www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM
2
www.chemometec.com NucleoCounter SCC-100 Stand-alone fluorescence microscope with built in computer for image analysis Disposable sampling and measuring device, preloaded with nuclei staining dye
3
www.chemometec.com Easy operation 1. Lyse2. Load3. Count
4
www.chemometec.com Fast Result + = 30 seconds ------------------ max 1 minute
5
www.chemometec.com Base technology of ChemoMetec Fluorescence Microscopy
6
www.chemometec.com Integrated Fluorescence Microscope Lenses: magnification <1X. Excitation filter: sorts out green light. Emission filter: sorts out long wave length including red light. fixture
7
www.chemometec.com NucleoCassette
8
www.chemometec.com Propidium iodide (PI) Immobilized in the interior of the NucleoCassette, 2.8 ug in each A fluorescence dye, which intercalates with DNA of cell nuclei Cell membrane permeable
9
www.chemometec.com 1 dye molecule per 4-5 bp Genome between 2.5 x 10 9 – 4.0 x 10 9 bp. Propidium Iodide DNA base pair (bp): A: Adenine T: Thymine G: Guanine C: Cytosine C 27 H 34 N 4 2+.2I - Mr 668.4
10
www.chemometec.com Reagent C-100 Disrupts the plasma membranes of the mammalian cells, rendering the nuclei susceptible to staining with PI
11
www.chemometec.com Total cell count Viable Non-viable Free nucleus PI PI-stained nucleus Lysis buffer (Reagents C)
12
www.chemometec.com Sensitive parts, moving parts and “wear and tear” parts have been taken out of the instrument and integrated in the disposable NucleoCassette - Pump - Tubing’s - Mixer - Precision optics - Chemicals - Calibration - Waste container The NucleoCassette
13
www.chemometec.com Calibrated SCC-cassettes
14
www.chemometec.com Highly reliable LEDs Thermal drift in optical intensity is approximately 0.2 %/°C
15
www.chemometec.com Key Advantages Simple Operation –Ready in 5 seconds –5 minutes operator training –No cleaning Calibration Free –LED's, CCD and fixed optics assure stable factory calibration –Individually calibrated cassettes assure long term stability Maintenance Free –Robust and reliable parts used in design –2 years product guarantee Operator Independent –Objective cell counting Safe sample handling –PI immobilized inside NucleoCassette
16
www.chemometec.com Good reproducibility Results from different NucleoCounters can be directly compared: –All NucleoCounters are calibrated to respond as the same Master NucleoCounter at ChemoMetec. –The initial calibration will maintain throughout the lifetime of a NucleoCounter.
17
www.chemometec.com Repeatability Repeatability error dominated by count statistics (Poisson distribution) CV = 1/ x 100%
18
www.chemometec.com Repeatability Repeatability*: - CV < 8% at 150,000 cells/ml - CV < 6% at 300,000 cells/ml - CV < 4% at 500,000 cells/ml - CV < 3% at 1,000,000 cells/ml CV = Repeatability error Coefficient of Variance
19
www.chemometec.com Reproducibility Comparison of two instruments from different production series - Factory Settings!
20
www.chemometec.com Reproducibility Only significant method contribution is repeatability –In test more than 95% of the intra-laboratory reproducibility explained by repeatability
21
www.chemometec.com Routine Milk Samples
22
www.chemometec.com Milk Application SCC Stability –The inherent stability of the NucleoCounter SCC-100 offers tools to monitor and assure stability of SCC results obtained by routine instruments Pilot Samples –The NucleoCounter SCC-100 allows fast and reliable determination of SCC pilot samples
23
www.chemometec.com Somatic Cell Counting NucleoCounter SCC-100 –Stand-alone fluorescence microscope –Maintenance free –Calibration free NucleoCassette –Pre-loaded with nuclei staining dye –Analysed volume 1µl (working factor 1,000) –Calibrated measurement chamber (Dot code) Cell Counting –Operator independent –Reproducibility - Appr. Equal to Repeatability! –Repeatability Error - Appr. Poisson statistics (≈5% @ 500,000/ml) –Sample throughput – About 100 samples/hour –SCC - 10,000/ml to 1,000,000/ml