Download presentation
Presentation is loading. Please wait.
1
RAD 354 Chapt. 13 Intensifying Screens
Physical purpose: to convert x-ray photons into light photons (done at the phosphor layer). The RESULT does lower patient dose.
2
Most in use – if not ALL – are “rare earth”
Rare earth crystals include (but are NOT limited to): Gadolinium Lanthanum Yttrium
3
Other Intensifying Crystals Used
Barium lead sulfate (very early phosphor used) Calcium Tungstate
4
Desired Physical Properties of Crystals
High atomic number = high absorption (DETECTIVE QUANTUM EFFICIENCY {DQE}) Phosphor should emit a LARGE # of light photons for EACH x-ray photon – CONVERSION EFFICIENCY (CE) Color of light should match the color light the film is sensitive to – SPECTRAL MATCHING ZERO afterglow (“lag”)
5
Important Screen Terms
Luminescence – process of giving off light when stimulated Fluorescence – giving off light ONLY when stimulated Phosphorescence – continuing to give off light after stimualation Intensification factor – amount of radiation reduction WITH screens vs NO screens
6
Screen Speed Can be judged by intensification factor (IF)
Increasing speed INCREASES noise Increasing speed REDUCES spatial rresolution Increasing speed INCREASES quantum mottle (line-pair test pattern device is used to measure this)
7
Tech CONTROLABLE Screen Items
Screen attributes the tech can control: Radiation quality (kVp, grid/no grid, filters, etc.) Image processing and temperature Care of and cleaning of screens
8
Cassette Construction
Rigid, light proof protective housing for the film and screens Felt/rubber/sponge “compression” layer to assure good film-screen contact K-edge of crystals determines light spectrum
9
Screen Cleaning Compare/contrast screen cleaning solutions (home made vs commercially produced) Cotton balls vs 4 X 4’s
10
Screen – Film Contact Test
Wire mesh test for screen-film contact and proper resolution/visibility of detail
11
RAD 254 Chapt. 14 Control of Scatter
Break down into: Those that reduce patient dose and those that are geometrical in nature and those not
12
3 (primary) factors affecting scatter
Increased kVp Increased field size Increased patient thickness
13
Spatial Resolution & Contrast Resolution
Spatial resolution may be thought of as geometric in nature (F.S. size, emission spectrum, OID, SID – dealing with geometric image formation Contrast resolution – driven by scatter and other sources of “noise”
14
Scatter INCREASED filed sizes = MORE scatter – collimation is the MOST readily available and EASIEST thing to lower the amount of scatter Patient thickness also INCREASES scatter – compression may be used to help avoid this (IVP’s and mammos are examples where compression may be used)
15
Beam restricting devices limit the radiation to the patient
Aperature diaphram (size and resultant field size are a DIRECT proportion – draw the damn picture and figure the problem) Cones and cylinders – GREAT for absorbing scatter, but are circular shaped = great for improving contrast and removing scatter, BUT required MUCH MORE mAs as a result
16
Variable Aperature Diaphram
Mandated in 1974 by the Food and Drug Administration (mandate later removed) Positive Beam Limitation Device (PBL’s) Automatically collimate to the size of the cassette/receptor in the bucky and CANNOT be a BIGGER size than the cassette/receptor
17
Filtration Filtration also will DECREASE the low energy rays and LIMIT patient dose and some scatter
18
The Grid Only “FORWARD” scatter is of any benefit to the radiographic image – ALL other scatter degrades the image!
19
Scatter = LOWER Contrast
Using a grid (alternating strips of fine leaded strips with alternating radiolucent interspace material) can effectively reduce the amount of ANGLED scatter from reaching the cassette/receptor
20
Grid Terms Grid ratio = height of the lead lines divided by the interspace width Grid frequency/lines per inch = the MORE lines per inch, the more clean up Grid clean up = scatter w/o a grid vs scatter reaching the film/receptor with a grid AKA “Contrast Improvement Factor” Grid function = improved image contrast
21
Bucky Factor Refers to the AMOUNT of radiation to the patient with a grid vs W/O a grid The HIGHER the grid ratio, the HIGHER the “bucky factor” The HIGHER the kVp, the HIGHER the “bucky factor” Grid WEIGHT refers to how HEAVY the grid is – duhhhh- the MORE lead the heavier it is
22
Grid Types Parallel Crossed (cross hatch) Focused Focused crossed
23
Grid Problems Grid cut-off = short SID’s result in the vertical, parallel strips absorbing the “diverging” beam at the OUTER margins of the grid/film/receptor; MOST pronounced at SHORT SID’s Most grid problems are positioning related Uneven grid/off level grid Off centered (lateral decentering) Off focus grid Upside down, focused grid
24
Focused Grid Misalignment
Off level = grid cutoff across image; underexposed image (light OD) Off Center = ditto Off focus = CR centered to one side of the other of a focused grid Upside down grid = SEVER grid cut-off (NO density/OD) at BOTH sides of the image
25
Grid Ratio Selection 8:1 grid is the MOST widely used
5:1 grid is the most PORTABLE use grid ration Grid ratio is kVp driven Higher kVp’s warrant HIGHER grid ratios Higher grid ratios = HIGHER patient dose (more radiation needed to produce an image) As kVp increases pat MAXIUM OPTIMUM kVp, patient dose INCREASES
26
mAs – Grid Considerations
AS grid ratio INCREASES, so must mAs 5:1 = 2 X mAs 8:1 = 4 X mAs 12:1 = 5 X mAs 16:1 = 6 X mAs
27
Air Gap Technique By allowing the scatter radiation to “diffuse” in the atmosphere AFTER the patient but BEFORE the cassette/receptor, the image has HIGHER contrast, as the scatter diffuses and does NOT reach the receptor C-spine is a good example of this
28
RAD 354 Chap. 15 Radiographic Technique
Four PRIMARY exposure factors: kVp mA Time distance
29
In the next 5 minutes Write down “bullets” about what happens when on RAISES kVp
30
Memory “jerk” for grids
Write the following: 5 2 8 4 12 5 16 6
31
Now What??? 5:1 = 2X mAs 8:1 = 4 X mAs 12:1 = 5 X mAs 16:1 = 6 X mAs
32
kVp Beam Qualtiy Penatration Beam intensity HVL
Primarily responsible for quality, BUT INCREASES in kVp also make x-ray production SLIGHT more productive Penatration Beam intensity HVL Biggest exposure factor affecting CONTRAST
33
mA DIRECTLY responsible for AMOUNT of radiation produced (Quantity). As mAs is doubled, so is the number of photons produced and so is PATIENT DOSE mA stations are responsible for focal spot size selection
34
Time Exposure times should be practical and short enough to stop patient motion, but the shortest times also result in the most radiation output per unit of time – thus MORE wear and tear on the x-ray tube mAs = time X mA mAs is only measured by tube current Responsible for Optical Density (OD)
35
Distance (SID) The most “forgotten” exposure factor, but perhaps the most important Inverse Square Law Primarily effects Optical Density (OD) NO effect on quality Other distance related terms: FFD, FOD, OFD, FRD, ORD, SSD Other geometric factors (F.S. size, pt. size, part orientation to CR and receptor
36
Filtration kVp driven Inherent (.5 mm al equiv)
Added (2.0 which may also include some filtration from localizer light apparatus, etc.) in a kVp unit Total filtration : inherent + added (2.5 mm al equivalent)
37
Generators Half wave (120 cycles/sec = 60 impulses per second) – 100% ripple “self rectified” is also half wave where the X-RAY TUBE is the DIODE Full wave rectification (120 cycles per second = 120 impulses per second) – 1--% ripple 3 phase, 6 pulse = 14% ripple (33% more radiation per exposure over full wave) 3phase, 12 pulse = 4% ripple (40% more per exposure over full wave Hi frequency = <1% ripple
Similar presentations
© 2025 SlidePlayer.com. Inc.
All rights reserved.