Magnus Pro Training Presentation
Image analysis Software Note: The contents of this presentation are only showing the procedure for calibration, counting & generating reports. The dummy data have been shown just for training purpose
2
Calibration For any measurements calibration is required to be performed through standard stage micrometer. Formulae : Micrometer scale 1mm = 1000 Micron (e.g. 0.01mm= 10 micron) It should be performed on all objectives of microscope where digital camera or MIPS are installed. In case of readjustments or new replacement of any part of microscope or camera, the calibration should be done again. (contd.)
3
Select Magnification before calibration
Before going to measurement, one has to select appropriate magnification of calibration
4
Calibration (contd.) procedure
Focus stage micrometer under the first objective of microscope in X direction. Capture image with proper names. Focus stage micrometer under the first objective of microscope in Y direction. Repeat above procedure for any other objectives like 10X, 40X, 100X etc. (as per availability) (contd.)
5
Screen Dumps (calibration)
For objective 100X in X and Y direction
6
Calibration Procedure for X direction
Give the suitable name of calibration and save If line is for e.g.. 60 pixels Then write down same reading in the Scale X value Load image captured in X axis for 100X obj. Move line slowly with the help of Zoom Preview and release the line once it reaches to the other line Zoom Preview Calibration Procedure for X direction
7
Calibration Procedure for Y direction
Give the suitable name of calibration and save Move line slowly with the help of Zoom Preview and release the line once it reaches to the other line If line is for e.g.. 60 pixels Then write down same reading in the Scale Y value Load image captured in Y axis for 100X obj. Zoom Preview Calibration Procedure for Y direction
8
Measurement The measurements can be obtained after pixel calibration.
The units of measurements depends on the currently selected spatial calibration (for e.g. micron or inches or millimeter etc.)
9
Measurements of selected images
Images on which measurements are calculated. Selected calibration
10
Count Used to automatically identify objects in an image.
To get the several feature of the measurements.
11
Click Count to count the no. of objects
Automatically identified objects in an image obtained the several features of the measurements
12
Automatic Report Generation
13
Select File for report to be generated
Click Report data and select appropriate data on which report is to be generated