DOSIMETRY COMMISSIONING OF THE LNS-INFN PROTON THERAPY FACILITY THE DOSIMETRIC CHARACTERISTICS OF NARROW PROTON BEAMS USED IN EYE THERAPY HAVE BEEN DEFINED BY USING DIFFERENT RADIATION DETECTORS IONIZATION CHAMBERS (parallel-plate,minithimble chambers) SILICON DIODES RADIOCHROMIC AND RADIOGHRAPHIC FILMS THERMOLUMINESCENCE DETECTORS 1) CENTRAL AXIS DEPTH-DOSE DISTRIBUTIONS 2) LATERAL (TRANSVERSE) PROFILES 3) OUTPUT FACTORS (FSDF) TO PROVIDE INPUT TO THE TPS AND FOR M.U. CALCULATIONS. QUALITY CONTROL PROCEDURES DOSIMETRIC COMMISSIONING OF 60 MeV PROTON BEAMS PRODUCED AT LNS FOR EYE THERAPY SHAPED WITH 5 - 30 mm CIRCULAR BRASS COLLIMATORS. INFN-LNS
2
PROTON DEPTH-DOSE DISTRIBUTIONS THE PTW MARKUS PLANE-PARALLEL ION CHAMBER WAS USED IN THE COMMISSIONING AS THE REFERENCE DETECTOR FOR DEPTH-DOSE MEASUREMENT IN PROTON BEAMS (ICRU 59, IAEA 398) THE DESIGN OF THE MARKUS CHAMBER 1) Active volume=0.05 cm 3, 2) Electrode separation=2 mm, 3) Collector diameter=5.4 mm 4) t entrance window =2.3 mg/cm 2 5) Exact Location of p eff. PROVIDES DEPTH-DOSE DISTRIBUTIONS WITH HIGH SPATIAL RESOLUTION AND HIGH PRECISION ) PMMA CUP IS PROVIDED TO SEAL THE CHAMBER FOR USE IN WATER ) A SPACER IS PROVIDED FOR USE IN A SOLID PHANTOM, WHICH ALSO SERVES AS A HOLDER FOR A RADIOACTIVE SOURCE FOR STABILITY CHECKS. INFN-LNS
3
PTW MARKUS PLANE-PARALLER ION CHAMBER IN WATER PHANTOM INFN-LNS
4
^ taken from BJR Supplement 25 (1996) CENTRE MAXIMUM RANGE (mm) EQUIVALENT ENERGY (MeV) PEAK PLATEAU RATIO F.W.H.M. (mm ) Distal-dose falloff d 90%-10% (mm) CATANA 30.60 60.2 4.68 3.29 0.81 CCO^ 31.0 60.4 4.85 3.65 0.80 PSI^ 30.0 60.0 4.47 1.10
5
THE NEW SCANDITRONIX Si-DIODE IN PROTON BEAMS (Proton Field waterproof Detector-narrow beam no.DZA192 1001) INFN-LNS
6
THE NEW SCANDITRONIX Si-DIODE IN PROTON BEAMS (Proton Field waterproof Detector-narrow beam no.DZA192 1001) CYLINDRICAL MINIDIODE SPECIFICATIONS Detector Material: Hi-pSi, high doped p-type silicon (preirradiated for use in proton beams). Detector diameter: 0.6 mm (t = 60 m) DETECTOR PEAK DEPTH (mm) PEAK- PLATEAU RATIO F.W.H.M (mm) Distal-dose falloff (1) d 90%-10% (mm) Distal-dose falloff (2) d 80%-20% (mm PRACTICAL RANGE (d 10%, ICRU 59) MARKUS PTW 30.14 4.68 3.19 0.71 0.50 31.15 DIODE SCANDITRON IX 30.07 4.89 3.07 0.80 0.60 31.06 INFN-LNS
7
MODULATED PROTON DEPTH DOSES FOR EYE THERAPY MODULATOR RANGE SHIFTER MODULATION (SOBP) (mm eye tissue) DISTAL-DOSE FALLOFF (1) d 90%-10% (mm eye-tissue) DISTAL-DOSE FALLOFF (2) d 80%-20% (mm eye tissue) MAXIMUM DOSE IN SOBP % BEAM RANGE (90% DISTAL) (mm eye-tissue) MOD: 000/00 RS: 14 mm 14.5 0.91 0.65 101.6 14.38 MOD: 010/02 RS: 10 mm 10.5 0.90 0.60 104 18.27 MOD: 009/02 RS: 1.8 mm 20.30 0.80 0.55 103.9 27.04 CCO 17 0.900.75< 102 INFN-LNS
8
15 MM SOBP vs RANGE SHIFTER MODULATED PROTON DEPTH DOSES FOR EYE THERAPY
9
BEFORE EACH TREATMENT PERIOD TREATMENT DEPTH DOSE PROFILES HAVE TO BE VERIFIED. TO VERIFY RANGE MODULATION AND MAXIMUM RANGE OF THE BEAM DEPTH-DOSE MEASUREMENT WHEEL (CCO DIODE SCANNER) WHEEL IS COMPUTER CONTROLLED, AND RELATIVE DOSE MEASURED AS RATIO BETWEEN WHEEL’S DETECTOR AND REFERENCE DETECTOR TOLERANCES 1) [(MEASURED RANGE) / (REQUIRED RANGE)]: 0.2 mm 2) [(MEASURED MODULATION) / (REQUIRED MODULATION)]: [ -0.1 mm // +0.9 mm ] INFN-LNS
10
LATERAL OFF-AXIS BEAM PROFILES 1) LATERAL PENUMBRA: d 80%-20% 2) Field ratio: 3) L 95% 4) SIMMETRY (AREA RATIO): 5) FLATNESS: MD-55-2 RADIOCHROMIC FILMS WERE USED IN COMMISSIONING tissue equivalence – dose rate indipendence Linear dose response – high spatial resolution ONLY ONE CALIBRATION FILE IS NEEDED TO EVALUATE FILMS EXPOSED AT DIFFERENT DEPTHS Energy indipendence 1.3 mm
11
PMMA Phantom He-Ne Scanning Laser Densitomiter
12
INFN - LNS EXPERIMENTAL SET-UP ( =25 mm) SOBP (WIDTH)HL 95% (mm) S r% (simmetry) R t% (flatness) Lateral penumbra (mm) d 80%-20% MOD: 000/00 RS: 14 mm 14 mm z irr. =7 mm 0.9223 3% 1.40 MOD: 010/02 RS: 10 mm 10 mm z irr. = 12 mm 0.9223 2% 3% 1.30 MOD: 009/02 RS:1.8 mm 20 mm z irr. = 14 mm 0.9323 3% 1.15 CCO 18 z irr. = ½ SOBP 3% 1.10 CAL 18 z irr. = ½ SOBP 1.50
13
KODAK XV FILMS AND SCANDITRONIX DIODE WERE USED IN COMMISSIONING IF CALIBRATION FILES PRODUCED FOR KODAK XV FILMS AT DEPTH OF SOBP ARE USED FOR TRANSVERSE BEAM MEASUREMENTS GAF MD55-2 = KODAK XV FILM MOD 010/02 RS:10 MM z irr. =12 mm Lateral penumbra (mm) d 80%-20% H L 95% (mm) S r% (simmetry) KODAK XV 1.300.9222.80.50 MD-55-2 1.270.9222.70.90 INFN-LNS
14
BEFORE EACH TREATMENT PROTON BEAM PROFILES AT ISOCENTRE ARE TO BE CHECKED IN AIR X-Y DIODE SCANNING COMPUTER CONTROLLED DEVICE PLANNED TOLERANCES 1)LATERAL PENUMBRA (d 80%-20% ) 1.50 mm 2) BEAM SIMMETRY (S r ) 3% 3) BEAM FLATNESS: R t % 3% 4) FIELD RATIO: H 0.90 INFN-LNS
15
DOSE MONITORING SYSTEM (IN-BEAM DOSE MONITORS) THE PROTON DOSE IS MONITORED BY TWO INDEPENDENT TRANSMISSION UNSEALED ION CHAMBERS, PLACED IMMEDIATELY UPSTREAM OF THE PROTON NOZZLE. TRANSMISSION CHAMBERS HAVE SEPARATE CABLING, BIAS SUPPLIES (800 V) AND CURRENT INTEGRATORS AND ARE ARRANGED AS A REDUNDANT COMBINATION TO TERMINATE THE BEAM. TRANSMISSION ION CHAMBERS ARE CALIBRATED DAILY AGAINST A REFERENCE PARALLEL-PLATE MARKUS ION CHAMBER, LOCATED AT THE ISOCENTRE. COMMISSIONING TESTS 1) PRECISION 1a) SHORT TERM PRECISION: (CV) OF THE RATIO R OF DOSE MONITOR UNITS TO DOSIMETER SCALE READING FOR n=5 CONSECUTIVE IRRADIATIONS OF 15 Gy. MEASURED CV = 0.1% 1b) LONG TERM PRECISION (WEEKLY STABILITY): MEASURED [ (cGy/U.M.) WEEK ] : 1.5%. 2) LINEARITY THE RATIO V BETWEEN THE MEASURED PROTON DOSE AND DOSE MONITOR UNITS SHALL BE WITHIN 1% AT ALL MONITOR SCALE READINGS, i.e. IN THE CLINICAL RANGE UP TO 15–20 Gy. MEASURED RATIO INFN-LNS
16
DOSE MONITORING SISTEM TRANSMISSION ION CHAMBER INFN - LNS
17
DOSE RATE MONITORING (BEAM INTENSITY MONITOR) THE PROTON BEAM RATE IS MONITORED BY THE VOLTAGE SIGNAL (V SF ) PROVIDED BY THE FIRST SCATTERING FOIL, LOCATED IN VACUUM IMMEDIATELY UPSTREAM OF THE EXIT WINDOW. I BEAM (nA) = 0.0412 + 6.0835 V SF (R 2 =0.9999) (V SF ) IS USED BY THE C.S. TO STOP THE BEAM IF THE RATE XCEEDS A PRESET LIMIT. NORMAL TREATMENT BEAM CURRENT 4 nA CLINICAL DOSE RATE =12 20 Gy/min. Normal current Alarm current Z markus = 1 mm Full energy beam U.M. =K INFN-LNS
18
ABSOLUTE DOSIMETRY (BEAM CALIBRATION, cGy/U.M.) FOR REFERENCE DOSIMETRY, AT LNS, A PLANE-PARALLEL MARKUS ION CHAMBER IS USED IN A WATER PHANTOM, BY EXTENDING TO PROTON BEAMS THE FORMALISM OF THE IAEA 381 CODE OF PRACTICE ON THE USE OF PLANE-PARALLEL ION CHAMBERS IN HIGH ENERGY ELECTRON DOSIMETRY. cGy (nC) cGy/nC (s w,air ) q p Q = 1.00 (ICRU 59, IAEA TRS-398, MED.PHYS. 1995 :22) ( TAB. 7.1 ICRU 59) (W air /e) c p = 1.031 (ICRU 59) FOR DOSE CALIBRATIONS OF INDIVIDUAL PATIENTS, THE CALIBRATION IS MADE IN A UNIFORM DOSE REGION AT THE MIDDLE OF SOBP, AT ISOCENTRE, WITH THE 25 mm DIAMETER REFERENCE COLLIMATOR ON BEAM DELIVERY NOZZLE. FOR DAILY DOSE CHECKS, A PMMA PHANTOM BLOCK IS USED: A RADIOACTIVE CHECK DEVICE ( 90 Sr) IS USED FOR OPERATIONAL AND CONSTANCY CHECKS OF DOSIMETRY SYSTEM ( 1%, NCRP).
19
AN EXRADIN T1 THIMBLE ION CHAMBER [A-150 TE, 0.05 cm 3, N K ( 60 Co)] CAN BE USED AT LNS IN THE WATER PHANTOM, ACCORDING TO ICRU 59 BECAUSE OF THE Cavity length AND Outer diameter of the chamber, USE IS LIMITED TO ( ) 15 mm AND (SOBP WIDTH) 12 mm. WORK IN PROGRESS NOW AVAILABLE BY PTW N D,w, 60 Co FOR T1 EXRADIN AND MARKUS LNS CHAMBERS ADOPTION OF IAEA TRS-398 CODE OF PRACTICE D w, Q = M Q N D,w,Q0 k Q,Q0 ESTIMATED UNCERTAINTY OF D w, Q UNDER REFERENCE CONDITIONS (1 S.D.) ICRU 59 ( N k =1%) IAEA TRS-398 ( N D,w,Q0 =1%) EXRADIN T1 2.6% 2.1% PTW MARKUS 3.1% 2.5% INFN-LNS
20
PROTON DOSIMETRY INTERCOMPARISON (ECHED, ICRU 59) BEFORE STARTING LNS PROTON THERAPY FACILITY ) INTERCOMPARISON (1) AT CCO (PHYSICA MEDICA, VOL.XV, N.3) LNS (PTW MARKUS, T1-EXRADIN) CCO (A-150 FW-IC18, REFERENCE DOSE) (D w ) [(PTW MARKUS) / (FW-IC18)]: 1% (D w ) [(T1-exradin) / (FW-IC18)]: 1.4% ) INTERCOMPARISON (2) AT PSI (PHYSICA MEDICA VOL.XVII, S.3, PTCOG XXX 1999) 1) CCO (FW-IC18, MARKUS) 2) LNS (MARKUS, EXRADIN-T1, PTW PMMA FARMER) 3) PSI (PTW GRAPHITE FARMER, MARKUS) 4) TERA ( MARKUS, EXRADIN-T1) % = 1.2% max (%) = 3.5% INFN-LNS
21
DOSIMETRY INTERCOMPARISON AT PSI (FULLY MODULATED BEAM) INFN-LNS
22
OUTPUT FACTORS (FSDF) ) AT THE INITIAL CALIBRATION OF CYCLOTRON THE OUTPUT DOSE RATE HAS BEEN MEASURED FOR ALL COLLIMATORS AREA ENCOUNTERD IN THE CLINICAL PRACTICE. ) TO EVALUATE IF THERE ARE SIGNIFICANT DROPS OF DOSE PER MONITOR UNIT WHEN COLLIMATOR DIAMETER DECREASES TO A FEW MILLIMETERS. ) THE MOST RELIABLE RELATIVE OUTPUT VALUES ARE PROVIDED FROM RADIOCHROMIC DETECTORS, ESPECIALLY FOR NARROWEST BEAMS (VATNITSKY). NO SIGNIFICANT DECREASE OF BEAM OUTPUT (cGy/U.M.) FOR COLLIMATOR DIAMETER UP TO 5 mm INFN-LNS
23
OUTPUT FACTORS (FSDF) ) FROM EXPERIMENTAL RESULTS WE CAN STATE THE LOWER BOUND OF COLLIMATOR DIAMETER FOR WHICH FRELATIVE OUTPUT IS MEASURED ACCURATELY WITH EACH DETECTOR 1) MARKUS CHAMBER CAN BE USED IN PROTON BEAMS WITH 12 mm 2) TLDs AND SCANDITRONIX DIODE CAN BE USED UP TO =8 mm IN CLINICAL PRACTICE AT LNS GAF-DETECTOR IS USED FOR EVALUATION OF OUTPUT FACTOR FOR SHAPED NARROWEST BEAMS INFN-LNS
24
PATIENT DOSES (STRAY RADIATION) (PERSONAL MONITORING, TWO PATIENTS TESTED) A) TWO PERSONAL BADGE-DOSIMETERS WERE PLACED ON THE CHEST OF PATIENTS DURING THE WHOLE PROTON TREATMENT. 1) ENEA FILM BADGE PERSONAL DOSIMETER ( , , x) 0.05 mSv 2) NRPB PADC NEUTRON PERSONAL DOSEMETER 0.20 mSv PATIENTS COMPLETING TREATMENT INDUCED DOSE RATE 15 Sv/h (10’) 1.5 Sv/h INFN-LNS