RHOSONICS In-line Concentration Analysis With Ultrasound Presented By: Willem Van B. De Jong
3
RHOSONICS PRODUCTS
4
ULTRASONIC IN-LINE ANALYSIS MAIN ADVANGTAGES Based on wave propagation (movement of molecules) Independent of conductivity Independent of density Independent of transparency
5
RHOSONICS: MAIN FEATURES Completely non-destructive No moving parts Absolute measurement (time) In-line Very high accuracy
6
PULSE-ECHO (REFLECTION) & TRANSMISSION PRINCIPLE
7
THE RHOSONICS LIQUID ANALYZER The Model 8100 The MCS-1-cell transmission system The MP-130 measuring probe The MP-54 measuring probe
8
MODEL 8100 – BENCH TOP MODEL
9
THE FLOW-THROUGH CELL
10
MP-130-EX REFLECTION PROBE Explosion proof 2” and larger In-line Temp. compensation
11
8100 WITH MP-54 SENSOR Sanitary CIP-able For Beer & drinks Inline Temp. compensation
12
MEASUREMENT OF SOUND VELOCITY Principles Block diagram Time measurement Speed calculation Calculation of concentration from c, T
13
BLOCK DIAGRAM OF MODEL 8100
14
TIME MEASUREMENT: BY DIGITIZED WAVEFORM ANALYSIS
15
MAGNIFIED WAVEFORM
16
SPEED CALCULATION
17
THE TRANSLATION OF ULTRASOUND Calculation of concentration Sound speed curves of some liquids
18
CALCULATION OF CONCENTRATION Polynomial for % w/w, density, etc:
19
APPLIED TO HYDROCARBON LIQUIDS Sound speeds of some hydrocarbons Composition of some C10…C60 samples Sound speed curves of these samples
20
USED FOR Conversion processes In-line density SOUND SPEEDS OF HYDROCARBONS
21
CALIBRATION METHOD Gather sound speed data: Sound velocities @ 25 temperatures Sound velocities @ 3 to 5 concentrations Calculation of polynomial data (R) Enter data into model 8100 Each set is called a “recipe”
22
EXAMPLE 2: NITRIC ACID
23
EXAMPLE 3: SULPHURIC ACID
24
EXAMPLE 4: ACETIC ACID
25
EXPECTED RESOLUTION AC. ACID: Change in c per %: 9.2 m/s/% Change in c per ºC: 3.65 m/s/ºC C accuracy: + 0.05 m/s T accuracy: + 0.02ºC Resulting in resolution of 100 PPM
26
EXAMPLE 5: AMMONIUM HYDROX.
27
MAINTENANCE Functional checks Calibration Cell replacement Editing liquid parameters
28
TYPICAL ADVANTAGES No moving parts No drift In-line Highly accurate
29
FUNCTIONAL CHECKS Sample check Pure water check (checks both c & T) Calibration of temperature circuit Negligible drift of sound speed circuit
30
CALIBRATION INSTRUMENT ELECTRONICS Calibration of analog circuits: mA outputs PT100 inputs PROBE PT100 verification with water check
31
CELL REPLACEMENT NEW CELL - ENTRY OF NEW PARAMETERS: Path length Dead time Temperature offset ADVANTAGES: Sensor is pre-calibrated Independent of liquid
32
EDITING LIQUID PARAMETERS Liquid name Liquid number Polynomial constants Field calibration factors
33
RHOSONICS ANALYSIS INSTRUMENTS YOUR PARTNER FOR ULTRASONIC IN-LINE CONCENTRATION ANALYSIS