undersampled Green: MTF from sampling grid Blue: FT of ir psf + grid 25mm focal length lens, 13 m psf width, 30 m pixels
21
imagesc(horiz,[-1000 6000])imagesc(zero,[-1000 5000]) imagesc(vert,[-1000 6000])imagesc(quad,[-1000 2000]) Diffractive PSF coding for nondegenerate measurement on IR system
22
80100120 30 40 50 60 70 1.3 1.4 1.5 x 10 4 80100120 30 40 50 60 70 1.5 2 2.5 3 3.5 x 10 4 80100 120 30 40 50 60 70 1.4 1.6 1.8 x 10 4 80100120 30 40 50 60 70 1.4 1.6 1.8 x 10 4 Performance of the Fabricated Device
23
8.9 m Bandpass Filter, 650ºC Blackbody Testing of Delware diffractive coder
24
Focal Plane Coding Sony CCD: Pixel size: 5.6 micron by 5.6 micron Pixel number is: 480 by 600 Pixel Mask on CCD/CMOS Sensors
25
Experiment Setup 15 micron pin hole at ~ 2 meter away Wavelength: 0.6328 micron Lens: NA= 0.5
26
Aperture Pattern One Pitch is matched to the pixel size of 5.6 micron 4 micron2 micron3 micron1 micron
27
Scan Crossing Apertures
28
Characterization of Lumina CMOS Sensor Pixel Size: 5.2 X 5.2 micron square Number of Pixels: 1024 X 1280 Focus light on a Single Pixel
29
Pixel Scan (without mask) in X and Y direction
30
Sub-pixel Response of the CMOS Sensors (with the mask)
31
Coded aperture spectroscopy Focal plane coding microspectrometer using Hadamard codes, 0.5 nm resolution in a 5 cubic centimeter volume Coding technology and algorithms very similar to COMP-I imaging algorithms Advantages as spectrometer include very compact volume, high throughput
32
Data Recorded with Coded Aperture Mask For a single grating, system records spectrum convolved with coded mask on CCD K-fold multiplex grating records sum of K spectra convolved with coded mask on CCD
33
System Details 640x480 Firewire Webcam Relay Lenses Multiplex Hologram Coded Aperture Mask 1000 Spectral Channels between 400 – 700nm
34
Spectral reconstruction Ocean Optics Spectrum DMS Spectrum
35
DISP Tissue Spectrometer
36
Raman Spectroscopy Raman spectra uniquely identify molecules by revealing information about ro-vibrational level spacing IR Raman spectroscopy is very useful for biomedical applications because of the “diagnostic window”, an IR region in which many biological tissues have low absorption Raman signals are very weak and spatially incoherent Shown to be effective in analyzing blood components in ex vivo samples
37
Light Scattering in Tissue Satisfies diffusion equation, in the scattering- dominated limit ( κ d =1 / diffusion length) Spot sizes become very large (2-4 mm) in tissue creating difficulty in coupling light to conventional instruments T. Vo-Dinh, Biomedical photonics handbook, CRC Press, Boca Raton, Fla., 2003.
38
Simulation of Photon Distribution for 800-850 nm Light in Tissue and Blood
39
Throughput- Resolution Tradeoff Narrow slit to achieve high spectral resolution on detector Detector Plane Dispersion Element Detector Plane Dispersion Element Increasing slit width creates ambiguity on detector plane
40
Coded Aperture Multiplexing Use coded aperture to measure combinations of spectral channels on each detector, preserving resolution and increasing throughput Detector Plane Dispersion Element Use multiple rows of detector in order to measure a family of spectral combinations, leading to a well conditioned matrix inversion
41
Reasons for Coded Aperture Fully incoherent sources can not be locally increased in intensity (constant radiance theorem) D. J. Brady, “Multiplex sensors and the constant radiance theorem," Optics Letters 27(1), pp. 16-18, 2002. Turbid samples typical in biological systems Spatial filtering typically done through a slit to eliminate spatio-spectral ambiguity Source Optic “Focused” Source
42
Common Solution Structured fiber bundle to go from circular aperture to rectangular aperture Our prototype system- 36 um apertures, 4x2 mm area with 50% throughput Source can be magnified to increase angular acceptance and reduce interrogation area For 36 um slit, 4 mm 2 becomes 36 um x 111 mm High performance scientific cameras ~8 x 8 mm sensor, would require 14 cameras!!!
43
Excitation source: 8 ~808nm laser diodes providing multi-wavelength excitation for fluorescence rejection and noise reduction Mask: 64 row aperture mask based on N=32 Hadamard matrix Optical system: High-throughput, low- distortion optical system Sensor: Andor CCD camera with high QE, low noise, 16-bit digitization Software: Custom algorithms for spectral reconstruction and fluorescence rejection DISP Raman Spectrometer
44
Optical Design
45
Data Inversion Process
46
Molecular spectral model 1.a sufficient “dictionary” S of analytes spectra is calibrated 2.any mixture with concentration vector c has a Raman spectrum r
47
Challenge Estimation of ethanol concentration in the presence of multiple analytes of unknown concentrations
48
Methodology Linear multivariate calibration enables extraction of concentration from observed spectra Calculate a regression vector w, such that a weighted linear combination of the measured spectral intensities estimates the concentration Each element of w reveals the contribution of that spectral region Different w vectors may estimate the concentration of different analytes from the same spectrum
49
The problem statement Given a set of training spectra A of known ethanol concentrations c Find weights vector w such that
50
Singular Value Decomposition
51
Different algorithms Least squares Principal component regression Partial least squares Hybrid linear analysis
52
Cross validation calibration of training data Leave-one-out cross validation determines the rank of the approximation used for the LS solution
53
Ethanol Tests Sample solution –tissue-like lipid (20%) based in soybean oil –3 trials @ 4 min each –30 mW laser power (ANSI maximal permissible exposure for tissue at 808 nm) Eleven concentrations –20%, 10%, 5%, 2%, 1%, – 0.5%, 0.2%, 0.1% – 0.05%, 0.02%, 0.01% Two different sets of testing Ten single and dual laser excitations
54
LD2 cross validation
55
LD6 cross validation
56
LD8 cross validation
57
LD7 cross validation
58
2 Lasers
60
4 Lasers
62
Blood Analysis Tests –3 trials @ 4 min each –30 mW laser power (ANSI maximal permissible exposure for tissue at 808 nm) Analytes –Hemoglobin –Hematocrit –Glucose –Urea Ten single and dual laser excitations
63
Hemoglobin
64
Hematocrit
65
Glucose
66
Urea
67
New algorithmic approaches Bypass spectrum reconstruction altogether, using “economy” SVD decomposition –Raw CCD data –Random linear combinations and projections Pure ethanol spectra in HLA Non-linear estimators Classification algorithms
68
Hybrid linear analysis Let T be the dictionary with all spectra but the ethanol spectrum We would like to write where
69
Current work Data collection and manipulation –database software for access and validation –large volume of data –data must span over a period of time Consistency –correct for laser wavelength drift power variation Preprocessing strategies –scaling and normalization
70
Conclusions Coded aperture spectroscopy shows to be an effective method to measure the Raman spectra of ethanol in diffuse, fluorescent, media Such systems show promise for making systems with a combination of smaller excitation powers, shorter exposure times, and/or less costly components
71
Acknowledgements This work is supported by DARPA, AFOSR and the National Institute on Alcoholism and Alcohol Abuse Thanks to the members of DISP
73
optical imaging is a field that n also is ripe for mathematical attention